Skip to content

Direxion recallZ-0238-2027 · Boston Scientific Cork, Ltd. · initiated Sep 16, 2026

Class II Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Cork, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

FDA root cause category: Process control

What the company is doing

On September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Forward this notice to relevant personnel and any facilities to which the product was transferred. 3. Continue to follow the standard of practice when monitoring patients treated with the product. 4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 5. Return product to Boston Scientific in accordance with the enclosed instructions.

Recall details

Recall number
Z-0238-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 16, 2026
Posted by the FDA
Oct 22, 2026
Recalling firm
Boston Scientific Cork, Ltd.
Quantity
384
Product code
KRA
FDA submissions
K163701
Distribution
Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.

Product description

Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/155; Material Number: M001195350

Code information (models, lots, serial numbers)

Material Number: M001195350; UDI-DI: 08714729839460; Batch Numbers: 37881102, 37881103, 37943142, 37943143, 37957743, 38439681, 38439682, 38484292, 38527325, 38876320, 38904816, 38904817, 38916510, 38916512, 38990272, 38990273, 38990274, 39029898, 39092695, 39092696;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99906).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page