INTERSTIM recallZ-0267-2027 · Medtronic Neuromodulation · initiated Sep 8, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm began distributing Urgent Medical Device Correction letters to consignees via email or mail starting 9/15/26. The letter included recommendations for system and patient management: - Verify InterStim product names and model numbers are compatible prior to the procedure. - Review letter for a summary of compatible product information. - Consult the System Overview and Compatibility Insert at www.manuals.medtronic.com for additional product compatibility information. - Ensure a system integrity check is performed prior to completing the implant procedure per the applicable product manual. - Contact your Medtronic Representative or Technical Services for any additional support as needed.
Recall details
- Recall number
- Z-0267-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 8, 2026
- Posted by the FDA
- Oct 22, 2026
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 56,331 (US); 40,211 (OUS)
- Product code
- EZW
- FDA submissions
- P970004S302
- Distribution
- Worldwide distribution - US Nationwide and territory of Puerto Rico. The countries of Brazil, Colombia, Chile, Canada, Japan, New Zealand, Australia, United Kingdom, Switzerland, Sweden, Spain, Portugal, Italy, Israel, Germany, France, Finland, Canary Islands, Belgium, Austria, Norway, Northern Ireland, Netherlands, Iceland, Denmark, South Africa, Turkey, Slovakia, Saudi Arabia, Poland, Hungary, United Arab Emirates (UAE), Kuwait, Greece.
Product description
InterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 3560030
Code information (models, lots, serial numbers)
Show the full code information (25k characters)
3560022 UDI-DI 00763000203894 Lots VA24HE7 VA24R5J VA24TXT VA24UYB VA24WJE VA24WJF VA250XH VA25144 VA2515A VA25TEF VA25TEH VA25TEJ VA25TEL VA26A7F VA26AJA VA26AJB VA26AKS VA26AKT VA26AKU VA26AKV VA26C4T VA26CB1 VA26CB2 VA26EHL VA26EHN VA26EHP VA2725R VA2725S VA2726H VA2726J VA2728H VA2728K VA2728L VA275QE VA2760F VA277W0 VA277W1 VA27C85 VA27C86 VA27CCG VA27CEE VA27D4B VA27D62 VA27D63 VA27KEH VA27KGX VA27KHG VA27KHH VA27KHJ VA27KHK VA27KHL VA27NS0 VA27NSC VA27NSG VA27NSH VA27NSJ VA27NSL VA27ST1 VA27ST2 VA27UGK VA27UGL VA27UXN VA27UXP VA27VDZ VA27VED VA27VEE VA27VEF VA27VJF VA27VJG VA27VJJ VA27VWY VA27Y21 VA27Y22 VA27YF8 VA27YFG VA280D1 VA280E6 VA280LQ VA2848G VA28499 VA284MR VA284PJ VA284S3 VA286PS VA286PT VA2872W VA28AQH VA28BDU VA28BEW VA28BR6 VA28BR7 VA28DZR VA28E02 VA28EC2 VA28EC3 VA28GE3 VA28HC2 VA28HM9 VA28HSG VA28J1R VA28J1S VA28K8K VA28MHH VA28MHS VA28MYN VA28N0Y VA28N15 VA28N1R VA28T7D VA28T7F VA28TSM VA28TSP VA28UE7 VA28UHR VA28UHY VA28Z99 VA28ZJ5 VA28ZMP VA2976F VA297CK VA297CR VA297PV VA2990L VA29939 VA299J7 VA299J8 VA29EQE