Skip to content

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in… recallZ-0002-2027 · Instrumentation Laboratory · initiated Sep 3, 2026

Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.

FDA root cause category: Under Investigation by firm

What the company is doing

Instrumentation Laboratory notified consignees on about 09/03/2026 via emailed and mailed URGENT MEDICAL DEVICE REALL letter. Consignees were made aware of the issue and hazard involved. They were instructed to check inventory for affected lots, destroy any remaining product following each facility's standard procedures, and document the destruction on the enclosed Urgent Mandatory Response Tracking Form, including quantity destroyed and date of destruction. Consignees were asked to identify an alternative method for HIT antibody testing as a supply interruption is expected and work local Werfen representatives for guidance and transition support. Consignees were also requested to inform all laboratory personnel of the notification, forward the notification to all affected departments and locations within each facility, retain a copy of the notification for records and complete and return the Urgent Mandatory Response Tracking Form within 10 days of receipt even if there is no remaining inventory. On 09/29/2026, an expanded URGENT MEDICAL DEVICE RECALL Notification was sent to consignees through registered mail or email. The updated notification informed consignees of additional lot numbers affected. Instructions remained the same.

Recall details

Recall number
Z-0002-2027
Recall class
Not yet classified by the FDA
Status
Open, Classified
Initiated by the company
Sep 3, 2026
Posted by the FDA
Oct 10, 2026
Recalling firm
Instrumentation Laboratory
Quantity
1,414 units
Product code
LCO
FDA submissions
K170854
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.

Product description

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

Code information (models, lots, serial numbers)

Part Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026).

Other devices under the same FDA submission

1 device cleared under K170854 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99843).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page