Cardinal Health Presource PBDS kits, Containing BD Alaris… recallZ-0218-2027 · Cardinal Health 200, LLC · initiated Sep 24, 2026

Class I Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits contain infusion sets which were recalled for performance issues.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Cardinal Health issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/24/2026 via US Mail. The notice explained the issue and requested the discontinuation of the use of the affected BD Alaris Pump Infusion Sets as per the enclosed notice from Becton Dickinson. The notice instructed the following: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: Hospitals Cardinal Customer Service at 866.551.0533 Distributors Distributor Management at 800.635.6021 Federal Government facilities Government Customer Service at 800.444.1166 Distributors were directed to notify their customers of the recall. For questions contact: Cardinal Health Quality Systems at 800.292.9332.

Recall details

Recall number
Z-0218-2027
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Sep 24, 2026
Posted by the FDA
Nov 2, 2026
Recalling firm
Cardinal Health 200, LLC
Quantity
221554 kits
Product code
LRS
Distribution
US and Canada

Product description

Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) IV START LD KIT , CATALOG # IVS5342277; 2) IV START LD KIT , CATALOG # IVS5342276; 3) IV START LD KIT , CATALOG # IVS5342275; 4) IV START LD KIT , CATALOG # IVS5342274; 5) IV START LD KIT , CATALOG # IVS5342273; 6) IV START LD KIT , CATALOG # IVS5342272; 7) IV START LD KIT , CATALOG # IVS5342271; 8) IV START LD KIT , CATALOG # IVS534227B; 9) IV START BLOOD DRAW,KIT , CATALOG # BLDV125142; 10) RUN W/O SPKVM0102X , CATALOG # SPKVM01011; 11) IV START BLOOD DRAW, KIT , CATALOG # BLDHD03262; 12) IV START BLOOD DRAW, KIT , CATALOG # BLDHD0326C; 13) IV START BLOOD DRAW,KIT , CATALOG # BLDV125141; 14) IV START BLOOD DRAW,KIT , CATALOG # BLDHD03261; 15) KIT,IV START BLOOD DRAW , CATALOG # BLDV12514A; 16) KIT,IV START, , CATALOG # IVSJV83001; 17) IV START LD KIT , CATALOG # IVS534227A; 18) IV START BLOOD DRAW,KIT , CATALOG # PLHDBDHCH; 19) IV START BLOOD DRAW, KIT , CATALOG # PLHDBDHCG; 20) IV START, ,KIT , CATALOG # PJJVIVMMU; 21) KIT,IV START L D , CATALOG # PZVQIVPAN; 22) IV START L D KIT , CATALOG # PZVQIVPAM; 23) KIT,IV START L D , CATALOG # PZVQIVPAM; 24) IV START BLOOD DRAW KIT , CATALOG # BLDHD0326B; 25) KIT,IV START BLOOD DRAW , CATALOG # BLDHD0326B; 26) KIT,IV START L D , CATALOG # PZVQIVPAO; 27) IV START, KIT , CATALOG # PJJVIVMMV; 28) IV START, , KIT , CATALOG # PJJVIVMMV; 29) KIT, IV START, , CATALOG # PJJVIVMMV; 30) IV START ,CHILD KIT, CATALOG # PX29IVACAC; 31) IV START BLOOD DRAW KIT , CATALOG # BLDHD0326A; 32) IV START KIT , CATALOG # PZ82IVTCI; 33) I V START,NMH KIT , CATALOG # PZVBIVNMQ.

Code information (models, lots, serial numbers)

Show the full code information (43k characters)

