Azurion recallZ-0297-2027 · Philips Medical Systems Nederland B.V. · initiated Oct 6, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
FDA root cause category: Equipment maintenance
What the company is doing
Philips Medical notified consignees on about 10/06/2026 via certified mailed letter titled "URGENT Medical Device Correction." Consignees were instructed that affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the additional instructions provided, which include: if motorized movements become unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2), and rotational (3) stand movements remain available. Note: Transversal stand movements (2) are only available for ceiling-mounted systems with the FlexMove option. For floor-mounted systems, only rotational (3) stand movements remain available. The table longitudinal position (4) can be adjusted manually when the table is not tilted. Additionally, consignees were instructed to ensure the notification letter is distributed to all relevant users, maintained with the device documentation until corrective actions are completed, to post the letter near affected devices for ease of reference, notify customers/organizations if affected systems have been distributed or transferred, and complete and return the response form provided. Philips will inspect all systems in scope and replace SSDs that are not specified in the Medical Device File. Philips expects to start the implementation of this correction in December 2026. A local Philips representative will contact consignees to schedule a visit for a Field Service Engineer to perform these activities.
Recall details
- Recall number
- Z-0297-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 6, 2026
- Posted by the FDA
- Oct 27, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 457 units
- Product code
- OWB
- FDA submissions
- K172822, K181830, K200917
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Product description
Philips Azurion 7 B20. Model Numbers: 722068, 722226.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
1. Model Number: 722068. UDI Numbers: (01)00884838085367(21)178, (01)00884838085367(21)147, (01)00884838085367(21)205, (01)00884838085367(21)170, (01)00884838085367(21)494, (01)00884838085367(21)247, (01)00884838085367(21)149, (01)00884838085367(21)153, (01)00884838085367(21)85, (01)00884838085367(21)320, (01)00884838085367(21)89, (01)00884838085367(21)172, (01)00884838085367(21)505, (01)00884838085367(21)271, (01)00884838085367(21)443, (01)00884838085367(21)354, (01)00884838085367(21)196, (01)00884838085367(21)73, (01)00884838085367(21)285, (01)00884838085367(21)406, (01)00884838085367(21)528, (01)00884838085367(21)350, (01)00884838085367(21)462, (01)00884838085367(21)319, (01)00884838085367(21)520, (01)00884838085367(21)117, (01)00884838085367(21)143, (01)00884838085367(21)84, (01)00884838085367(21)469, (01)00884838085367(21)361, (01)00884838085367(21)114, (01)00884838085367(21)477, (01)00884838085367(21)232, (01)00884838085367(21)107, (01)00884838085367(21)282, (01)00884838085367(21)276, (01)00884838085367(21)608, (01)00884838085367(21)473, (01)00884838085367(21)417, (01)00884838085367(21)302, (01)00884838085367(21)88, (01)00884838085367(21)126, (01)00884838085367(21)211, (01)00884838085367(21)468, (01)00884838085367(21)410, (01)00884838085367(21)466, (01)00884838085367(21)495, (01)00884838085367(21)167, (01)00884838085367(21)537, (01)00884838085367(21)173, (01)00884838085367(21)289, (01)00884838085367(21)233, (01)00884838085367(21)536, (01)00884838085367(21)204, (01)00884838085367(21)313, (01)00884838085367(21)237, (01)00884838085367(21)133, (01)00884838085367(21)481, (01)00884838085367(21)307, (01)00884838085367(21)577, (01)00884838085367(21)210, (01)00884838085367(21)258, (01)00884838085367(21)328, (01)00884838085367(21)199, (01)00884838085367(21)480, (01)00884838085367(21)72, (01)00884838085367(21)491, (01)00884838085367(21)423, (01)00884838085367(21)509, (01)00884838085367(21)246, (01)00884838085367(21)614, (01)00884838085367(21)539, (01)00884838085367(21)257, (01)00884838085367(21)369, (01)00884838085367(21)508, (01)00884838085367(21)97, (01)00884838085367(21)303, (01)00884838085367(21)48, (01)00884838085367(21)44, (01)00884838085367(21)550, (01)00884838085367(21)460, (01)00884838085367(21)551, (01)00884838085367(21)478, (01)00884838085367(21)522, (01)00884838085367(21)238, (01)00884838085367(21)544, (01)00884838085367(21)230, (01)00884838085367(21)434, (01)00884838085367(21)93, (01)00884838085367(21)306, (01)00884838085367(21)192, (01)00884838085367(21)215, (01)00884838085367(21)447, (01)00884838085367(21)316, (01)00884838085367(21)552, (01)00884838085367(21)570, (01)00884838085367(21)387, (01)00884838085367(21)134, (01)00884838085367(21)560, (01)00884838085367(21)158, (01)00884838085367(21)518, (01)00884838085367(21)318, (01)00884838085367(21)250, (01)00884838085367(21)118, (01)00884838085367(21)185, (01)00884838085367(21)206, (01)00884838085367(21)489, (01)00884838085367(21)521, (01)00884838085367(21)254, (01)00884838085367(21)534, (01)00884838085367(21)292, (01)00884838085367(21)497, (01)00884838085367(21)293, (01)00884838085367(21)359, (01)00884838085367(21)527, (01)00884838085367(21)129, (01)00884838085367(21)128, (01)00884838085367(21)396, (01)00884838085367(21)334, (01)00884838085367(21)425, (01)00884838085367(21)415, (01)00884838085367(21)471, (01)00884838085367(21)540, (01)00884838085367(21)584, (01)00884838085367(21)553, (01)00884838085367(21)602, (01)00884838085367(21)240, (01)00884838085367(21)617, (01)00884838085367(21)604, (01)00884838085367(21)589, (01)00884838085367(21)612, (01)00884838085367(21)556, (01)00884838085367(21)557, (01)00884838085367(21)574, (01)00884838085367(21)603, (01)00884838085367(21)601, (01)00884838085367(21)609, (01)00884838085367(21)616, (01)00884838085367(21)597, (01)00884838085367(21)605, (01)00884838085367(21)575, (01)00884838085367(21)516, (01)00884838085367(21)624, (01)00884838085367(21)563, (01)00884838085367(21)590, (01)00884838085367(21)548, (01)00884838085367(21)420, (01)00884838085367(21)475, (01)00884838085367(21)180, (01)00884838085367(21)414, (01)00884838085367(21)349, (01)00884838085367(21)519, (01)00884838085367(21)234, (01)00884838085367(21)219, (01)00884838085367(21)195, (01)00884838085367(21)533, (01)00884838085367(21)398, (01)00884838085367(21)618, (01)00884838085367(21)549, (01)00884838085367(21)46, (01)00884838085367(21)531, (01)00884838085367(21)426, (01)00884838085367(21)290, (01)00884838085367(21)49, (01)00884838085367(21)326, (01)00884838085367(21)286, (01)00884838085367(21)124, (01)00884838085367(21)362, (01)00884838085367(21)317, (01)00884838085367(21)571, (01)00884838085367(21)331, (01)00884838085367(21)273, (01)00884838085367(21)188, (01)00884838085367(21)264, (01)00884838085367(21)122, (01)00884838085367(21)300, (01)00884838085367(21)100, (01)00884838085367(21)253, (01)00884838085367(21)391, (01)00884838085367(21)226, (01)00884838085367(21)315, (01)00884838085367(21)63, (01)00884838085367(21)252, (01)00884838085367(21)175, (01)00884838085367(21)186, (01)00884838085367(21)409, (01)00884838085367(21)298, (01)00884838085367(21)368. Serial