Magnetom Avanto MRI System, Model 7391167 recallZ-1395-04 · Siemens Medical Solutions USA, Inc · initiated Jul 28, 2004

Terminated All recalls posted in August 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An input error can be made with Magnetom MRI software. Software allows users to manually change Transmit Reference Voltage in certain windows

FDA root cause category: Other

What the company is doing

The recalling firm issued a Customer Safety Advisory letter via certified mail to their customers per Update Instructions MR028/04/S. The letter informs customers of the potential issue and provides instructions to avoid its occurence.

Recall details

Recall number
Z-1395-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 28, 2004
Posted by the FDA
Aug 24, 2004
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
29 units
Product code
LNH
FDA submissions
K032428
Distribution
The products were shipped to 39 medical facilities in CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, MA, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, and VA.

Product description

Magnetom Avanto MRI System, Model 7391167

Code information (models, lots, serial numbers)

Serial numbers 25001 - 25129

Other devices under the same FDA submission

1 device cleared under K032428 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates