Cardinal Health Presource PBDS kits, Containing BD Alaris… recallZ-0220-2027 · Cardinal Health 200, LLC · initiated Sep 24, 2026
Class I Open, Classified All recalls posted in November 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain infusion sets which were recalled for performance issues.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Cardinal Health issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/24/2026 via US Mail. The notice explained the issue and requested the discontinuation of the use of the affected BD Alaris Pump Infusion Sets as per the enclosed notice from Becton Dickinson. The notice instructed the following: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: [email protected] For component credit or replacement, please contact the following: Hospitals Cardinal Customer Service at 866.551.0533 Distributors Distributor Management at 800.635.6021 Federal Government facilities Government Customer Service at 800.444.1166 Distributors were directed to notify their customers of the recall. For questions contact: Cardinal Health Quality Systems at 800.292.9332.
Recall details
- Recall number
- Z-0220-2027
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Sep 24, 2026
- Posted by the FDA
- Nov 2, 2026
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 47707 kits
- Product code
- MLS
- Distribution
- US and Canada
Product description
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT,LABOR KIT, , CATALOG # MA1JV8300C; 2) KIT,LABOR KIT, , CATALOG # MA1JV8300B; 3) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATM3; 4) KIT,LABOR KIT, , CATALOG # MA1JV83001; 5) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATM2; 6) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMN; 7) KIT,LABOR KIT, , CATALOG # MA1JV8300A; 8) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMB; 9) LABOR, KIT , CATALOG # PBJVAD00S; 10) KIT,LABOR, , CATALOG # PL41OTHS4; 11) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMA; 12) KIT,LABOR, , CATALOG # PL41OTHSV; 13) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATML; 14) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMK; 15) KIT,LABOR KIT, , CATALOG # PBJVAD00X; 16) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMJ; 17) LABOR KIT, ,KIT , CATALOG # PBJVAD00W; 18) LABOR KIT, ,KIT , CATALOG # PBJVAD00U; 19) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMH; 20) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATMG; 21) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATM1; 22) KIT, VAGINAL DELIVERY , CATALOG # PL42VDATME; 23) KIT,LABOR, , CATALOG # PL41OTHSW; 24) LABOR KIT, KIT , CATALOG # PBJVAD00T; 25) LABOR, KIT , CATALOG # PBJVAD00T; 26) LABOR, , KIT , CATALOG # PBJVAD00T; 27) CENTER KIT , CATALOG # PL73FBDW5; 28) CENTER, KIT , CATALOG # PL73FBDW4; 29) CENTER, KIT , CATALOG # PL73FBDW3; 30) CENTER, KIT , CATALOG # PL73FBDW2; 31) KIT, CENTER,, CATALOG # PL73FBDW1; 32) KIT, CENTER,, CATALOG # PL73OTDW1; 33) KIT, CENTER,, CATALOG # PL73OTDWU; 34) KIT, CENTER,, CATALOG # PL73OTDWT; 35) KIT, CENTER,, CATALOG # PL73OTDWS.
Code information (models, lots, serial numbers)
Show the full code information (18k characters)
CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #500060; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #500060; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #500060; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #500060; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #519838; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #519838; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #541166; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #541166; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #541166; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #541166; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #561757; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #561757; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #561757; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #582974; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #605722; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #605722; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #605722; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #605722; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #630532; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #630532; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #649740; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #649740; CATALOG #MA1JV8300C, GTIN 10887488483146 (ea) 50887488483144 (cs), LOT #676253; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #408506; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #408506; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #424817; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #451873; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #451873; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #451873; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #472033; CATALOG #MA1JV8300B, GTIN 10887488323428 (ea) 50887488323426 (cs), LOT #472033; CATALOG #PL42VDATM3, GTIN 10887488304687 (ea) 50887488304685 (cs), LOT #378304; CATALOG #PL42VDATM3, GTIN 10887488304687 (ea) 50887488304685 (cs), LOT #378304; CATALOG #MA1JV83001, GTIN 10887488268576 (ea) 50887488268574 (cs), LOT #356036; CATALOG #MA1JV83001, GTIN 10887488268576 (ea) 50887488268574 (cs), LOT #356036; CATALOG #MA1JV83001, GTIN 10887488268576 (ea) 50887488268574 (cs), LOT #356036; CATALOG #MA1JV83001, GTIN 10887488268576 (ea) 50887488268574 (cs), LOT #356036; CATALOG #MA1JV83001, GTIN 10887488268576 (ea) 50887488268574 (cs), LOT #356036; CATALOG #PL42VDATM2, GTIN 10887488212449 (ea) 50887488212447 (cs), LOT #311891; CATALOG #PL42VDATM2, GTIN 10887488212449 (ea) 50887488212447 (cs), LOT #311891; CATALOG #PL42VDATM2, GTIN 10887488212449 (ea) 50887488212447 (cs), LOT #311891; CATALOG #PL42VDATMN, GTIN 10887488165011 (ea) 50887488165019 (cs), LOT #324429; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #221591; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #221591; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #221591; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #221591; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #233296; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #233296; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #233296; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #252473; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #252473; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #274401; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #295340; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #295340; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #309090; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #309090; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #309090; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #320084; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #327759; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #327759; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #345386; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #358433; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #358433; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #358433; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #358433; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #358433; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #377461; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #377461; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #377461; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #377461; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #391108; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #391108; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #391108; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #391108; CATALOG #MA1JV8300A, GTIN 10887488061689 (ea) 50887488061687 (cs), LOT #391108; CATALOG #PL42VDATMB, GTIN 8854251488672 (ea) 58854251488677 (cs), LOT #815004; CATALOG #PL42VDATMB, GTIN 8854251488672 (ea) 58854251488677 (cs), LOT #815004; CATALOG #PBJVAD00S, GTIN 8854251374814 (ea) 58854251374819 (cs), LOT #96940; CATALOG #PBJVAD00S, GTIN 8854251374814 (ea) 58854251374819 (cs), LOT #96940; CATALOG #PL41OTHS4, GTIN 8854251296116 (ea) 58854251296111 (cs), LOT #739011; CATALOG #PL41OTHS4, GTIN 8854251296116 (ea) 58854251296111 (cs), LOT #739011; CATALOG #PL42VDATMA, GTIN 8854251273698 (ea) 58854251273693 (cs), LOT #764603; CATALOG #PL42VDATMA, GTIN 8854251273698 (ea) 58854251273693 (cs), LOT #799434; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #703867; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #718045; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #722172; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #722172; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #751701; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #751701; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #751701; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #757930; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #769962; CATALOG #PL41OTHSV, GTIN 8854251177231 (ea) 58854251177236 (cs), LOT #769962; CATALOG #PL42VDATML, GTIN 885425992198 (ea) 50885425992193 (cs), LOT #193818; CATALOG #PL42VDATML, GTIN 885425992198 (ea) 50885425992193 (cs), LOT #204464; CATALOG #PL42VDATML, GTIN 885425992198 (ea) 50885425992193 (cs), LOT #204464; CATALOG #PL42VDATML, GTIN 885425992198 (ea) 50885425992193 (cs), LOT #248433; CATALOG #PL42VDATML, GTIN 885425992198 (ea) 50885425992193 (cs), LOT #253705; CATALOG #PL42VDATMK, GTIN 885425928630 (ea) 50885425928635 (cs), LOT #121460; CATALOG #PL42VDATMK, GTIN 885425928630 (ea) 50885425928635 (cs), LOT #132774; CATALOG #PL42VDATMK, GTIN 885425928630 (ea) 50885425928635 (cs), LOT #140701; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #23421; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #33554; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #33554; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #33554; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #33554; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #33554; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #43203; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #49540; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #49540; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #57465; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #67261; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #67261; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #91489; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #91489; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #103954; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #130262; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #146244; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #173286; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #173286; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #173286; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #173286; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #173286; CATALOG #PBJVAD00X, GTIN 885425805863 (ea) 50885425805868 (cs), LOT #198680; CATALOG #PL42VDATMJ, GTIN 885425800899 (ea) 50885425800894 (cs), LOT #82497; CATALOG #PBJVAD00W, GTIN 885425779133 (ea) 50885425779138 (cs), LOT #218461; CATALOG #PBJVAD00W, GTIN 885425779133 (ea) 50885425779138 (cs), LOT #218461; CATALOG #PBJVAD00W, GTIN 885425779133 (ea) 50885425779138 (cs), LOT #218461; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #197672; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #197672; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #206659; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #206659; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #206659; CATALOG #PBJVAD00U, GTIN 885425673134 (ea) 50885425673139 (cs), LOT #206659; CATALOG #PL42VDATMH, GTIN 885425670416 (ea) 50885425670411 (cs), LOT #32603; CATALOG #PL42VDATMH, GTIN 885425670416 (ea) 50885425670411 (cs), LOT #32603; CATALOG #PL42VDATMH, GTIN 885425670416 (ea) 