TS 7500 recallZ-0333-2027 · Baxter Healthcare Corporation · initiated Sep 14, 2026
Class II Open, Classified All recalls posted in November 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
FDA root cause category: Equipment maintenance
What the company is doing
Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
Recall details
- Recall number
- Z-0333-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 14, 2026
- Posted by the FDA
- Nov 3, 2026
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 1 unit
- Product code
- GDC
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
Product description
Mobile column TruSystem 7500 U, Product Code 1730720
Code information (models, lots, serial numbers)
UDI-DI: 00887761974388; Serial number: 110183858
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TS 7500 1730720Baxter Medical Systems GmbH + Co. KG · Table, radiologic2 recalls1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99826).
- Z-0332-2027 · Open, ClassifiedTruSystem 7000 U Operating Table (dV), Product Code 1723633
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.