Numbers: 178, 147, 205, 170, 494, 247, 149, 153, 85, 320, 89, 172, 505, 271, 443, 354, 196, 73, 285, 406, 528, 350, 462, 319, 520, 117, 143, 84, 469, 361, 114, 477, 232, 107, 282, 276, 608, 473, 417, 302, 88, 126, 211, 468, 410, 466, 495, 167, 537, 173, 289, 233, 536, 204, 313, 237, 133, 481, 307, 577, 210, 258, 328, 199, 480, 72, 491, 423, 509, 246, 614, 539, 257, 369, 508, 97, 303, 48, 44, 550, 460, 551, 478, 522, 238, 544, 230, 434, 93, 306, 192, 215, 447, 316, 552, 570, 387, 134, 560, 158, 518, 318, 250, 118, 185, 206, 489, 521, 254, 534, 292, 497, 293, 359, 527, 129, 128, 396, 334, 425, 415, 471, 540, 584, 553, 602, 240, 617, 604, 589, 612, 556, 557, 574, 603, 601, 609, 616, 597, 605, 575, 516, 624, 563, 590, 548, 420, 475, 180, 414, 349, 519, 234, 219, 195, 533, 398, 618, 549, 46, 531, 426, 290, 49, 326, 286, 124, 362, 317, 571, 331, 273, 188, 264, 122, 300, 100, 253, 391, 226, 315, 63, 252, 175, 186, 409, 298, 368. 2. Model Number: 722226. UDI Numbers: (01)00884838099272(21)209, (01)00884838099272(21)675, (01)00884838099272(21)1091, (01)00884838099272(21)113, (01)00884838099272(21)1145, (01)00884838099272(21)1078, (01)00884838099272(21)197, (01)00884838099272(21)687, (01)00884838099272(21)680, (01)00884838099272(21)280, (01)00884838099272(21)343, (01)00884838099272(21)98, (01)00884838099272(21)332, (01)00884838099272(21)341, (01)00884838099272(21)1162, (01)00884838099272(21)554, (01)00884838099272(21)337, (01)00884838099272(21)400, (01)00884838099272(21)757, (01)00884838099272(21)326, (01)00884838099272(21)323, (01)00884838099272(21)262, (01)00884838099272(21)1017, (01)00884838099272(21)379, (01)00884838099272(21)786, (01)00884838099272(21)462, (01)00884838099272(21)573, (01)00884838099272(21)458, (01)00884838099272(21)987, (01)00884838099272(21)880, (01)00884838099272(21)141, (01)00884838099272(21)693, (01)00884838099272(21)510, (01)00884838099272(21)882, (01)00884838099272(21)1157, (01)00884838099272(21)21, (01)00884838099272(21)1050, (01)00884838099272(21)367, (01)00884838099272(21)184, (01)00884838099272(21)178, (01)00884838099272(21)179, (01)00884838099272(21)250, (01)00884838099272(21)1139, (01)00884838099272(21)1062, (01)00884838099272(21)575, (01)00884838099272(21)1121, (01)00884838099272(21)950, (01)00884838099272(21)368, (01)00884838099272(21)176, (01)00884838099272(21)712, (01)00884838099272(21)829, (01)00884838099272(21)249, (01)00884838099272(21)576, (01)00884838099272(21)1046, (01)00884838099272(21)708, (01)00884838099272(21)738, (01)00884838099272(21)1083, (01)00884838099272(21)1098, (01)00884838099272(21)717, (01)00884838099272(21)1126, (01)00884838099272(21)665, (01)00884838099272(21)764, (01)00884838099272(21)386, (01)00884838099272(21)977, (01)00884838099272(21)721, (01)00884838099272(21)1006, (01)00884838099272(21)232, (01)00884838099272(21)682, (01)00884838099272(21)789, (01)00884838099272(21)698, (01)00884838099272(21)331, (01)00884838099272(21)1097, (01)00884838099272(21)1094, (01)00884838099272(21)378, (01)00884838099272(21)808, (01)00884838099272(21)718, (01)00884838099272(21)195, (01)00884838099272(21)412, (01)00884838099272(21)715, (01)00884838099272(21)1143, (01)00884838099272(21)775, (01)00884838099272(21)1073, (01)00884838099272(21)982, (01)00884838099272(21)282, (01)00884838099272(21)617, (01)00884838099272(21)958, (01)00884838099272(21)222, (01)00884838099272(21)430, (01)00884838099272(21)1067, (01)00884838099272(21)810, (01)00884838099272(21)1114, (01)00884838099272(21)314, (01)00884838099272(21)762, (01)00884838099272(21)1102, (01)00884838099272(21)359, (01)00884838099272(21)978, (01)00884838099272(21)1060, (01)00884838099272(21)920, (01)00884838099272(21)1136, (01)00884838099272(21)836, (01)00884838099272(21)835, (01)00884838099272(21)795, (01)00884838099272(21)800, (01)00884838099272(21)899, (01)00884838099272(21)838, (01)00884838099272(21)944, (01)00884838099272(21)1123, (01)00884838099272(21)335, (01)00884838099272(21)1039, (01)00884838099272(21)203, (01)00884838099272(21)731, (01)00884838099272(21)725, (01)00884838099272(21)701, (01)00884838099272(21)967, (01)00884838099272(21)501, (01)00884838099272(21)992, (01)00884838099272(21)119, (01)00884838099272(21)732, (01)00884838099272(21)974, (01)00884838099272(21)919, (01)00884838099272(21)605, (01)00884838099272(21)686, (01)00884838099272(21)230, (01)00884838099272(21)523, (01)00884838099272(21)327, (01)00884838099272(21)477, (01)00884838099272(21)1049, (01)00884838099272(21)298, (01)00884838099272(21)993, (01)00884838099272(21)522, (01)00884838099272(21)520, (01)00884838099272(21)521, (01)00884838099272(21)259, (01)00884838099272(21)909, (01)00884838099272(21)142, (01)00884838099272(21)263, (01)00884838099272(21)422, (01)00884838099272(21)17, (01)00884838099272(21)1057, (01)00884838099272(21)36, (01)00884838099272(21)1103, (01)00884838099272(21)164, (01)00884838099272(21)485, (01)00884838099272(21)1155, (01)00884838099272(21)1054, (01)00884838099272(21)366, (01)00884838099272(21)1140, (01)00884838099272(21)1048, (01)00884838099272(21)752, (01)00884838099272(21)1088, (01)00884838099272(21)47, (01)00884838099272(21)544, (01)00884838099272(21)906, (01)00884838099272(21)289, (01)00884838099272(21)548, (01)00884838099272(21)238, (01)00884838099272(21)299, (01)00884838099272(21)284, (01)00884838099272(21)404, (01)00884838099272(21)283, (01)00884838099272(21)401, (01)00884838099272(21)833, (01)00884838099272(21)895, (01)00884838099272(21)233, (01)00884838099272(21)224, (01)00884838099272(21)1087, (01)00884838099272(21)395, (01)00884838099272(21)444, (01)00884838099272(21)710, (01)00884838099272(21)820, (01)00884838099272(21)235, (01)00884838099272(21)846, (01)00884838099272(21)727, (01)00884838099272(21)1026, (01)00884838099272(21)1129, (01)00884838099272(21)777, (01)00884838099272(21)481, (01)00884838099272(21)431, (01)00884838099272(21)696, (01)00884838099272(21)1099, (01)00884838099272(21)438, (01)00884838099272(21)552, (01)00884838099272(21)630, (01)00884838099272(21)941, (01)00884838099272(21)429, (01)00884838099272(21)854, (01)00884838099272(21)888, (01)00884838099272(21)596, (01)00884838099272(21)957, (01)00884838099272(21)645, (01)00884838099272(21)739, (01)00884838099272(21)736, (01)00884838099272(21)1127, (01)00884838099272(21)1117, (01)00884838099272(21)527, (01)00884838099272(21)251, (01)00884838099272(21)1000, (01)00884838099272(21)965, (01)00884838099272(21)553, (01)00884838099272(21)1160, (01)00884838099272(21)1042, (01)00884838099272(21)699, (01)00884838099272(21)1064, (01)00884838099272(21)574, (01)00884838099272(21)741, (01)00884838099272(21)740, (01)00884838099272(21)1137, (01)00884838099272(21)900, (01)00884838099272(21)901, (01)00884838099272(21)887, (01)00884838099272(21)949, (01)00884838099272(21)632, (01)00884838099272(21)1112, (01)00884838099272(21)1156, (01)00884838099272(21)737, (01)00884838099272(21)907, (01)00884838099272(21)825, (01)00884838099272(21)877, (01)00884838099272(21)1134, (01)00884838099272(21)1161, (01)