CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch… recallZ-0325-2027 · Artivion, Inc · initiated Sep 18, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Artivion, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 9/18/2026, the notification was delivered via mail to customer who was advised of the following: Available biomechanical and histological evidence does not indicate an increased risk that would warrant additional patient testing, enhanced clinical monitoring, or other clinical intervention beyond the patient's routine standard of care. Accordingly, prophylactic removal or surgical replacement of the already implanted tissue solely because of the donor-age nonconformance is not clinically warranted. Complete and return the response form via email to [email protected] Firm requests that any observed deficiency, performance concern, or other issue potentially associated with the tissue be reported to them vial email: [email protected] or phone: +1-770-419-3355
Recall details
- Recall number
- Z-0325-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 18, 2026
- Posted by the FDA
- Oct 30, 2026
- Recalling firm
- Artivion, Inc
- Quantity
- 1
- Product code
- DXZ
- FDA submissions
- K101866
- Distribution
- US Nationwide distribution in the state of IL.
Product description
CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010
Code information (models, lots, serial numbers)
Serial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CryoPatch® SG Pulmonary Human Cardiac Trunk Patch SGP010Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall0 reports
Other devices under the same FDA submission
2 devices cleared under K101866 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CryoPatch® SG Pulmonary Human Cardiac Branch Patch SGP020Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene2 recalls1 report
- CryoPatch® SG Pulmonary Human Cardiac Hemi-Artery Patch SGPH00Artivion, Inc. · Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page