Skip to content

Telesheath Left Atrial Introducer System Right Superior.… recallZ-1426-04 · St. Jude Medical/Daig Division · initiated Aug 4, 2004

Terminated All recalls posted in September 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from St. Jude Medical/Daig Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

St. Jude Medical received two complaints from customers involving the Telesheath Left Atrial Introducer System side arm detaching from the hub of the inner sheath. Subsequent testing of Telesheath inventory at St. Jude Medical indicated that the side arm from the inner sheath can detach during routine handling and manipulation, and occurs on a random basis.

FDA root cause category: Other

What the company is doing

100% of customers were notified by a personal visit by a St. Jude Medical representative, starting 8/4/2004. Letters were given by the representative to the consignee concerning the recall.

Recall details

Recall number
Z-1426-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 4, 2004
Posted by the FDA
Sep 1, 2004
Recalling firm
St. Jude Medical/Daig Division
Quantity
55 kits
Product code
DYB
FDA submissions
K964518, K973840, K982187
Distribution
US, Canada, Belgium, Hong Kong, and Japan.

Product description

Telesheath Left Atrial Introducer System Right Superior. Telesheath Catheter Two-Piece Introducer Kit with Hemostasis Valve. Left Lateral Reorder #407901. The product is packaged in an open, semi-rigid plastic tray, which is then placed in a sterile barrier pouch consisting of Tyvek on one side and a clear plastic film on the other. This pouch is then heat sealed prior to sterilization.

Code information (models, lots, serial numbers)

1063074 - Exp Date: 03/2007, 1075104, - Exp. Date: 05/2007, 1076753 - Exp. Date: 06/2007

Other devices under the same FDA submission

8 devices cleared under K964518, K973840, K982187 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 29848).

Sources and updates