00884838099272(21)648, (01)00884838099272(21)831, (01)00884838099272(21)1116, (01)00884838099272(21)1149, (01)00884838099272(21)1104, (01)00884838099272(21)860, (01)00884838099272(21)798, (01)00884838099272(21)603, (01)00884838099272(21)938, (01)00884838099272(21)773, (01)00884838099272(21)939, (01)00884838099272(21)1065, (01)00884838099272(21)716, (01)00884838099272(21)1133, (01)00884838099272(21)672, (01)00884838099272(21)1109, (01)00884838099272(21)1124, (01)00884838099272(21)995, (01)00884838099272(21)813, (01)00884838099272(21)1163, (01)00884838099272(21)580, (01)00884838099272(21)384, (01)00884838099272(21)896, (01)00884838099272(21)636, (01)00884838099272(21)720, (01)00884838099272(21)742, (01)00884838099272(21)832, (01)00884838099272(21)902, (01)00884838099272(21)867, (01)00884838099272(21)872, (01)00884838099272(21)418, (01)00884838099272(21)231, (01)00884838099272(21)1035, (01)00884838099272(21)1055, (01)00884838099272(21)745, (01)00884838099272(21)1033, (01)00884838099272(21)1105, (01)00884838099272(21)637, (01)00884838099272(21)953, (01)00884838099272(21)546, (01)00884838099272(21)844, (01)00884838099272(21)276, (01)00884838099272(21)649, (01)00884838099272(21)855, (01)00884838099272(21)288, (01)00884838099272(21)137, (01)00884838099272(21)68, (01)00884838099272(21)811, (01)00884838099272(21)652. Serial Numbers: 209, 675, 1091, 113, 1145, 1078, 197, 687, 680, 280, 343, 98, 332, 341, 1162, 554, 337, 400, 757, 326, 323, 262, 1017, 379, 786, 462, 573, 458, 987, 880, 141, 693, 510, 882, 1157, 21, 1050, 367, 184, 178, 179, 250, 1139, 1062, 575, 1121, 950, 368, 176, 712, 829, 249, 576, 1046, 708, 738, 1083, 1098, 717, 1126, 665, 764, 386, 977, 721, 1006, 232, 682, 789, 698, 331, 1097, 1094, 378, 808, 718, 195, 412, 715, 1143, 775, 1073, 982, 282, 617, 958, 222, 430, 1067, 810, 1114, 314, 762, 1102, 359, 978, 1060, 920, 1136, 836, 835, 795, 800, 899, 838, 944, 1123, 335, 1039, 203, 731, 725, 701, 967, 501, 992, 119, 732, 974, 919, 605, 686, 230, 523, 327, 477, 1049, 298, 993, 522, 520, 521, 259, 909, 142, 263, 422, 17, 1057, 36, 1103, 164, 485, 1155, 1054, 366, 1140, 1048, 752, 1088, 47, 544, 906, 289, 548, 238, 299, 284, 404, 283, 401, 833, 895, 233, 224, 1087, 395, 444, 710, 820, 235, 846, 727, 1026, 1129, 777, 481, 431, 696, 1099, 438, 552, 630, 941, 429, 854, 888, 596, 957, 645, 739, 736, 1127, 1117, 527, 251, 1000, 965, 553, 1160, 1042, 699, 1064, 574, 741, 740, 1137, 900, 901, 887, 949, 632, 1112, 1156, 737, 907, 825, 877, 1134, 1161, 648, 831, 1116, 1149, 1104, 860, 798, 603, 938, 773, 939, 1065, 716, 1133, 672, 1109, 1124, 995, 813, 1163, 580, 384, 896, 636, 720, 742, 832, 902, 867, 872, 418, 231, 1035, 1055, 745, 1033, 1105, 637, 953, 546, 844, 276, 649, 855, 288, 137, 68, 811, 652.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls1 death632 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls594 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99939).
- Z-0292-2027 · Open, ClassifiedPhilips Azurion 3 M12. Model Numbers: 722063, 722221.
- Z-0293-2027 · Open, ClassifiedPhilips Azurion 3 M15. Model Number: 722064, 722222, 722280.
- Z-0294-2027 · Open, ClassifiedPhilips Azurion 5 M12. Model Number: 722227.
- Z-0295-2027 · Open, ClassifiedPhilips Azurion 5 M20. Model Numbers: 722228, 722281.
- Z-0296-2027 · Open, ClassifiedPhilips Azurion 7 B12. Model Numbers: 722067, 722225.
- Z-0298-2027 · Open, ClassifiedPhilips Azurion 7 M12. Model Numbers: 722078, 722223.
- Z-0299-2027 · Open, ClassifiedPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page