Skip to content

Medline surgical convenience kits labeled as recallZ-0179-2027 · Medline Industries, LP · initiated Sep 2, 2026

Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com, Recall Reference #: R-26-201-FG, Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable." Those consignees who further distributed the product were directed to notify the parties to whom the affected product was distributed. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-0179-2027
Recall class
Not yet classified by the FDA
Status
Open, Classified
Initiated by the company
Sep 2, 2026
Posted by the FDA
Oct 15, 2026
Recalling firm
Medline Industries, LP
Quantity
6069 kits
Product code
OJH
Distribution
US Nationwide distribution.

Product description

Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697C; 3) FEM POP PACK-LF, Medline Product Number/SKU DYNJ0159820F; 4) AV FISTULA PACK, Medline Product Number/SKU DYNJ67139D; 5) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AD; 6) LAMINECTOMY PACK, Medline Product Number/SKU DYNJ17333K; 7) EXTREMITY-LF, Medline Product Number/SKU DYNJ905276F; 8) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AC; 9) SPINAL CDS, Medline Product Number/SKU CDS940087AL; 10) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697A; 11) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055; 12) CRANIOTOMY-RICHMOND, Medline Product Number/SKU CDS980929U; 13) ACB EXTRAS PACK-6211, Medline Product Number/SKU DYNJ49516; 14) SPINAL CDS, Medline Product Number/SKU CDS940087AK; 15) SPINAL CDS, Medline Product Number/SKU CDS940087AJ; 16) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AB; 17) SPINAL CDS, Medline Product Number/SKU CDS940087AI; 18) LAMINECTOMY PACK, Medline Product Number/SKU DYNJ17333J; 19) SPINAL CDS, Medline Product Number/SKU CDS940087AG; 20) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AA.

Code information (models, lots, serial numbers)

Show the full code information (16k characters)

Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 26DMG028; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 26DMF181; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 26DMD186; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 26CMA323; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 26BMJ234; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 26ALA854; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 26ADB485; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 26ADA395; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 26ABM111; Medline Product Number/SKU CDS780025AD, UDI/DI 10198459311598 (each) 40198459311599 (case), Lot Number 25LBH825; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25KMC475; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25KMB629; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25KLA526; Medline Product Number/SKU CDS780025AD, UDI/DI 10198459311598 (each) 40198459311599 (case), Lot Number 25KBP981; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 25JDB519; Medline Product Number/SKU CDS780025AD, UDI/DI 10198459311598 (each) 40198459311599 (case), Lot Number 25JBI857; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25IMC513; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25ILA706; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 25IDB222; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 25IBP736; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25HLA065; Medline Product Number/SKU CDS780025AD, UDI/DI 10198459311598 (each) 40198459311599 (case), Lot Number 25HBO604; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 25GDB198; Medline Product Number/SKU DYNJ905276F, UDI/DI 10198459247965 (each) 40198459247966 (case), Lot Number 25GBS381; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25GBQ923; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 25GBJ981; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25FLA160; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 25FDA332; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25FBN175; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 25EBV301; Medline Product Number/SKU CDS940087AM, UDI/DI 10198459311499 (each) 40198459311490 (case), Lot Number 25EBH152; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25DLA962; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 25DDB837; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 25DDA286; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 25DBC339; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25CLA883; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 25CBS904; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 25CBP237; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 25CBG304; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 25CBE419; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25ALA976; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25ALA345; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 25ALA067; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 25ABU061; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24LBE127; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 24KLA163; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 24KBR641; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24KBD715; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 24JLA181; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 24JDA289; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24JBQ903; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24JBQ698; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24JBG248; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24IBK612; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 24HBT188; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 24GDA765; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24GBE280; Medline Product Number/SKU DYNJ907697C, UDI/DI 10195327697617 (each) 40195327697618 (case), Lot Number 24FLA764; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 24FDA385; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24FBN519; Medline Product Number/SKU DYNJ67139D, UDI/DI 10195327668273 (each) 40195327668274 (case), Lot Number 24FBM920; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24FBE685; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 24DDB513; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 24DDA174; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24DBH403; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24CBP710; Medline Product Number/SKU DYNJ907697A, UDI/DI 10195327536459 (each) 40195327536450 (case), Lot Number 24BLA540; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24BBS351; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24BBR356; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 24BBI372; Medline Product Number/SKU DYNJ80055, UDI/DI 10195327008741 (each) 40195327008742 (case), Lot Number 24BBD195; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 24BBB443; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 24ADC061; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 24ADA704; Medline Product Number/SKU DYNJ907697A, UDI/DI 10195327536459 (each) 40195327536450 (case), Lot Number 23LLA416; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23LDA917; Medline Product Number/SKU CDS980929U, UDI/DI 10193489976427 (each) 40193489976428 (case), Lot Number 23LBC788; Medline Product Number/SKU DYNJ907697A, UDI/DI 10195327536459 (each) 40195327536450 (case), Lot Number 23KLB157; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 23KDA615; Medline Product Number/SKU CDS980929U, UDI/DI 10193489976427 (each) 40193489976428 (case), Lot Number 23KBT094; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23KBP002; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23KBN905; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 23KBC083; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23JDB177; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23IDB906; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 23IDB810; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23IBW076; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 23IBI612; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23IBD027; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23HBP378; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 23GDC095; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23FDB307; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23FBI843; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 23FBH173; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23EBL944; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 23EBJ659; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23DDB819; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23DBN024; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23CDA324; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 23CBX130; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 23BDB285; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 23ADB644; Medline Product Number/SKU CDS940087AL, UDI/DI 10195327297824 (each) 40195327297825 (case), Lot Number 23ABB068; Medline Product Number/SKU DYNJ49516, UDI/DI 10888277854369 (each) 40888277854360 (case), Lot Number 22LBM685; Medline Product Number/SKU CDS940087AK, UDI/DI 10195327257514 (each) 40195327257515 (case), Lot Number 22LBG216; Medline Product Number/SKU CDS780025AC, UDI/DI 10195327212278 (each) 40195327212279 (case), Lot Number 22LBF788; Medline Product Number/SKU DYNJ49516, UDI/DI 10888277854369 (each) 40888277854360 (case), Lot Number 22LBC322; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 22KDB263; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 22KDA291; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 22JDA178; Medline Product Number/SKU DYNJ0159820F, UDI/DI 10193489207057 (each) 40193489207058 (case), Lot Number 22IDA027; Medline Product Number/SKU CDS940087AJ, UDI/DI 10195327210427 (each) 40195327210428 (case), Lot Number 22HBO525; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 22HBB637; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 22GDA892; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 22GBH493; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 22GBE442; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 22EBN090; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 22EBG548; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 22EBF695; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 22EBA274; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 22CBP174; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 22BDC466; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 22BBY060; Medline Product Number/SKU DYNJ49516, UDI/DI 10888277854369 (each) 40888277854360 (case), Lot Number 22BBP492; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 21LBU758; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 21KBH163; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 21JBF936; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 21IBL720; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 21IBJ260; Medline Product Number/SKU CDS780025AB, UDI/DI 10193489846966 (each) 40193489846967 (case), Lot Number 21HBL624; Medline Product Number/SKU DYNJ17333K, UDI/DI 10193489948998 (each) 40193489948999 (case), Lot Number 21GDD271; Medline Product Number/SKU CDS940087AI, UDI/DI 10193489933758 (each) 40193489933759 (case), Lot Number 21GBI183; Medline Product Number/SKU DYNJ49516, UDI/DI 10888277854369 (each) 40888277854360 (case), Lot Number 21FBB329; Medline Product Number/SKU DYNJ17333J, UDI/DI 10193489791525 (each) 40193489791526 (case), Lot Number 21EDB082; Medline Product Number/SKU CDS940087AG, UDI/DI 10193489847260 (each) 40193489847261 (case), Lot Number 21EBC900; Medline Product Number/SKU DYNJ17333J, UDI/DI 10193489791525 (each) 40193489791526 (case), Lot Number 21CDA517; Medline Product Number/SKU CDS780025AA, UDI/DI 10193489483932 (each) 40193489483933 (case), Lot Number 21CBD927.

Devices named in this recall

20 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99886).

Sources and updates