Fast-Cath™ 406842Recalls & reported problems
St. Jude Medical, Inc. · Transseptal Guiding Introducer · Class 2
All models of Fast-Cath (71)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05414734203364), received between Oct 15, 2019 and Feb 26, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K964518); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- Stroke/CVA1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2025Fluid/Blood Leak; Fluid/Blood Leak
DURING THE ATRIAL FIBRILLATION PROCEDURE, BLOOD LEAKED FROM THE HEMOSTASIS VALVE WHEN APPLYING NEGATIVE PRESSURE. THE INTRODUCER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. NO AIR MOVEMENT INTO THE PATIENT WAS CONFIRMED. NO ADDITIONAL PUNCTURE WAS PERFORMED WHEN THE INTRODUCER WAS REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21472284 ↗InjuryNov 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RELATED MANUFACTURER REPORT NUMBER: 3008452825-2020-00602, 3005334138-2020-00546, 3005334138-2020-00547. FOLLOWING AN AF ABLATION PROCEDURE, THE PATIENT EXPERIENCED A STROKE. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES. A TRANSESOPHAGEAL ECHO WAS PERFORMED TO EXCLUDE A LEFT ATRIAL APPENDAGE (LAA) THROMBUS, THERE WAS QUESTION OF A THROMBUS IN THE LAA BUT IT WAS DECIDED TO BE UNLIKELY. A TARGET ACT OF 300 OR ABOVE IS MAINTAINED FOR ALL LA ABLATIONS PER THE REPRESENTATIVE. THE PHYSICIAN DID NOT ALLEGE ANY ABBOTT PRODUCT WAS THE CAUSE FOR THE CVA.
Patient problems: Stroke/CVA
MAUDE report 10787557 ↗MalfunctionOct 15, 2019Fluid/Blood Leak; Fluid/Blood Leak
DURING AN ATRIAL FIBRILLATION PROCEDURE, BLOOD WAS LEAKING FROM THE SHEATH. THE SHEATH WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9190965 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964518. One submission often covers many models, so check the product and lot details in each record.
- Z-0617-2013 · Mar 28, 2012 · TerminatedFast-Cath Transseptal Guiding Introducer, 90, 8F, REF 406586, Length 60 cm, Sterile EO, Rx only, St. Jude Medical, 14901 DeVeau Place, Minnetonka, MN 55343 Indicated for introduci
- Z-0535-2007 · Oct 5, 2006 · TerminatedFastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and dilator; Product # 406850.
- Z-1425-04 · Aug 4, 2004 · TerminatedTelesheath Left Atrial Introducer System Right Superior. Telesheath Catheter Two-Piece Introducer Kit with Hemostasis Valve. Right Superior Reorder #407900. The product is packa
- Z-1426-04 · Aug 4, 2004 · TerminatedTelesheath Left Atrial Introducer System Right Superior. Telesheath Catheter Two-Piece Introducer Kit with Hemostasis Valve. Left Lateral Reorder #407901. The product is package
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page