VA29EQF VA29EQG VA29EQJ VA29FSC VA29FSD VA29FSS VA29G5U VA29G94 VA29GKR VA29HCQ VA29HCU VA29HM5 VA29HM6 VA29N97 VA29QAK VA29QAL VA29QBK VA29QBL VA29QCE VA29QUW VA29QUZ VA29QVN VA29S6N VA29S6Q VA29SAV VA29SDN VA29SEY VA29SEZ VA29SF4 VA29SF5 VA29SYJ VA29Y62 VA29Y63 VA29YGH VA29YGJ VA29YKY VA29YW5 VA29Z8P VA29Z8Q VA29ZEG VA29ZN3 VA29ZSA VA29ZU3 VA29ZU4 VA29ZUG VA2A2Z3 VA2A5JQ VA2A5MK VA2A6GU VA2A6GY VA2A6L5 VA2A73R VA2A73U VA2A7EC VA2A7EF VA2A7HW VA2AEKR VA2AEM9 VA2AEWP VA2ALQ9 VA2ALUD VA2ALV0 VA2ALVA VA2ALW5 VA2APNN VA2APQA VA2APSH VA2APSU VA2ARAX VA2ARCA VA2ARCT VA2AYRR VA2AYRU VA2AYZ3 VA2AZ0W VA2AZ7J VA2AZ7R VA2B6PX VA2B8WD VA2B8WE VA2B8XZ VA2B8ZN VA2BBDN VA2BBDQ VA2BBT6 VA2BBU0 VA2BBWC VA2BBZM VA2BD21 VA2BDCD VA2BDCE VA2BHB0 VA2BHB9 VA2BHCD VA2BHCF VA2BHDU VA2BHDX VA2BHJN VA2BHPH VA2BHPK VA2BHPR VA2BUNB VA2BUNP VA2BUSW VA2BUVT VA2BXXB VA2BY1R VA2BY2N VA2BY4M VA2C4A9 VA2C4JZ VA2C582 VA2C585 VA2C7JX VA2C7LT VA2C7SB VA2CCJ0 VA2CCJ1 VA2CCJK VA2CCJL VA2CCJM VA2CCJN VA2CCJP VA2CCJR VA2CH8Y VA2CHC0 VA2CHET VA2CM6R VA2CM9W VA2CMA3 VA2CMAG VA2CMAL VA2CN36 VA2CPWP VA2CQ0U VA2CQ0Z VA2CQ1G VA2CQ23 VA2CQ2A VA2CSVR VA2CSW9 VA2CSWG VA2CSX2 VA2CT5M VA2CUN8 VA2CVB8 VA2CW13 VA2CZSU VA2CZSV VA2CZSY VA2CZT4 VA2D4PR VA2D4Q5 VA2D4QJ VA2D4QM VA2D4Z4 VA2D4ZU VA2D657 VA2D658 VA2D659 VA2D65A VA2D65B VA2D8B5 VA2DDCK VA2DDCU VA2DDD8 VA2DDDG VA2DDDT VA2DP4Z VA2DPAJ VA2DPCV VA2DPF5 VA2DPNR VA2DVAY VA2DVB5 VA2DVBA VA2DVBF VA2DVDA VA2DVP5 VA2DVPQ VA2E1PW VA2E1PY VA2E1Q1 VA2E1QE VA2E2EK VA2E491 VA2E494 VA2E4A5 VA2E4AC VA2ECQY VA2ECRM VA2ECS4 VA2ECTK VA2ECTM VA2ECTV VA2EGXE VA2EGXJ VA2EGXK VA2EGZ4 VA2EHT8 VA2EMDT VA2EME1 VA2EME9 VA2EMEP VA2EMF5 VA2EMFB VA2EMM1 VA2EMN8 VA2ENF9 VA2ENH4 VA2ENH7 VA2ENJG VA2ENJQ VA2ENJT VA2ENJV VA2ENKX VA2ENW3 VA2EP8C VA2ESRT VA2ESSH VA2ESSW VA2ESTM VA2ESTV VA2ESU0 VA2ESUA VA2ESUK VA2EYMM VA2EYNF VA2EYQQ VA2EYRB VA2F0TM VA2F0U6 VA2F0UV VA2F1B5 VA2F1G2 VA2F1G6 VA2F1GQ VA2F1HH VA2F1XE VA2F1XY VA2F1ZF VA2F202 VA2F21Q VA2F222 VA2F259 VA2F25E VA2F25V VA2F38C VA2F38K VA2F38T VA2F3DE VA2F3ET VA2F3FG VA2F3FX VA2F3G6 VA2F3K6 VA2F4BJ VA2F4E5 VA2F4KD VA2F4M7 VA2F4MY VA2FBYT VA2FBYW VA2FBZJ VA2FBZZ VA2FC00 VA2FC0H VA2FC0T VA2FPRU VA2FWNY VA2FWP1 VA2FWPK VA2FWQJ VA2FZRU VA2FZRY VA2FZRZ VA2FZS1 VA2FZS8 VA2FZSB VA2G05M VA2G05U VA2G06S VA2G06W VA2G06Z VA2G074 VA2G07T VA2G4WX VA2G4X0 VA2G4X5 VA2G4XS VA2G4XV VA2G6MT VA2G6NL VA2G6NZ VA2G6R9 VA2G6RE VA2G79V VA2G7A4 VA2G8J5 VA2G8J7 VA2G8J8 VA2G8S6 VA2G8UC VA2G8UH VA2G8WC VA2G8WP VA2G8WV VA2G8X1 VA2G8X9 VA2G8XG VA2G8YN VA2G8Z6 VA2GDZE VA2GDZF VA2GDZH VA2GDZJ VA2GDZK VA2GDZM VA2GEA9 VA2GEAV VA2GEBH VA2GEDS VA2GEEP VA2GEF3 VA2GH1M VA2GH1W VA2GH25 VA2GH2F VA2GH3Q VA2GH9H VA2GKGQ VA2GKGW VA2GKH1 VA2GKH6 VA2GL2M VA2GL3F VA2GL3N VA2GL6Z VA2GT1C VA2GT1J VA2HF33 VA2HF3H VA2HF3Y VA2HJJU VA2HJJX VA2HJK5 VA2HJX0 VA2HJX4 