InterStim™ 3560022Recalls & reported problems
Medtronic, Inc. · EXT 3560022 ISTM QUAD PERC 4.32MM · Class 3
All models of INTERSTIM (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000203894), received between Aug 20, 2020 and Nov 19, 2025 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
7 other recalls cover devices cleared under the same FDA submission (P080025, P970004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 8, 2026Class IIOpen, Classified
Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.
Z-0267-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Break2
- Component Missing2
- Connection Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hypersensitivity/Allergic reaction1
- Insufficient Information1
- Local Reaction1
- No Known Impact Or Consequence To Patient1
- Rash1
- Swelling/ Edema1
- Unspecified Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryNov 19, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT EXTENSION FAILURE. THE EXTENSION DID NOT CAUSE ANY SENSATION OF IRRITATION. IT WAS REPLACED. ISSUE RESOLVED. WILL BE RETURNED.
Patient problems: Insufficient Information
MAUDE report 23595553 ↗MalfunctionOct 31, 2025Connection Problem; Connection Problem
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS IMPLAN TED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT DURING PROCEDURE, THE LEAD DID NOT GO INTO THE PERCUTANEOUS EXTENSION CABLE. IT TURNED OUT THAT THE SCREW WAS INSIDE AND BLOCKING LEAD ENTRANCE. CAUSE NOT SPECIFIED. SCREW WAS TAKEN OUT AND IT WAS SOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23437410 ↗MalfunctionDec 10, 2024Component Missing; Insufficient Device Problem Information; Component Missing; Insufficient Device Problem Information
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT THERE WAS A DEFECT WITH THE PERCUTANEOUS EXTENSION. IT WAS NOTED THAT THE PERCUTANEOUS EXTENSION CONNECTOR DID NOT HAVE A SET SCREW. THE CAUSE WAS NOT KNOWN. A DIFFERENT PERCUTANEOUS EXTENSION WAS OPENED AND USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20892450 ↗InjuryJul 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT THE TRIAL PATIENT STARTED THEIR TRIAL ON (B)(6) 2024. THE PATIENT HAS VISITED THEIR ABDOMINAL SURGEON FOR REGULAR VISIT FOR CHECKING THE WOUNDS AND THE WOUND MANAGEMENT. AFTER CHECKING THE PATIENT'S SKIN THERE WAS URTICARIA AND INITIAL SYSTEMIC URTICARIA, AND EDEMA ON PARTS OF THE SKIN WHERE THE EXTENSION WAS IN DIRECT CONTACT. THE DOCTOR ASSUME IT COULD BE AN ALLERGIC REACTION TO THE PERCUTANEOUS EXTENSION MATERIALS. THEY DID SKIN OBSERVATION, ANTI-ANAPHYLACTIC MEASURES, AND PATIENT HAD BEEN SENT TO OR TO IMMEDIATE REMOVAL/EXPLANTATION OF THE EXTENSION AND WAS TREATED ANTI-ANAPHYLACTIC. FURTHER DERMATOLOGY EXAMINATIONS ARE SCHEDULED ALLERGY TESTS TO SPECIFIC MATERIALS ARE SCHEDULED. THE PATIENT IS HEMODYNAMIC STABILE ALL THE TIME. THEY ARE AWAITING TO SEE IF THE PATIENT IS GOING TO BE SUITABLE FOR PERMANENT IMPLANTATION OF THE PERMANENT IMPLANTABLE NEUROSTIMULATOR (INS). THE PATIENT IS OK AND WILL BE HOSPITALIZED UNTIL NEXT WEEK DUE TO URTICARIA SYMPTOM
Patient problems: Hypersensitivity/Allergic reaction; Local Reaction; Rash; Swelling/ Edema; Unspecified Infection; Urticaria
MAUDE report 19646174 ↗MalfunctionMay 7, 2024Break; Break
INFORMATION WAS RECEIVED FROM AN UNKNOWN SOURCE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT DURING TRIAL CONVERSION, THE BARREL CONNECTOR ON THE PERCUTANEOUS EXTENSION WAS CUT FROM THE EXTENSION SO THAT THE EXTERNAL PORTION COULD BE REMOVED BY NON-SCRUB HOSPITAL PERSONNEL. AFTER DISCARDING THE BARREL CONNECTOR, THE SURGEON FOUND A SMALL PIECE OF PLASTIC TUBING IN THE POCKET WOUND, PRESUMABLY FROM THE PERCUTANEOUS EXTENSION. SURGEON RETRIEVED SMALL PIECE OF PLASTIC TUBING FROM THE WOUND AND DISCARDED IT. THE SURGEON IS CONCERNED THAT IF A SMALL PIECE OF PLASTIC IS INADVERTENTLY LEFT IN THE POCKET WOUND THEN THE PATIENT WILL BE AT A HIGHER RISK OF INFECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19261130 ↗InjuryMar 8, 2024Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT REGARDING EXTENSION LOT# VA2W455, ACCORDING TO THE DOCTOR, THE LEADS ON THE INSULATION ARE DEFECTIVE AND THEREFORE NO CONNECTION COULD BE ESTABLISHED TO THE ELECTRODES. A REVISION OPERATION IS SCHEDULED FOR (B)(6). MORE INFORMATION WILL FOLLOW AFTER THE REVISION SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18870772 ↗MalfunctionAug 30, 2023Connection Problem; Insufficient Device Problem Information; Connection Problem; Insufficient Device Problem Information
