Senographe Pristina recallZ-0326-2027 · GE Medical Systems, LLC · initiated Aug 28, 2026
Class II Open, Classified All recalls posted in November 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
FDA root cause category: Radiation Control for Health and Safety Act
What the company is doing
GE HealthCare will send a customer letter dated August 28, 2026, that provides safety instructions for immediate mitigation of the incorrect configuration of the collimator anode track / filter in manual mode during 3D+2D acquisitions. GE HealthCare will correct all affected Senographe Pristina VIA systems by upgrading the software at no cost to the customer. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0326-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 28, 2026
- Posted by the FDA
- Nov 4, 2026
- Recalling firm
- GE Medical Systems, LLC
- Product code
- MUE
- FDA submissions
- K260087
- Distribution
- USA and OUS.
Product description
Senographe Pristina
Code information (models, lots, serial numbers)
GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- Senographe Pristina EKB4Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina EKB6Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina NAY2Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina NAY3Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina EKB1Ge Medical Systems · Full field digital,system,x-ray,mammographic4 recalls0 reports
- Senographe Pristina EKB2Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina EKB3Ge Medical Systems · Full field digital, system, x-ray, mammographic4 recalls0 reports
- Senographe Pristina MBE2Ge Medical Systems · Full Field Digital, System, X-Ray, Mammographic3 recalls0 reports
- Senographe Pristina POZ2Ge Medical Systems · Full Field Digital, System, X-Ray, Mammographic2 recalls0 reports
- Not Applicable PRISTINA 2.9Ge Medical Systems · Full Field Digital, System, X-Ray, Mammographic2 recalls0 reports
- not applicable PRISTINA 2.10Ge Medical Systems · Full Field Digital, System, X-Ray, Mammographic2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 100070).
- Z-0327-2027 · Open, ClassifiedSenographe Pristina
- Z-0328-2027 · Open, ClassifiedSenographe Pristina 3D
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.