Baxter Interlink System Microbore T-Connector Extension Set… recallZ-0518-03 · Baxter Healthcare Corp. · initiated Oct 17, 2002
Terminated All recalls posted in February 2003
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
FDA root cause category: Other
What the company is doing
Recalled by letter dated 10/17/02. Accounts were informed of the potential for the filters to be 1.2 micron filters instead of the 0.22. micron filters as labeled. The accounts were requested to examine their inventories for the affected catalog numbers and destroy the units found for credit.
Recall details
- Recall number
- Z-0518-03
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 17, 2002
- Posted by the FDA
- Feb 6, 2003
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 265,240 units
- Product code
- FPA
- FDA submissions
- K791496, K921899, K925126
- Distribution
- Nationwide, Australia, Puerto Rico, Brazil, Canada, Sweden, New Zealand, Japan, Mexico, American Samoa, El Salvador, Turkey, Saudi Arabia, Austria, Cayman Islands, Macedonia, Honduras, Kuwait, Hong Kong and Nigeria.
Product description
Baxter Interlink System Microbore T-Connector Extension Set with 0.22 Micron High Pressure Pediatric Filter, catalog 2N3333; Baxter Healthcare Corporation, Deerfield, IL 60015; 200 units per case
Code information (models, lots, serial numbers)
All lots received prior to 10/23/2002
Other devices under the same FDA submission
9 devices cleared under K791496, K921899, K925126 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- INTERLINK 2N3328Baxter International Inc. · Set, administration, intravascular46 reports
- INTERLINK 2N3339Baxter International Inc. · Set, administration, intravascular14 reports
- INTERLINK 2N3374Baxter International Inc. · Set, administration, intravascular7 reports
- INTERLINK 2N3378Baxter International Inc. · Set, administration, intravascular7 reports
- INTERLINK 2N3379Baxter Healthcare Corporation · Set, administration, intravascular6 reports
- INTERLINK 2N3399Baxter Healthcare Corporation · Set, administration, intravascular2 reports
- NA 2C1145Baxter International Inc. · Set, administration, intravascular2 reports
- INTERLINK 2N3327Baxter International Inc. · Set, administration, intravascular1 report
- INTERLINK 2N3340Baxter International Inc. · Set, administration, intravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 25088).
- Z-0505-03 · TerminatedBaxter Interlink System Minivolume Extension Set; catalog 1C8412; Baxter Healthcare Corporation, Deerfield, IL 60015
- Z-0507-03 · TerminatedBaxter Interlink System Extension Set with 0.22 Micron Filter, catalog 1C8546, Baxter Healthcare Corporation, Deerfield, IL 60015
- Z-0509-03 · TerminatedBaxter Interlink System Minivolume Extension Set with 0.22 Micron Filter, catalog 1C8577, Baxter Healthcare Corporation, Deerfield, IL 60015
- Z-0517-03 · TerminatedBaxter Interlink System Microbore T-Connector Extension Set with 0.22 Micron High Pressure Pediatric Filter, catalog 2N3332; Baxter Healthcare Corporation, Deerfield, IL 60015; 200…
- Z-0519-03 · TerminatedBaxter Microbore 3-Lead Extension Set with 0.22 Micron High Pressure Pediatric Filter, catalog 2N3342; Baxter Healthcare Corporation, Deerfield, IL 60015; 50 units per case
- Z-0520-03 · TerminatedBaxter 36'' Micro-Volume Extension Set, catalog 2N3347; Baxter Healthcare Corporation, Deerfield, IL 60015
- Z-0521-03 · TerminatedBaxter 60'' Micro-Volume Extension Set, catalog 2N3350; Baxter Healthcare Corporation, Deerfield, IL 60015
- Z-0522-03 · TerminatedBaxter Interlink System Minivolume Extension Set; catalog 1C8483; Baxter Healthcare Corporation, Deerfield, IL 60015
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page