Model 8561 Lioresal (Baclofen injection) Intrathecal Refill… recallZ-0391-05 · Medtronic Inc, Neurological & Spinal Division · initiated Nov 19, 2004
Terminated All recalls posted in January 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc, Neurological & Spinal Division or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Outer package of Lioresal Intrethecal Refill Kit, model 8561, identifies the kit to contain 1-20 ml ampule of baclofen injection for intrathecal administration, 500 mcg/ml (10 mg/20 ml). Lot N0016487 incorrectly contains 1-20 ml ampule of baclofen concentration 2000 mcg/ml (40 mg/20 ml).
FDA root cause category: Other
What the company is doing
Customers were contacted 11/19/04 via telephone. Customers were asked to set aside product and return the affected lot to Medtronic, Inc., Moundsview, MN. Recalled product will be replaced by Medtronic.
Recall details
- Recall number
- Z-0391-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 19, 2004
- Posted by the FDA
- Jan 13, 2005
- Recalling firm
- Medtronic Inc, Neurological & Spinal Division
- Quantity
- 96 kits were distributed to 34 customers
- Product code
- LKK
- FDA submissions
- P860004
- Distribution
- within the United States only
Product description
Model 8561 Lioresal (Baclofen injection) Intrathecal Refill Kit is designed for refilling the Medtronic SynchroMed family of pumps. The 8561 kit contains a 1-20 ml ampule of 500 mcg/ml Baclofen injection along with a sterile refill kit tray and sterile drug preparation tray. Lioresal drug ampules contained witin Medtronic Lot N0016487 bear the primary manufacturer''s label and lot number 007J2583. The drug box is labeled with Lot 017H1925. Medtronic packages a manufacturer''s labeled ampule into a protective drug box. A lot number and drug lot expiration date are printed on the outside of the drug box. The filled and labeled drug box is placed inside a larger kit assembly that contains other components necessary to complete a pump refill. Only the large kit assembly (''Lioresal Refill Kit'') will display the Medtronic lot number (N0016487). The Model 8561 refill kit does not have a product code for itself; it is approved under PMA P860004. PMA P860004 is classified as: pump infusion, implanted, programmable under product code LKK.
Code information (models, lots, serial numbers)
N0016487
Other devices under the same FDA submission
51 devices cleared under P860004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
- SynchroMed® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths1,920 reports
- SYNCHROMED™ III 8667-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 death1,937 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths1,766 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page