8870 software application card versions AAA 02, BBB 04, BBC… recallZ-1334-04 · Medtronic Inc, Neurological & Spinal Division · initiated Aug 24, 2004

Terminated All recalls posted in September 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc, Neurological & Spinal Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

With the Model 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N'Vision Clinician Programmer, users may have mistakenly entered a periodic bolus interval into the minutes field, rather than the hours field, resulting in drug overdoses. This issue is limited to programming the SynchroMed and SynchroMed EL pumps.

FDA root cause category: Other

What the company is doing

Letters were distributed 8/24/2004 utilizing a courier with signed receipt upon delivery. The letters were signed for upon receipt at the consignee address, and the contents of the letter included a response form which can be faxed or mailed back to Medtronic. If any letters are returned as 'undeliverable', attempts will be made to telephone these consignees. A press release was issued on September 22, 2004 by Medtronic Inc. for all Version AAA 02 Model 8870 software application cards in the U.S.

Recall details

Recall number
Z-1334-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 24, 2004
Posted by the FDA
Sep 23, 2004
Recalling firm
Medtronic Inc, Neurological & Spinal Division
Quantity
3,218 cards distributed in US, plus 1,248 distributed outside of the US
Product code
LKK
FDA submissions
P860004
Distribution
These cards were distributed primarily in the USA, but also to countries in Europe, Middle East and Africa, Canada, Australia, Hong Kong, Latin America and Japan.

Product description

8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N''Vision Clinician Programmer.

Code information (models, lots, serial numbers)

Software Version AAA 02, BBB 04, BBC 02, and BBD 01

Other devices under the same FDA submission

51 devices cleared under P860004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.