Product is labeled as Cordis Pre-shaped Prowler Plus MP… recallZ-0018-04 · Cordis Neurovascular, Inc. · initiated Oct 19, 2001
Terminated All recalls posted in October 2003
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cordis Neurovascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
These catheters have J-shape tips instead of the labeled MP (multipurpose) tips and also are packed into the incorrect tip tray.
FDA root cause category: Other
What the company is doing
The firm initiated a recall of the product shipped to Japan via an e-mail notification on 8/20/2001 requesting return of the devices. The firm mailed recall notification letters to accounts in the United States on 10/19/2001. The firm considers the recall complete and effective and has accounted for all devices involved.
Recall details
- Recall number
- Z-0018-04
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 19, 2001
- Posted by the FDA
- Oct 15, 2003
- Recalling firm
- Cordis Neurovascular, Inc.
- Quantity
- 54
- FDA submissions
- K003925
- Distribution
- Product was distributed to 4 US hospital accounts and to one Distirbutor in Japan.
Product description
Product is labeled as Cordis Pre-shaped Prowler Plus MP Infusion Catheter (3 F proximal OD, 2.3 F Distal OD, 45 cm distal length, total length 155 cm) with an MP (45 degree) dual marker band tip. Also the package states that it is distributed by Cordis Endovascular systems, Inc, Miami, FL 33102 a Johnson & Johnson Company.
Code information (models, lots, serial numbers)
Catalog No. 606-2511 FX UPN H7396062511FX3 Lot T0601010, expiration date 2004-05
Other devices under the same FDA submission
8 devices cleared under K003925 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PROWLER 606-151XMedos International Sàrl · Catheter, continuous flush8 reports
- PROWLER 606-151FXMedos International Sàrl · Catheter, continuous flush7 reports
- PROWLER 606-171Medos International Sàrl · Catheter, continuous flush2 reports
- PROWLER 606-051FXMedos International Sàrl · Catheter, continuous flush1 report
- PROWLER 606-051XMedos International Sàrl · Catheter, continuous flush1 report
- PROWLER 606-151JXMedos International Sàrl · Catheter, continuous flush1 report
- PROWLER 606-2510Medos International Sàrl · Catheter, continuous flush1 report
- PROWLER 606-2510XMedos International Sàrl · Catheter, continuous flush1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page