Direxion recallZ-0233-2027 · Boston Scientific Cork, Ltd. · initiated Sep 16, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Cork, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA root cause category: Process control
What the company is doing
On September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Forward this notice to relevant personnel and any facilities to which the product was transferred. 3. Continue to follow the standard of practice when monitoring patients treated with the product. 4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 5. Return product to Boston Scientific in accordance with the enclosed instructions.
Recall details
- Recall number
- Z-0233-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 16, 2026
- Posted by the FDA
- Oct 22, 2026
- Recalling firm
- Boston Scientific Cork, Ltd.
- Quantity
- 3284
- Product code
- KRA
- FDA submissions
- K163701
- Distribution
- Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
Product description
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Material Number: M001195300; UDI-DI: 08714729839415; Batch Numbers: 37818213, 37818214, 37818215, 37818216, 37881115, 37886786, 37886787, 37886788, 37914463, 37914464, 37914465, 37943140, 37956679, 37957740, 37977611, 38002444, 38002445, 38085044, 38089225, 38092404, 38092405, 38092406, 38112470, 38112471, 38112472, 38149820, 38181217, 38191752, 38220950, 38330637, 38330638, 38415914, 38415915, 38444073, 38444074, 38484290, 38527320, 38527321, 38560452, 38573062, 38573063, 38585014, 38596980, 38596981, 38629571, 38629572, 38769972, 38769973, 38827700, 38827701, 38827702, 38827703, 38827704, 38827705, 38904809, 38904810, 38916514, 38916515, 38937265, 38990849, 38990852, 38990856, 38990857, 39029815, 39143228, 39143230, 40466130, 40466131, 40466631, 40471499, 40471940, 40476290, 40476291, 40476292, 40479473, 40479474, 40479716, 40479717, 40492475, 40493319, 40493333, 40493440, 40497672, 40499680, 40499681, 40505968, 40505969, 40512591, 40512592, 40512593, 40513026, 40513027, 40521923, 40521924, 40525798, 40525799, 40525860, 40531461, 40531462, 40533302, 40535971, 40535972;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Direxion™ M001195300Boston Scientific Corporation · Catheter, continuous flush1 recall15 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99906).
- Z-0223-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200
- Z-0224-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210
- Z-0225-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/155; Material Number: M001195220
- Z-0226-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/105; Material Number: M001195230
- Z-0227-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/130; Material Number: M001195240
- Z-0228-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/155; Material Number: M001195250
- Z-0229-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260
- Z-0230-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/130; Material Number: M001195270
- Z-0231-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280
- Z-0232-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290
- Z-0234-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/155; Material Number: M001195310
- Z-0235-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320
- Z-0236-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/155; Material Number: M001195330
- Z-0237-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/130; Material Number: M001195340
- Z-0238-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/155; Material Number: M001195350
- Z-0239-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400
- Z-0240-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/130; Material Number: M001195410
- Z-0241-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/155; Material Number: M001195420
- Z-0242-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430
- Z-0243-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440
- Z-0244-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/155; Material Number: M001195450
- Z-0245-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470
- Z-0246-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/155; Material Number: M001195480
- Z-0247-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/130; Material Number: M001195500
- Z-0248-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/155; Material Number: M001195510
- Z-0249-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/130; Material Number: M001195610
- Z-0250-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620
- Z-0251-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/130; Material Number: M001195640
- Z-0252-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650
- Z-0253-2027 · Open, ClassifiedDirexion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/2RO/130; Material Number: M001195660
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page