50885425670411 (cs), LOT #32603; CATALOG #PL42VDATMH, GTIN 885425670416 (ea) 50885425670411 (cs), LOT #32603; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #905014; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #905014; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #905014; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #934476; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #951889; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #972754; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #998599; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #998599; CATALOG #PL42VDATMG, GTIN 885425572888 (ea) 50885425572883 (cs), LOT #998599; CATALOG #PL42VDATM1, GTIN 885425540290 (ea) 50885425540295 (cs), LOT #875392; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #869818; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #869818; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #869818; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #879050; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #879050; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #893335; CATALOG #PL42VDATME, GTIN 885425334776 (ea) 50885425334771 (cs), LOT #893335; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #784070; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #784070; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #784070; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #799430; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #799430; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #809338; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #809338; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #824874; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #824874; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #840214; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #840214; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #858782; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #878523; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #882411; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #894096; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #904576; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #904576; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #915474; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #915474; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #934509; CATALOG #PL41OTHSW, GTIN 885425334660 (ea) 50885425334665 (cs), LOT #934509; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #128733; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #128733; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #146470; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #146470; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #146470; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #146470; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #155463; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #160016; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #167090; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #167090; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #168600; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #174272; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #174272; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #178681; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #178681; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #178681; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #189774; CATALOG #PBJVAD00T, GTIN 885425212364 (ea) 50885425212369 (cs), LOT #189774; CATALOG #PL73FBDW5, GTIN 382603511337 (ea) 50382603511332 (cs), LOT #119317; CATALOG #PL73FBDW5, GTIN 382603511337 (ea) 50382603511332 (cs), LOT #119317; CATALOG #PL73FBDW5, GTIN 382603511337 (ea) 50382603511332 (cs), LOT #119317; CATALOG #PL73FBDW4, GTIN 382603461434 (ea) 50382603461439 (cs), LOT #115246; CATALOG #PL73FBDW3, GTIN 382603417103 (ea) 50382603417108 (cs), LOT #110645; CATALOG #PL73FBDW3, GTIN 382603417103 (ea) 50382603417108 (cs), LOT #110645; CATALOG #PL73FBDW2, GTIN 382603368252 (ea) 50382603368257 (cs), LOT #107365; CATALOG #PL73FBDW2, GTIN 382603368252 (ea) 50382603368257 (cs), LOT #107365; CATALOG #PL73FBDW2, GTIN 382603368252 (ea) 50382603368257 (cs), LOT #107365; CATALOG #PL73FBDW1, GTIN 382603349398 (ea) 50382603349393 (cs), LOT #105930; CATALOG #PL73OTDW1, GTIN 382603325774 (ea) 50382603325779 (cs), LOT #103603; CATALOG #PL73OTDW1, GTIN 382603325774 (ea) 50382603325779 (cs), LOT #103603; CATALOG #PL73OTDW1, GTIN 382603325774 (ea) 50382603325779 (cs), LOT #103603; CATALOG #PL73OTDW1, GTIN 382603325774 (ea) 50382603325779 (cs), LOT #103603; CATALOG #PL73OTDWU, GTIN 382603276151 (ea) 50382603276156 (cs), LOT #100510; CATALOG #PL73OTDWU, GTIN 382603276151 (ea) 50382603276156 (cs), LOT #100510; CATALOG #PL73OTDWU, GTIN 382603276151 (ea) 50382603276156 (cs), LOT #100510; CATALOG #PL73OTDWT, GTIN 382603166049 (ea) 50382603166044 (cs), LOT #994976; CATALOG #PL73OTDWT, GTIN 382603166049 (ea) 50382603166044 (cs), LOT #994976; CATALOG #PL73OTDWS, GTIN 382603112350 (ea) 50382603112355 (cs), LOT #991899; CATALOG #PL73OTDWS, GTIN 382603112350 (ea) 50382603112355 (cs), LOT #991899.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CARDINAL HEALTH MA1JV8300CCardinal Health 200, LLC · Labor and Delivery Kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 100058).
- Z-0212-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, ANESTHESIA, CATALOG # IVSVM41612; 2) …
- Z-0213-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) STR CATH ANGIO PEDIATRIC PACK , CATALOG #SA…
- Z-0215-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, CABG, , CATALOG # PCOCCG3A…
- Z-0216-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT,C SECTION ADMIT,, CATALOG # MATJV8300D…
- Z-0217-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, ANTEPARTUM IV , CATALOG # M…
- Z-0218-2027 · Open, ClassifiedCardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) IV START LD KIT , CATALOG # …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page