Skip to content

Pulse Generator recallZ-0052-04 · St Jude Medical · initiated Jul 31, 2003

Terminated All recalls posted in October 2003

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from St Jude Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Under certain circumstances, pacemaker has a potential to deliver a short coupled pacing interval of approx. 300 msec (200ppm). It is also possible that a patient may experience (up to max of 12) shorter than anticipated pacing intervals.

FDA root cause category: Other

What the company is doing

Sales Representatives will personally contact the physicians, explain the situation, leave a technical memo and a list of implanted patients for the doctor to follow-up with. The ''Dear Doctor'' letter to accompany the representatives is dated 7/31/2003. Instruments not implanted will be returned for reprocessing.

Recall details

Recall number
Z-0052-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 31, 2003
Posted by the FDA
Oct 24, 2003
Recalling firm
St Jude Medical
Quantity
775
Product code
DXY
FDA submissions
P880086
Distribution
Nationwide and to four VA Medical Centers in IA and MO.

Product description

Pulse Generator

Code information (models, lots, serial numbers)

Integrity ADx DR Model 5360 & 5366 Identity ADx DR model 5286, 5380, 5386 &5480 Verity ADx DR 5256, 5356 & 5456

Other devices under the same FDA submission

38 devices cleared under P880086 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page