Skip to content

Unity Knee ASC Modular Handle included in the Corin UNITY… recallZ-0356-2027 · Corin Medical, Ltd. · initiated Oct 2, 2026

Class II Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Corin Medical, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

FDA root cause category: Under Investigation by firm

What the company is doing

Corin notified consignees on about 10/02/2026 via emailed "URGENT MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to identify the products present in inventory and quarantine them, return the devices to Corin (replacement parts will be provided), provide the notification letter to all involved parties, and complete and return the acknowledgement form. Additionally, Corin requested notification if affected units have been further distributed as well as any affected customers so that they may be notified.

Recall details

Recall number
Z-0356-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 2, 2026
Posted by the FDA
Nov 3, 2026
Recalling firm
Corin Medical, Ltd.
Quantity
22 units
Product code
JWH
FDA submissions
K113060
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.

Product description

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Code information (models, lots, serial numbers)

Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.

Other devices under the same FDA submission

3 devices cleared under K113060 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page