Skip to content

Veritas rTSA recallZ-0348-2027 · restor3d Inc. · initiated Oct 1, 2026

Class II Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from restor3d Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.

FDA root cause category: Device Design

What the company is doing

On 1, 2026 Restor3d reached out to affected consignees via phone call to make them aware of the recall. A Urgent; Medical Device Recall was issued to affected consignees on October 7, 2026 via email. Restor3d ask consignees to take the following actions: 1. Do not implant or otherwise use any affected units identified above. 2. Complete the attached Field Action Response Form (Attachment A) and return it to restor3d within 10 business days, so that restor3d can confirm effective distribution of this notice. 3. Promptly report to restor3d any intraoperative issue, patient adverse event, suspected device related complication, loosening, migration, revision surgery, or other clinical concern potentially associated with the affected product.

Recall details

Recall number
Z-0348-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 1, 2026
Posted by the FDA
Nov 3, 2026
Recalling firm
restor3d Inc.
Quantity
240 units
Product code
PHX
FDA submissions
K243643
Distribution
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.

Product description

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Code information (models, lots, serial numbers)

All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K243643 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page