Skip to content

TS 7000 recallZ-0332-2027 · Baxter Healthcare Corporation · initiated Sep 14, 2026

Class II Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

FDA root cause category: Equipment maintenance

What the company is doing

Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.

Recall details

Recall number
Z-0332-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 14, 2026
Posted by the FDA
Nov 3, 2026
Recalling firm
Baxter Healthcare Corporation
Quantity
62 units
Product code
GDC
Distribution
US Nationwide distribution in the states of IL, MO, TX.

Product description

TruSystem 7000 U Operating Table (dV), Product Code 1723633

Code information (models, lots, serial numbers)

UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99826).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page