VA2HJXN VA2HK92 VA2HK98 VA2HK9B VA2HK9C VA2HKK0 VA2HKKE VA2HLVK VA2HLVQ VA2HLW7 VA2HLW9 VA2HLWC VA2HLWD VA2HLWE VA2HMQR VA2HMR1 VA2HMRS VA2HMS7 VA2HMSN VA2HMSQ VA2HMTK VA2HT5K VA2HTGH VA2HTGU VA2HTJ6 VA2HTJP VA2HTK8 VA2HTKL VA2HTLR VA2HTP3 VA2HTPB VA2J72W VA2J73J VA2J74K VA2J74R VA2J7C2 VA2J7D5 VA2J7KQ VA2J7L8 VA2J7LN VA2J7ME VA2J7ND VA2J7VP VA2J7VV VA2J7VZ VA2J7WD VA2JL82 VA2JL87 VA2JL8P VA2JLBK VA2JLBU VA2JLUV VA2JLXK VA2JLXM VA2JLYG VA2JLZS VA2JLZT VA2JLZU VA2JLZV VA2JLZW VA2JLZX VA2JMW2 VA2JMWK VA2JMXM VA2JMYA VA2JMYG VA2JMZJ VA2JN1B VA2JN1M VA2JN2V VA2JQF8 VA2JQG0 VA2JQG5 VA2JQGJ VA2JQGQ VA2JQZG VA2JQZP VA2JQZY VA2JR04 VA2JR0F VA2JR0J VA2JR7R VA2JR86 VA2JR8A VA2JR8D VA2JR8L VA2JR8M VA2JRVJ VA2JRXU VA2JRYA VA2JRYE VA2JRYZ VA2JS0M VA2JSSZ VA2JST1 VA2JSTF VA2JSTH VA2JT5P VA2JT5Q VA2JT5Z VA2JT60 VA2JTCE VA2JTDT VA2JWBU VA2JWCA VA2JWDV VA2JWE5 VA2JWEC VA2JWF4 VA2JWT7 VA2JWTU VA2JWU8 VA2JWVF VA2JWVT VA2JWWC VA2JWXG VA2JX2S VA2JX2Z VA2JX3E VA2K89B VA2K8A0 VA2K8AA VA2KUVM VA2KUWM VA2KUWT VA2KVCU VA2KVD5 VA2KVDC VA2KVP4 VA2KVQU VA2KVQY VA2KVRA VA2KVU4 VA2KVX9 VA2KVXC VA2KW4P VA2KW4R VA2KWQW VA2KWSG VA2KWSJ VA2KWSP VA2KWSU VA2KWSY VA2KWTS VA2KWU1 VA2KWUE VA2KWUZ VA2KX2P VA2KX51 VA2KX5D VA2KX5N VA2KX61 VA2KX6F VA2L1YH VA2L1YK VA2L1YM VA2L263 VA2L26M VA2L2C7 VA2L2DP VA2L2E0 VA2L7J0 VA2L7JE VA2L7KW VA2L88B VA2L88N VA2L892 VA2L898 VA2L89K VA2L9ZG VA2L9ZK VA2LA10 VA2LA1B VA2LA1F VA2LA1J VA2LA8G VA2LA9N VA2LAA8 VA2LABL VA2LAC8 VA2LACS VA2LH21 VA2LH35 VA2LH4C VA2LH4S VA2LH63 VA2LH6F VA2LH6G VA2LH6L VA2LTAZ VA2LTC1 VA2LTE6 VA2LTEQ VA2LU6K VA2LU6P VA2LUA5 VA2LUCH VA2LUES VA2LUFW VA2LUJ4 VA2LUJG VA2LVG8 VA2LVGG VA2LVGR VA2LVLA VA2LVT2 VA2LVTM VA2LVTV VA2LVUK VA2LVX2 VA2LVXH VA2LWED VA2LWEE VA2LWEH VA2LWM7 VA2M0GB VA2M0GK VA2MHSZ VA2N7R7 VA2N7S3 VA2N7TZ VA2N7W1 VA2N7W4 VA2N7Z5 VA2N7ZQ VA2N7ZY VA2N80G VA2N81E VA2N8GG VA2N8HQ VA2N8RZ VA2N8TM VA2NCAS VA2NCB1 VA2NCDL VA2NCE0 VA2NCF7 VA2NCGR VA2NCJ2 VA2NCLH VA2NCLR VA2NCLX VA2NCMW VA2NCPG VA2NCR3 VA2NCRG VA2NCS5 VA2NCS7 VA2NN6S VA2NN75 VA2NN78 VA2NN7R VA2NN7W VA2NN95 VA2NN96 VA2NN9V VA2NNP5 VA2NNP8 VA2NNPP VA2NNQ8 VA2NNRP VA2NNSP VA2NNUW VA2NP1Q VA2NP2B VA2NP2H VA2NP2K VA2NP2Z VA2NWP5 VA2NWQ7 VA2NWQN VA2NWRH VA2NWRY VA2NWS1 VA2NWS2 VA2NX2N VA2NX8Y VA2NX97 VA2NXBV VA2NXCA VA2NXCU VA2NXDP VA2NXHQ VA2NXJ6 VA2NXK3 VA2NXKQ VA2NXKZ VA2NXLA VA2P2MD VA2P2ML VA2P2MU VA2P2MZ VA2P2N4 VA2P2NB VA2P39D VA2P39V VA2P39Y VA2P3A8 VA2P3AA VA2P3AK VA2P4J8 VA2P4JF VA2P4JS VA2P4K8 VA2P4MJ VA2P4WF VA2P4WV VA2P4X5 VA2P4ZG VA2P503 VA2P50E VA2P50W VA2P9W3 VA2P9YX VA2P9ZG VA2P9ZZ VA2PA13 VA2PA1Q VA2PA8W VA2PA9V VA2PAA8 VA2PAAP VA2PABJ VA2PADJ VA2PAE8 VA2PCL7 VA2PCM8 VA2PCNW VA2PCP0 VA2PCP6 VA2PCPS VA2PCQE VA2PD2E VA2PD57 VA2PD6T VA2PD7J VA2PD7R VA2PFK1 VA2PFLU VA2PFLZ VA2PG39 VA2PG3N VA2PG8M VA2PG8W VA2PGA4 VA2PGC5 VA2PGCE VA2PGGZ VA2PGH0 VA2PGH2 VA2PGJ7 VA2PGP6 VA2PGPW VA2PGS3 VA2PGSK VA2PGUT VA2PGV7 VA2PPDU VA2PPEN VA2PPEU VA2PPFA VA2PPFB VA2PPFH VA2PQ01 VA2PQ0R VA2PQ18 VA2PQ43 VA2PQ5C VA2PQ5U VA2PTY2 VA2PTY5 VA2PTY9 VA2PTYM VA2PTYZ VA2PTZ3 VA2PUBM VA2PUBT VA2Q575 VA2Q57K VA2Q57V VA2Q58C VA2QEVH VA2QEVM VA2QEZ4 VA2QEZA VA2QFGA VA2QFHF VA2QJGC VA2QJGL VA2QJGQ VA2QJGU VA2QJRL VA2QJRS VA2QJW5 VA2QJWB VA2QWB5 VA2QWB9 VA2QWDY VA2QWH2 VA2QWHM VA2QWJ0 VA2QWJK VA2QWJY VA2QWLT VA2QWMM VA2QWN2 VA2QWN6 VA2QWNG VA2QWNJ VA2QWWT VA2QWX9 VA2QWXF VA2QWXX VA2QWY8 VA2QWYV VA2QX54 VA2QX5H VA2QX5N VA2QXAN VA2QXBP VA2QXC7 VA2RABG VA2RABL VA2RABQ VA2RABW VA2RACE VA2RACH VA2RAJL VA2RAJM VA2RAJN VA2RAJP VA2RAJQ VA2RAJR VA2RDMC VA2RDMK VA2RDN2 VA2RDPY VA2RDQ8 VA2RDR1 VA2RDRW VA2RDRZ VA2RDS2 VA2RDS4 VA2RDS5 VA2RDS6 VA2RDS7 VA2RDS8 VA2RE4D VA2RE4E VA2RE4R VA2RE4T VA2RE4X VA2RE57 VA2RE5F VA2RNM7 VA2RNMF VA2RNNB VA2RNNF VA2RNNG VA2RNNK VA2RNNQ VA2RP22 VA2RP3D VA2RP5A VA2RP5Y VA2RZ12 VA2RZ1K VA2RZ2G VA2RZ2T VA2RZ36 VA2RZ3G VA2RZ3P VA2RZ6S VA2S015 VA2S0KR VA2S1BU VA2S1CG VA2S1D1 VA2S1KD VA2S1KF VA2S1KY VA2S1L5 VA2S1SY VA2S1T3 VA2S1TZ VA2S1V2 VA2S1V6 VA2S1VA VA2S4ZL VA2S501 VA2S515 VA2S51H VA2S51L VA2S51N VA2S52L VA2S52N VA2S52S VA2S52Y VA2S53G VA2S6RH VA2S6RX VA2S6SJ VA2S6U3 VA2S6UB VA2S70Y VA2S720 VA2S8UH VA2S8UY VA2S8VJ VA2S8W8 VA2S8WB VA2S8ZS VA2S8ZW VA2S90K VA2S91X VA2S92R VA2S93G VA2S94M VA2S9V8 VA2SCTH VA2SCTN VA2SCTR VA2SCU1 VA2SCV3 VA2SDKL VA2SDKP VA2SDLB VA2SDMY VA2SDP3 VA2SX34 VA2SX4F VA2SX67 VA2SX6J VA2SXBN VA2SXCU VA2SYRT VA2SYS4 VA2SYSE VA2SYSS VA2SYTB VA2SYVA VA2SYYA VA2SYYB VA2SYYE VA2SYYJ VA2T019 VA2T01G VA2T03K VA2T04C VA2T0C2 VA2TNME VA2TNMG VA2TNN7 VA2TNNK VA2TNNZ VA2TNP2 VA2TNXK VA2TNXW VA2TP2X VA2TP34 VA2TP4F VA2TP51 VA2TP9U VA2TP9X VA2TPC7 VA2TPD6 VA2TPDQ VA2TQGN VA2TQH1 VA2TQKB VA2TW88 VA2TW95 VA2TW9E VA2TWNB VA2TWPC VA2TWR5 VA2TWRM VA2TWRX VA2TWTB VA2TXZC VA2TY0Q VA2TY1A VA2TY35 VA2TZ0X VA2TZ19 VA2TZ30 VA2TZ3E VA2TZ5E VA2TZ5M VA2U6R1 VA2U6X1 VA2U6XA VA2U6XH VA2U6XP VA2U6XY VA2U72Q VA2U75T VA2U75W VA2U75Z VA2U76B VA2U76P VA2UBPS VA2UBQB VA2UBQF VA2UBQH VA2UBQM VA2UBR2 VA2UBS5 VA2UBSN VA2UBTF VA2UBU9 VA2UBUU VA2UDXS VA2UDZM VA2UE1A VA2UE61 VA2UE8A VA2UGPV VA2UGRH VA2UGUM VA2UGVD VA2UGW7 VA2UGXQ VA2UHF9 VA2UHFH VA2UHJ4 VA2UHJF VA2UK9S VA2UKA0 VA2UKAD VA2UKAQ VA2UKBH VA2UKBX VA2ULXB VA2UM04 VA2UM14 VA2UM1W VA2UM1Z VA2UM2B VA2UM7B VA2UM8E VA2UM9D VA2UM9L VA2UMC3 VA2UW1Q VA2UW6A VA2UW6Z VA2UX32 VA2UX34 VA2UX3H VA2UX43 VA2UX4W VA2UX5G VA2UX5S VA2UX5W VA2UYKL VA2UYKP VA2V1PC VA2V1R7 VA2V1RH VA2V1RL VA2V1S1 VA2V1S5 VA2V1S8 VA2V1T5 VA2V1U5 VA2V1UH VA2V1UM VA2V9SP VA2V9U1 VA2V9V3 VA2V9W5 VA2VA59 VA2VA5A VA2VA5B VA2VBND VA2VBP1 VA2VBPG VA2VBPQ VA2VBQY VA2VBW9 VA2VBWD VA2VBWF VA2VBWW VA2VD3Q VA2VD46 VA2VDBA VA2VE15 VA2VFFP VA2VFHT VA2VFJ1 VA2VFJG VA2VJLK VA2VKFP VA2VKJT VA2VKKP VA2VL7Z VA2VL80 VA2VL83 VA2VL86 VA2VL89 VA2VL8B VA2VNEP VA2VNEY VA2VNEZ VA2VNF5 VA2VNF7 VA2VNFH VA2VRNG VA2VRNZ VA2VRPE VA2VRPT VA2VSW4 VA2VUX0 VA2VUY7 VA2VV0H VA2VV0K VA2VV10 VA2VV2X VA2VWDC VA2VWDG VA2VWER VA2VWEV VA2VWF2 VA2VWF8 VA2VWWE VA2VWWU VA2VZUS VA2VZV6 VA2VZVK VA2W02V VA2W02X VA2W033 VA2W11G VA2W12E VA2W1CC VA2W1CR VA2W43H VA2W443 VA2W44B VA2W455 VA2W459 VA2W45N VA2W6VE VA2WB8K VA2WB9N VA2WG8E VA2WG8H VA2WG9J VA2WG9P VA2WG9R VA2WG9T VA2WPL0 VA2WPLH VA2WPLK VA2WPLM VA2WPLW VA2WPM5 VA2WW4S VA2WW5M VA2WW60 VA2WW7F VA2WW7R VA2WW85 VA2WW8J VA2WW8V VA2WW9C VA2WWAQ VA2WWB2 VA2WWB5 VA2WWBF VA2WWBZ VA2WXRT VA2WXS3 VA2X5WM VA2X60Z VA2X61K VA2X62C VA2X6LH VA2X6R9 VA2X6RX VA2X6TG VA2X6U2 VA2X6UH VA2X6UM VA2X6UP VA2XE76 VA2XE9B VA2XEAS VA2XEBU VA2XEEP VA2XEF1 VA2XFSF VA2XFSH VA2XFSJ VA2XFSL VA2XFSV VA2XK0C VA2XK0Y VA2XK16 VA2XK1C VA2XK1S VA2XK3W VA2XKX5 VA2XLSY VA2XLTJ VA2XLTU VA2XLUE VA2XLVU VA2XLW8 VA2XLX7 VA2XLXJ VA2Y0BT VA2Y0BZ VA2Y0C7 VA2Y0C9 VA2Y0CA VA2Y0CE VA2Y0CJ VA2Y0CM VA2Y1DV VA2Y1E0 VA2Y1EY VA2Y1F9 VA2Y1H9 VA2Y1HZ VA2Y1J3 VA2Y1JA VA2Y1JL VA2Y1JN VA2Y1SY VA2Y1WH VA2Y5KP VA2Y5KT VA2Y5LD VA2Y5LH VA2Y5LN VA2Y5N2 VA2Y5Q7 VA2Y5QM VA2YAMK VA2YAP1 VA2YAXC VA2YB18 VA2YB3B VA2YB3W VA2YB3X VA2YB4C VA2YBEJ VA2YBEN VA2YBEU VA2YBEW VA2YBEZ VA2YBF1 VA2YBF2 VA2YBF3 VA2YKDK VA2YKE7 VA2YKEQ VA2YKF3 VA2YKF9 VA2YKFR VA2YKG3 VA2YKGE VA2YKJB VA2YKM9 VA2YLDT VA2YLGN VA2YMTF VA2YMTJ VA2YN6E VA2YN6Y VA2YPJS VA2YPMD VA2YPSP VA2YPSR VA2YRTW VA2YRU3 VA2YRUC VA2YRUQ VA2YRUV VA2YRUW VA2YSUF VA2YSUR VA2YSVR VA2YSVZ VA2YSW3 VA2YVK8 VA2YVM2 VA2YVND VA2YVQ5 VA2YVQX VA2YVR1 VA2YVR9 VA2YVRF VA2YVRG VA2YVRK VA2YVS5 VA2YXJ2 VA2YXJA VA2YXJD VA2YXRP VA2YXVH VA2YYZB VA2YZ02 VA2YZ4R VA2YZ4S VA2YZ56 VA2YZ5B VA2YZ5J VA2YZ5X VA2Z4UT VA2Z4V1 VA2Z4V9 VA2Z4VB VA2Z4VC VA2Z4VN VA2Z8GS VA2Z8JH VA2Z8MH VA2Z8PD VA2Z8RS VA2ZA6C VA2ZA71 VA2ZA9C VA2ZA9H VA2ZA9N VA2ZAA5 VA2ZAAG VA2ZL9C VA2ZL9G VA2ZL9H VA2ZL9U VA2ZLA0 VA2ZLDV VA2ZLG0 VA2ZLG9 VA2ZMM5 VA2ZMR1 VA2ZN4Z VA2ZN50 VA2ZN51 VA2ZN52 VA2ZT9J VA2ZT9P VA2ZT9S VA2ZT9X VA2ZT9Y VA2ZY12 VA2ZY13 VA2ZY9S VA2ZYB5 VA2ZYBD VA2ZYBV VA2ZYFJ VA2ZYFM VA2ZYFQ VA2ZYFS VA3017J VA3017M VA3017N VA3017Q VA30195 VA30196 VA308JF VA308K3 VA308K7 VA308KK VA309BH VA309BS VA309BX VA309C7 VA30BG3 VA30BLW VA30CH5 VA30CHZ VA30CKQ VA30CP8 VA30CQ6 VA30CQB VA30DD4 VA30DDB VA30E3Q VA30E3Y VA30ERW VA30ESW VA30ETN VA30EUC VA30EXH VA30EYP VA30LQR VA30LRU VA30LS3 VA30LS6 VA30LSJ VA30LT7 VA30MS5 VA30MS7 VA30MSH VA30MUA VA30P8H VA30PAN VA30PDY VA30PEZ VA30PL2 VA30PNL VA30PP3 VA30WRQ VA30WRR VA30WRS VA30WRT VA30WRU VA30WRV VA30ZMP VA30ZMU VA30ZNG VA30ZNX VA30ZP6 VA30ZPX VA30ZSN VA30ZT4 VA311DM VA311DX VA3185T VA3185U VA3189M VA3189N VA318KK VA318KZ VA318L9 VA318LL VA318LT VA318M1 VA31JVE VA31JVK VA31K9P VA31K9Q VA31K9R VA31K9S VA31K9T VA31K9U VA31KG9 VA31KGA VA31KGB