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED AN EXTENSION MALFUNCTION AND THAT THE LEAD WOULD NOT PASS INTO THE CONNECTOR OF THE EXTENSION EVEN AFTER BACKING THE SCREW OUT AND MULTIPLE ATTEMPTS. THE EXTENSION WAS REPLACED AND WILL BE RETURNED. ISSUE RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17653355 ↗MalfunctionAug 20, 2020Component Missing; Component Missing
ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER'S REPRESENTATIVE: PATIENT WEIGHT 318 POUNDS (PREVIOUSLY REPORTED AS 308 PO UNDS). ***MDR DECISION CORRECTED TO NOT REPORTABLE. NO ADDITIONAL SUPPLEMENTS REQUIRED UNLESS ADDITIONAL INFORMATION RECEIVED INDICATES REPORTABLE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10432982 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P080025 on Mar 14, 2011. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Sacral Nerve Stimulation Therapy System”. The FDA review took 899 days. The FDA has since approved 350 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
Approved through premarket approval (PMA) P970004 on Sep 29, 1997. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Therapy System For Urinary Control”. The FDA review took 242 days. The FDA has since approved 449 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
The FDA regulates product code EZW (stimulator, electrical, implantable, for incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under P080025, P970004. One submission often covers many models, so check the product and lot details in each record.
- Z-1704-2025 · Apr 3, 2025 · Open, ClassifiedEnhanced Verify Evaluation Handset (CFN HH90130FA)
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-2365-2023 · Jul 6, 2023 · Open, ClassifiedMedtronic InterStim Ground Pad REF 041826
- Z-0998-2022 · Feb 28, 2022 · TerminatedMEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-0040-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim
- Z-0037-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta
Common questions about the InterStim™ 3560022
Has the InterStim™ 3560022 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0267-2027 (Class II), began on Sep 8, 2026: Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the InterStim™ 3560022?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Aug 20, 2020 and Nov 19, 2025. In those reports, the most common specific device problems are “Break” (2), “Component Missing” (2) and “Connection Problem” (2); the most common patient problems are “Hypersensitivity/Allergic reaction” (1), “Local Reaction” (1) and “Rash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the InterStim™ 3560022?
None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the InterStim™ 3560022 FDA approved?
It was approved through premarket approval (PMA) P970004 on Sep 29, 1997; its record lists 2 FDA submissions in total.
Who makes the InterStim™ 3560022, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand InterStim™, model 3560022). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stimulator, electrical, implantable, for incontinence devices
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence6 deaths23,076 reports
- InterStim™ Micro SureScan™ 97810Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence2,762 reports
- InterStim™ X 97800Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1,840 reports
- Axonics 4101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death818 reports
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence818 reports
- Axonics 1101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death399 reports
- Axonics 5101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence162 reports
- INTERSTIM® 3023Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1 death147 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.