MAGNETOM Avanto 10684332Recalls & reported problems
Siemens Healthcare GmbH · MRI Scanner · Class 2
All models of MAGNETOM Avanto (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04056869006635), received between Oct 28, 2020 and Aug 26, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
8 other recalls cover devices cleared under the same FDA submission (K032428, K052164, K052423, K062454, K121434, K130885); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Entrapment of Device1
- Improper or Incorrect Procedure or Method1
Patient problems
- Bone Fracture(s)1
- Joint Dislocation1
- Laceration(s)1
- Skin Tears1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SIEMENS HEALTHINEERS WAS NOTIFIED OF A RIB INJURY TO A PATIENT UNDERGOING A BREAST MRI EXAM USING A MAGNETOM AVANTO SYSTEM. ALLEGEDLY, THE PATIENT HAD HIGH BODY MASS LIMITING THE SPACE INSIDE THE MRI BORE. THE PATIENT ALLEGEDLY PANICKED INSIDE THE BORE THRASHING AND COMPLAINING OF A RIB INJURY. OUR INVESTIGATION COULDN¿T CONFIRM THE EXACT NATURE AND EXTENT OF THE ALLEGED INJURY AND WHETHER IT NEEDED ANY MEDICAL ATTENTION OR INTERVENTION.
Patient problems: Bone Fracture(s)
MAUDE report 22892962 ↗InjuryDec 15, 2022Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE PATIENT WAS INJURED DURING MOVEMENT INTO THE GANTRY AFTER THE PATIENT POSITIONING PROCEDURE. ADDITIONAL INFORMATION RECEIVED BY SIEMENS ON DECEMBER 13, 2022, CLARIFIED THAT THE PATIENT'S FIFTH DIGIT ON THE RIGHT HAND BECAME STUCK BETWEEN TABLETOP AND BODY COIL. ACCORDING TO THE HOSPITAL STAFF, SOFT TISSUE SWELLING OF THE DIGIT WAS OBSERVED AT THE TIME OF INCIDENT. FURTHERMORE, IT WAS STATED THAT THERE WAS A LACERATION WHICH REQUIRED THREE (3) SUTURES AND THE DIGIT WAS DISLOCATED. THE PATIENT REQUIRED A SURGICAL PROCEDURE TO RESOLVED SEVERE ULNAR DEVIATION OF THE METACARPOPHALANGEAL JOINT OF THE FIFTH DIGIT.
Patient problems: Joint Dislocation; Laceration(s)
MAUDE report 15988001 ↗InjuryOct 28, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE MAGNETOM AVANTO DOT SYSTEM. DURING TABLE MOVEMENTS, A PATIENT´S FINGER TIP WAS CAUGHT BETWEEN THE TABLETOP AND TROLLEY. AS A RESULT, THE PATIENTS FINGER TIP WAS SEVERED. THE PATIENT WAS EVALUATED IN THE EMERGENCY ROOM AND SURGERY WAS PERFORMED TO REATTACH THE SEVERED TISSUE. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Skin Tears
MAUDE report 10747159 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032428, K052164, K052423, K062454, K121434, K130885. One submission often covers many models, so check the product and lot details in each record.
- Z-2661-2025 · Aug 28, 2025 · Open, ClassifiedMAGNETOM Skyra fit. Model Number: 10849580.
- Z-2665-2025 · Aug 28, 2025 · Open, ClassifiedMAGNETOM Verio Dot. Model Number: 10684333.
- Z-2666-2025 · Aug 28, 2025 · Open, ClassifiedMAGNETOM Verio Dot Upgrade. Model Number: 10684334.
- Z-0467-2016 · Dec 7, 2015 · TerminatedMAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
- Z-1308-2015 · Mar 9, 2015 · TerminatedMAGNETOM systems Aera/Skyra/Avanto/Verio with software syngo MR D13A ; indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal
- Z-2021-2011 · Mar 3, 2011 · TerminatedMAGNETOM Avanto, Espree, Trio A Tim System. Nuclear magnetic resonance imaging.
- Z-1528-05 · Aug 12, 2005 · TerminatedMagnetom Avanto. Magnetic Resonance Imaging System. Model number 7391167
- Z-1395-04 · Jul 28, 2004 · TerminatedMagnetom Avanto MRI System, Model 7391167
Other System, nuclear magnetic resonance imaging devices
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
- Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging13 recalls2 reports
- NA Discovery MR750w 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging13 recalls0 reports
- Ingenia Ingenia Ambition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging12 recalls12 reports
- Ingenia Ingenia 3.0T CXPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls59 reports
- Ingenia Smartpath to dStream for 3.0TPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls44 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls5 reports
- MR 7000 Series MR 7700Philips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page