VA31LJZ VA31LK5 VA31LK7 VA31LN5 VA31LNK VA31LNQ VA31LNS VA31LNU VA31LNV VA31LP1 VA31LP4 VA31LP9 VA31MZ9 VA31MZB VA31MZC VA31MZE VA31MZG VA31NLS VA31SVZ VA31SW0 VA31SW3 VA31SWZ VA31TFR VA31THX VA31VE9 VA31VEA VA31VEB VA31VEC VA31VED VA31VUC VA31VZ9 VA31WX7 VA31WXM VA321PS VA321PT VA321PU VA321QA VA321QB VA3226N VA3226P VA322AL VA322AM VA322DC VA322DD VA326VL VA326VS VA326W8 VA326WB VA326WQ VA326WT VA326X7 VA326XB VA326XE VA326XZ VA327G5 VA328BZ VA328C0 VA328Z1 VA328Z8 VA32B1R VA32B1U VA32FZG VA32FZH VA32FZJ VA32FZL VA32FZM VA32FZP VA32FZW VA32G0Q VA32G0T VA32G15 VA32GXM VA32H8N VA32H8P VA32H94 VA32H96 VA32LKD VA32LPQ VA32LQB VA32LQU VA32LRR VA32LRW VA32NFC VA32NFD VA32NFE VA32NFF VA32NHJ VA32NHK VA32Q6J VA32Q6K VA32QGR VA32QGV VA32QGX VA32QH2 VA32T05 VA32T09 VA32T3V VA32T41 VA32T4G VA331AR VA331BF VA331C1 VA331C3 VA331DW VA331EG VA331Y0 VA332XV VA334AP VA334AQ VA334P8 VA334PF VA33508 VA33509 VA336RZ VA337BC VA337BE VA337CV VA337CW VA3399N VA33BMK VA33BTS VA33BUA VA33BVX VA33BWN VA33BY2 VA33C1W VA33C21 VA33DH5 VA33DHN VA33DTG VA33DTM VA33FKU VA33FLC VA33FLG VA33K80 VA33L1L VA33L1M VA33L1Y VA33L1Z VA33L4F VA33L4L VA33NUA VA33NUP VA33NUZ VA33Y3N VA33Y49 VA33Y4P VA33Y5L VA33Y5N VA33Y5S VA344NZ VA344P2 VA344P5 VA344PJ VA344PN VA344Q1 VA345KX VA345KZ VA3477Q VA34789 VA347AZ VA347C5 VA348X8 VA348Y0 VA34D3A VA34D3N VA34D3V VA34D3X VA34D3Z VA34D42 VA34GR2 VA34GRV VA34GS8 VA34GSJ VA34GSK VA34GSP VA34K4L VA34K4S VA34K61 VA34K6D VA34NA5 VA34NA8 VA34NBW VA34NBX VA34NC0 VA34NC4 VA34P48 VA34P5S VA34P68 VA34P76 VA34P79 VA34P7C VA34QNU VA34QP7 VA34QXR VA34QXY VA34SHX VA34SHZ VA34SJ4 VA34SJ5 VA34U54 VA34U5W VA34U6M VA34U9H VA34UBK VA34UXR VA34XYF VA34XYM VA34XZJ VA34Y1C VA34Y1W VA350MC VA350MH VA350N6 VA350N9 VA350NB VA350ND VA352YK VA352ZN VA3530X VA35311 VA35314 VA3531D VA354QE VA354QV VA354RE VA354RU VA35503 VA35509 VA356JP VA356M3 VA356MR VA356UJ VA356VE VA3579A VA359BS VA359C6 VA359NX VA359NZ VA359P1 VA359PB VA35ASM VA35ATX VA35AWC VA35AYF VA35EQT VA35EQW VA35ERB VA35ES0 VA35JLR VA35JLY VA35JML VA35JND VA35JNH VA35JNK VA35MFG VA35MFH VA35MFJ VA35MFK VA35MFN VA35NXF VA35NZ1 VA35NZZ VA35P0D VA35P0H VA35P0V VA35P0X VA35S6A VA35S6H VA35S6L VA35S6Q VA35S6R VA35TUR VA35TUS VA35TUT VA35TUV VA35V8B VA35V8D VA35V8Y VA35V90 VA35V97 VA35V9C VA35VY2 VA35VY8 VA35VYH VA35VYP VA35VYQ VA35VYV VA35Y35 VA35Y36 VA35Y3S VA35Y3T VA360DV VA360DW VA360E0 VA360EA VA360ED VA360FP VA362PM VA362PN VA362RN VA362RP VA362UQ VA362V1 VA366MH VA366ML VA366S3 VA366S5 VA366SN VA366TC VA368H6 VA368H9 VA36C9C VA36C9D VA36C9V VA36C9W VA36CA0 VA36CA1 VA36FD0 VA36FD3 VA36FDA VA36FDE VA36FDK VA36FE1 VA36JDW VA36JE4 VA36JMT VA36JNA VA36JND VA36JNQ VA36K3J VA36K3K VA36K9D VA36KGU VA36KH0 VA36KH1 VA36N6M VA36NEB VA36NEC VA36NQN VA36QMX VA36QMY VA36QPG VA36QQ9 VA36QQZ VA36QRV VA36QRX VA36V2Y VA36V5S VA36V5T VA36V5U VA36XQG VA36XQK VA36XR8 VA36XRF VA36XRT VA36XS0 VA36Z5A VA36Z5D VA36Z5H VA36ZBC VA36ZBD VA36ZBR VA3705Q VA3705X VA37066 VA3706W VA3707Y VA3708B VA376SR VA376SS VA376T5 VA376T6 VA376TA VA376TC VA37A02 VA37A07 VA37A0S VA37A18 VA37F22 VA37F26 VA37F27 VA37F28 VA37F29 VA37F2A VA37KD3 VA37KGK VA37KGN VA37KGU VA37KGY VA37KHA VA37NUS VA37NUU VA37NUX VA37NVS VA37NVT VA37NW4 VA37NWA VA37P3D VA37P3E VA37P5G VA37P5H VA37PAL VA37PAT VA37TVC VA37TVE VA37TVW VA37TW7 VA37TWA VA37TWB VA37TWC VA37U3K VA37U3Q VA37U3R VA37U3S VA37UAS VA37UAT VA37Z7R VA37Z8Q VA37Z8T VA37Z90 VA37ZAH VA37ZAQ VA383AS VA383BP VA383BY VA383BZ VA3847B VA38489 VA3848S VA3848T VA3848U VA3848V VA387YX VA387Z0 VA387Z4 VA387Z9 VA38DJE VA38DJG VA38DJJ VA38DJN VA38DJQ VA38DJR VA38G2M VA38G32 VA38G3F VA38G3M VA38GHW VA38GHX VA38GJY VA38GJZ VA38N46 VA38N4B VA38N4P VA38N5Z VA38PGW VA38PGY VA38UDS VA38UDV VA38UDX VA38UDZ VA38UE1 VA38Y7V VA38Y7W VA38YD7 VA38YKM VA38YL2 VA38YL3 VA3913K VA3913M VA3913P VA39141 VA394UN VA394VH VA394W1 VA3956H VA3956N VA3956U VA3963Q VA3963S VA3963Y VA39646 VA39AG5 VA39AHY VA39AJ0 VA39B7T VA39B80 VA39B88 VA39B8J VA39B8P VA39B8X VA39G3N VA39G3P VA39G3R VA39G5D VA39G5H VA39G5J VA39GXQ VA39GZE VA39H00 VA39H02 VA39L2K VA39L2L VA39L2X VA39L31 VA39L38 VA39L3B VA39MJA VA39MJK VA39MJM VA39MK3 VA39MK6 VA39P98 VA39P9D VA39SLZ VA39SM4 VA39SMG VA39SMQ VA39WQ4 VA39WQ5 VA39X01 VA39X02 VA39X0R VA39X0T VA39YMF VA39YMH VA39YTM VA39YTN VA39YTP VA39YTQ VA3A060 VA3A061 VA3A062 VA3A063 VA3A064 VA3A065 VA3A08C VA3A08D VA3A08E VA3A08F VA3A08Z VA3A09A VA3A227 VA3A22F VA3A22K VA3A22Q VA3A22U VA3A3MR VA3A3MU VA3A3MY VA3A3MZ VA3A3N0 VA3A3N1 VA3A428 VA3A439 VA3A495 VA3A497 VA3A49H VA3A49L VA3A4A0 VA3A4A6 VA3A503 VA3A509 VA3A50G VA3A50K VA3A50N VA3A50Q VA3A5E8 VA3A5J2 VA3A5NX VA3A5NZ VA3A5P1 VA3A5P3 VA3A5P7 VA3A5QT VA3A69Z VA3A6A1 VA3A6A9 VA3A6AA VA3A6AE VA3A6AJ VA3A7WW VA3A817 VA3A81M VA3A825 VA3A83E VA3A84G VA3A9RP VA3A9RU VA3A9RZ VA3A9SQ VA3A9SV VA3A9SZ VA3ACSP VA3ACSV VA3ACT5 VA3ACZ3 VA3AD0J VA3AD39 VA3AEK6 VA3AEKP VA3AEL5 VA3AENB VA3AENG VA3AENT VA3AGRH VA3AGRK VA3AGT1 VA3AGY5 VA3AGYF VA3AGZC VA3AJ6B VA3AJ6L VA3AJ96 VA3AJ97 VA3ALEY VA3ALEZ VA3ALMH VA3ALMJ VA3ALNS VA3ALNU VA3ANFL VA3ANKD VA3ANKZ VA3AQ01 VA3AQ0B VA3AQ0C VA3ARMG VA3ARMM VA3ARMW VA3AT4L VA3AT56 VA3AUJA VA3AUJB VA3AYPR VA3AYQU VA3AYRD VA3B09A VA3B3DC VA3B4ZL VA3B5UX VA3B6G6 VA3B6G9 VA3B93P VA3BBAA VA3BCWB VA3BCXJ VA3BG5R VA3BKX1 VA3BL5P UDI-DI 00763000587062 Lots VA2JLZT VA2JMYG VA2JQGQ VA2NCB1 VA2NN7R VA2NN95 VA2NN9V VA2P2N4 VA2P9VM VA2PA8W VA2PCP0 VA2S1KD VA2S1KF VA2S1V2 VA2SCUY VA30ZT4 VA31KGB VA31LNS VA31LNV VA31LP1 VA31LP4 VA31MZB VA31SVZ VA31VE9 VA321PU VA322AM VA322DD VA34D3Z VA34K6D VA34P48 VA34U6M VA34U9H VA350N9 VA35509 VA356M3 VA35JLY VA35JML VA36Z5H VA376T5 VA37KGK VA37P5G VA37TVW VA383BP