Medline surgical convenience kits labeled as: CABG… recallZ-0168-2027 · Medline Industries, LP · initiated Sep 2, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com, Recall Reference #: R-26-201-FG, Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable." Those consignees who further distributed the product were directed to notify the parties to whom the affected product was distributed. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0168-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Sep 2, 2026
- Posted by the FDA
- Oct 15, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 3125 kits
- Product code
- OEQ
- Distribution
- US Nationwide distribution.
Product description
Medline surgical convenience kits labeled as: CABG AUXILIARY PACK, Medline Product Number/SKU DYNJ44925D
Code information (models, lots, serial numbers)
UDI/DI 10193489461725 (each) 40193489461726 (case), Lot Numbers: 25LMF813, 25IMF341, 25HBR484, 25FBP179, 25EBJ732, 25DBK745, 25CBK191, 25BBE801, 24LBB945, 24IBT658, 24IBD265, 24GBG100, 24EBC752, 24CBQ893, 24BBJ088, 24ABE008, 23JBA374, 23HBV035, 23GBL681, 23EBK074, 23DBL164, 23DBB447, 23ABV418, 22LBG352, 22IBK243, 22HBX829, 22EBA032, 22DBO915, 22BBR573, 21LBB905, 21IBX719, 21HBG260, 21GBH209, 21DBR261, 21BBJ878.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ44925DMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99886).
- Z-0169-2027 · Open, Classified1) Product Name AUD OPEN HEART PACK, Medline Kit Number/SKU DYNJ22052I; 2) Product Name AV SHUNT PACK-MARYVIEW, Medline Kit Number/SKU DYNJ24528N; 3) Product Name AV SHUNT PACK-M…
- Z-0170-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) OPEN HEART CDS, Medline product number/SKU CDS840094R; 2) OPEN HEART CDS, Medline product number/SKU CDS840171AG; 3) MHC AV FIS…
- Z-0171-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR PACK-LF, Medline product number/SKU DYNJ06658…
- Z-0172-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D
- Z-0173-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) EAR, Medline product number/SKU DYNJ909020B; 2) EAR, Medline product number/SKU DYNJ909020A; 3) CAROTID PACK, Medline product n…
- Z-0174-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) MAIN/LAPAROSCOPY TRANSPLA PACK, Medline product number/SKU DYNJ0656274AC; 2) MAIN/LAPAROSCOPY TRANSPLA PACK, Medline product num…
- Z-0175-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K; 2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J; 3) CV…
- Z-0176-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKU DYNJ04584L; 2) OPEN HEART ACCESSORY PACK, Medline Product number/SKU DYNJT5715A; …
- Z-0177-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M.
- Z-0178-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D
- Z-0179-2027 · Open, ClassifiedMedline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697C; 3) FEM POP PACK-LF…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 21, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 21, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 21, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page