CATALOG # IVS5342277, GTIN 10888439967814 (ea) 50888439967812 (cs), LOT #540480; CATALOG # IVS5342277, GTIN 10888439967814 (ea) 50888439967812 (cs), LOT #540480; CATALOG # IVS5342276, GTIN 10888439820690 (ea) 50888439820698 (cs), LOT #425675; CATALOG # IVS5342276, GTIN 10888439820690 (ea) 50888439820698 (cs), LOT #425675; CATALOG # IVS5342275, GTIN 10888439793628 (ea) 50888439793626 (cs), LOT #404845; CATALOG # IVS5342275, GTIN 10888439793628 (ea) 50888439793626 (cs), LOT #404845; CATALOG # IVS5342275, GTIN 10888439793628 (ea) 50888439793626 (cs), LOT #404845; CATALOG # IVS5342275, GTIN 10888439793628 (ea) 50888439793626 (cs), LOT #404845; CATALOG # IVS5342274, GTIN 10888439783810 (ea) 50888439783818 (cs), LOT #395356; CATALOG # IVS5342274, GTIN 10888439783810 (ea) 50888439783818 (cs), LOT #395356; CATALOG # IVS5342273, GTIN 10888439676181 (ea) 50888439676189 (cs), LOT #365114; CATALOG # IVS5342273, GTIN 10888439676181 (ea) 50888439676189 (cs), LOT #365114; CATALOG # IVS5342272, GTIN 10888439626469 (ea) 50888439626467 (cs), LOT #331290; CATALOG # IVS5342272, GTIN 10888439626469 (ea) 50888439626467 (cs), LOT #331290; CATALOG # IVS5342271, GTIN 10888439542912 (ea) 50888439542910 (cs), LOT #267592; CATALOG # IVS5342271, GTIN 10888439542912 (ea) 50888439542910 (cs), LOT #267592; CATALOG # IVS5342271, GTIN 10888439542912 (ea) 50888439542910 (cs), LOT #267592; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #208006; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #208006; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #208006; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #233758; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #233758; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #254639; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #254639; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #254639; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #276605; CATALOG # IVS534227B, GTIN 10888439419214 (ea) 50888439419212 (cs), LOT #305961; CATALOG # BLDV125142, GTIN 10887488259383 (ea) 50887488259381 (cs), LOT #334078; CATALOG # BLDV125142, GTIN 10887488259383 (ea) 50887488259381 (cs), LOT #334078; CATALOG # SPKVM01011, GTIN 10887488254340 (ea) 50887488254348 (cs), LOT #345717; CATALOG # BLDHD03262, GTIN 10887488237480 (ea) 50887488237488 (cs), LOT #329751; CATALOG # BLDHD03262, GTIN 10887488237480 (ea) 50887488237488 (cs), LOT #329751; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #335787; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #339093; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #339093; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #344894; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #344894; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #344894; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #344894; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #352206; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #358893; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #358893; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #362820; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #362820; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #362820; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #362820; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #368507; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #368507; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #373755; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #373755; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDHD0326C, GTIN 10887488237473 (ea) 50887488237471 (cs), LOT #379335; CATALOG # BLDV125141, GTIN 10887488221472 (ea) 50887488221470 (cs), LOT #325692; CATALOG # BLDHD03261, GTIN 10887488190129 (ea) 50887488190127 (cs), LOT #318976; CATALOG # BLDV12514A, GTIN 10887488134284 (ea) 50887488134282 (cs), LOT #310119; CATALOG # BLDV12514A, GTIN 10887488134284 (ea) 50887488134282 (cs), LOT #310119; CATALOG # IVSJV83001, GTIN 10887488077956 (ea) 50887488077954 (cs), LOT #217474; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #548118; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #548118; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #562119; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #580728; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #580728; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #580728; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #580728; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #610236; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #610236; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #642791; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #695910; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #695910; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #714857; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #714857; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #714857; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #714857; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #742013; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #742013; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #742013; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #762841; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #762841; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #762841; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #762841; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #790031; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #808290; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #808290; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #837632; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #837632; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #837632; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #859353; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #859353; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #889251; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #889251; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #913349; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #938139; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #938139; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #938139; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #938139; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #960010; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #978100; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #978100; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #978100; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #118310; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #162943; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #162943; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #162943; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #183872; CATALOG # IVS534227A, GTIN 10885425705986 (ea) 50885425705984 (cs), LOT #183872; CATALOG # PLHDBDHCH, GTIN 8854251476211 (ea) 58854251476216 (cs), LOT #121476; CATALOG # PLHDBDHCH, GTIN 8854251476211 (ea) 58854251476216 (cs), LOT #121476; CATALOG # PLHDBDHCH, GTIN 8854251476211 (ea) 58854251476216 (cs), LOT #135117; CATALOG # PLHDBDHCH, GTIN 8854251476211 (ea) 58854251476216 (cs), LOT #138549; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #82792; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #82792; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #82792; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #82792; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #94864; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #94864; CATALOG # PLHDBDHCG, GTIN 8854251315305 (ea) 58854251315300 (cs), LOT #94864; CATALOG # PJJVIVMMU, GTIN 8854251288555 (ea) 58854251288550 (cs), LOT #79656; CATALOG # PJJVIVMMU, GTIN 8854251288555 (ea) 58854251288550 (cs), LOT #79656; CATALOG # PJJVIVMMU, GTIN 8854251288555 (ea) 58854251288550 (cs), LOT #79656; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #721926; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #721926; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #721926; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #730336; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #758494; CATALOG # PZVQIVPAN, GTIN 8854251229510 (ea) 58854251229515 (cs), LOT #758494; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #149722; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #149722; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #153098; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #153098; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #155787; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #155787; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #158373; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #161894; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #161894; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #424591; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #447260; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #447260; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #487668; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #507227; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #507227; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #507227; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #516033; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #532703; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #532703; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #532703; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #532703; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #547528; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #547528; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #547528; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #556150; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #556150; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #559013; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #559013; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #559013; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #559013; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #572290; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #582550; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #592289; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #600958; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #611680; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #624648; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #624648; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #633491; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #633491; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #633491; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #646439; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #664281; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #673237; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #678217; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #678217; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #678217; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #694919; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #694919; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #694919; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #694919; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #694919; CATALOG # PZVQIVPAM, GTIN 8854250199463 (ea) 58854250199468 (cs), LOT #713945; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #146482; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #154387; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #161162; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #161162; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #165167; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #167987; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #167987; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #167987; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #167987; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #172512; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #172512; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #178359; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #178359; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #178359; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #181358; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #181358; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #145224; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #145224; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #187359; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #211061; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #211061; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #211061; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #229844; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #246634; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #254793; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #254793; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #254793; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #281173; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #281173; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #304567; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #304567; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #336039; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #336039; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #341699; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #359680; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #359680; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #392697; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #392697; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #424548; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #424548; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #444960; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #444960; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #444960; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #471964; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #471964; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #471964; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #471964; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #492187; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #492187; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #492187; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #492187; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #523360; CATALOG # IVS534227A, GTIN 885425705989 (ea) 50885425705984 (cs), LOT #523360; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #122981; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #122981; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #122981; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #122981; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #130611; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #130611; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #130611; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #130611; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #135443; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #135443; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #142516; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #142516; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #147541; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #147541; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #147541; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #151838; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #156082; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #156082; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #156082; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #156082; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #156082; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #157164; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #157164; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #157164; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #161485; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #161485; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #164678; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #164678; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #164678; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #167379; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #167379; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #167379; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #167379; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #169805; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #170872; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #170872; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #173717; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #173717; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #173717; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #262997; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #262997; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #262997; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #263647; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #263647; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #263647; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #272714; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #272714; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #272714; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #272714; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #281107; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #287096; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #287096; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #291680; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #296997; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #296997; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #302525; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #302525; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #302525; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #306574; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #306574; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #306574; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #309125; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #309125; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #309125; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #309125; CATALOG # BLDHD0326B, GTIN 885425572444 (ea) 50885425572449 (cs), LOT #309125; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #765863; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #778631; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #778631; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #778631; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #778631; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #793624; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #807238; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #807238; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #807238; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #807238; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #807238; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #824947; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #839013; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #852644; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #852644; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #852644; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #852644; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #852644; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #858224; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #858224; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #873261; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #880039; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #888756; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #898042; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #898042; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #898042; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #898042; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #898042; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #911606; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #911606; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #911606; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #911606; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #934402; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #934402; CATALOG # PZVQIVPAO, GTIN 885425345604 (ea) 50885425345609 (cs), LOT #948747; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #99001; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #99001; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #99001; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #121921; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #121921; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #129995; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #129995; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #132429; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #140727; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #140727; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #154735; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #154735; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #163407; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #166921; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #170615; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #182174; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #182174; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #182174; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #189925; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #189925; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #189925; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #189925; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #189925; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #198472; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #209218; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #218286; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #218286; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #218286; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #218286; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #224131; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #44365; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #44365; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #44365; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #59690; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #89791; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #116134; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #173288; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #173288; CATALOG # PJJVIVMMV, GTIN 885425223483 (ea) 50885425223488 (cs), LOT #196798; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #858864; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #870654; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #870654; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #875827; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #875827; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #875827; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #879589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #879589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #885399; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #885399; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #887613; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #887613; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #887613; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #893289; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #893289; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #896927; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #903356; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #903356; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #903356; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #908691; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #908691; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #914403; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #914403; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #922250; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #922250; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #922250; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #922250; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #922250; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #929254; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #929254; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #936821; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #936821; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #936821; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #944629; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #944629; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #944629; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #944629; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #951147; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #951147; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #956600; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #961372; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #961372; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #961372; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #969632; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #969632; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #969632; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #969632; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #976077; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #976077; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #2174; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #2174; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #2174; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #8963; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #17568; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #17568; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #17568; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #17568; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #25586; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #31863; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #31863; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #31863; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #45064; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #59657; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #59657; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #59657; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #66936; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #66936; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #80161; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #80161; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #80161; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #80161; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #87589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #87589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #87589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #87589; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #96074; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #96074; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #100605; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #108951; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #115390; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #115390; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #121177; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #121177; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #121177; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #127624; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #129893; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #129893; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #134096; CATALOG # PX29IVACAC, GTIN 885425124445 (ea) 50885425124440 (cs), LOT #134096; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #87460; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #87460; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #101367; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #101367; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #101367; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #101367; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #101367; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #113519; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #113519; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #113519; CATALOG # BLDHD0326A, GTIN 885425103938 (ea) 50885425103933 (cs), LOT #113519; CATALOG # PZ82IVTCI, GTIN 383810631535 (ea) 50383810675725 (cs), LOT #127379; CATALOG # PZ82IVTCI, GTIN 383810631535 (ea) 50383810675725 (cs), LOT #127379; CATALOG # PZ82IVTCI, GTIN 383810631535 (ea) 50383810675725 (cs), LOT #127379; CATALOG # PZ82IVTCI, GTIN 383810631535 (ea) 50383810675725 (cs), LOT #127379; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #849862; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #849862; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #849862; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #849862; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #851775; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #854070; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #854070; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #856399; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #856399; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #856399; CATALOG # PZVBIVNMQ, GTIN 382603428932 (ea) 50382603428937 (cs), LOT #856399.