VA3848V VA38GHX VA38N4P VA38YL3 VA3A22U VA3A3MU; 3560030 UDI-DI 00763000203887 Lots VA24HE0 VA24Q0F VA24QED VA24QEF VA24WJD VA25181 VA25182 VA26CMD VA26CMZ VA26CN0 VA26CN1 VA26CN3 VA26E7A VA26EGF VA26EGG VA275QF VA275QG VA2760G VA276AW VA276AX VA276GM VA276GN VA276GP VA277MV VA277PR VA277V8 VA277V9 VA277VU VA277VV VA277VW VA27CCU VA27CGB VA27CGC VA27CGD VA27D43 VA27D44 VA27D5W VA27UGH VA27UW3 VA27YHA VA27YKP VA280D2 VA280MV VA28488 VA2871D VA2872X VA28BHF VA28BHG VA28BLL VA28BQ1 VA28BR1 VA28BR2 VA28K4U VA28Z67 VA28Z6B VA28Z6N VA28Z6P VA28Z6X VA28Z9M VA29ES2 VA29ES3 VA29ES4 VA29FCW VA29MH7 VA29MKB VA29MTM VA29N1U VA29ZUL VA2A2Z8 VA2A2Z9 VA2A2ZA VA2AEYH VA2ALWG VA2APTS VA2ARCX VA2B686 VA2B687 VA2B6NC VA2B8H1 VA2B8HP VA2BCYM VA2BD3Z VA2BG7Z VA2BG83 VA2BG87 VA2BG8X VA2BG9P VA2BLFF VA2BLFH VA2BLFJ VA2BLFK VA2BM7Q VA2BT60 VA2BU2P VA2BUMF VA2C5LL VA2C7N9 VA2CC3M VA2CCF6 VA2CCJT VA2CSQV VA2CSQW VA2CSQX VA2CSQY VA2CT5C VA2CUNA VA2CZT5 VA2CZT6 VA2D8TH VA2DDM4 VA2DV4U VA2DV70 VA2DV94 VA2E1Q6 VA2E1QA VA2E4AW VA2E4D1 VA2EGXS VA2EGXT VA2EGZB VA2EJF6 VA2EMNG VA2EMPB VA2EP63 VA2EP76 VA2F0A5 VA2F0MJ VA2F0PS VA2F0R7 VA2F0S7 VA2F32V VA2F349 VA2F34L VA2F37J VA2FBWW VA2FBXW VA2FBYC VA2FBYK VA2FBYS VA2FJ04 VA2FJ0E VA2FJ3T VA2FJDG VA2FJDK VA2FJDP VA2FJU0 VA2FL2Z VA2FL90 VA2FL99 VA2FLA7 VA2FLA9 VA2FLAG VA2FLPL VA2FLSF VA2FLSN VA2FLTG VA2FLV1 VA2FLVY VA2FM0B VA2FMQY VA2FMRW VA2FMSW VA2FMTT VA2FP7E VA2FP7G VA2FP83 VA2FPAV VA2GKGD VA2GT10 VA2HET9 VA2HEUH VA2HEUM VA2HEUW VA2HT55 VA2HT5F VA2J72L VA2J8ZF VA2JLWF VA2JLX7 VA2JLXB VA2JLXE VA2JTN9 VA2KVRP VA2L1XV VA2L1Y3 VA2L1Y7 VA2L1YD VA2L7LK VA2L7M1 VA2LGZK VA2LH09 VA2LH1X VA2LH1Z VA2LWMF VA2LWN1 VA2M0HN VA2M0K2 VA2M7EH VA2M7EP VA2M7F4 VA2M7F8 VA2PD3B VA2PD41 VA2PPFK VA2PPGP VA2QESR VA2QET4 VA2QETF VA2QEUE VA2QYGH VA2QYJT VA2QYK0 VA2RAD2 VA2RP80 VA2RPBR VA2T1BR VA2T1HA VA2TSXS VA2TSYM VA2TSYR VA2TSZ3 VA2TSZB VA2TSZF VA2UW75 VA2UW84 VA2VKHF VA2VKJM VA2W6JR VA2WPMQ VA2WPMS VA2YBF4 VA2YVJA VA2Z4UG VA2ZMK3 VA2ZMKC VA30BN3 VA30BNV VA30PG1 VA30PGX VA31985 VA32NHL VA32NHN VA331Y9 VA332Y2 VA332Y7 VA332YQ VA334P0 VA34Y1Z VA35MC0 VA35Y60 VA35Y63 VA37A1N VA3839X VA383A3 VA38UE5 VA3913T VA3ARMA UDI-DI 00763000587000 Lots VA2F0R7 VA2GT10 VA2LH1Z VA2M7F8 VA2QET4 VA2QEUE VA2T1HA VA2TSXS VA2TSYM VA2VKJM VA2WPMQ VA2WPMS VA2YVJA VA2Z4UG VA334P0
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- InterStim™ 3560022Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1 recall8 reports
- InterStim™ 3560030Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1 recall3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99949).
- Z-0266-2027 · Open, ClassifiedInterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTM MICRO 33CM SSMRI TINED 978A133 LEAD 978A141 ISTM MICRO 41CM SSMR…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