Equator Forced Air Convective Warming System, Reference… recallZ-0497-05 · Smiths Medical ASD, Inc. · initiated Dec 15, 2004
Terminated All recalls posted in February 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Warmer heater coil may come in contact with the filter housing, causing a small area of the plastic housing to melt
FDA root cause category: Other
What the company is doing
Smiths Medical ASD notified customers by letter issued Fed''X on 12/15/04. Accounts are requested not to use the device and return the unit if the the unit has not yet been placed into service. If the unit has been placed into service and operated for more than one hour then the problem will not occur. EQ-5000 unit(s) will be replaced.
Recall details
- Recall number
- Z-0497-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 15, 2004
- Posted by the FDA
- Feb 3, 2005
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 206 units
- Product code
- DWJ
- FDA submissions
- K011907
- Distribution
- Nationwide Foreign: Nederland BV, UK, Finland
Product description
Equator Forced Air Convective Warming System, Reference Number EQ 5000
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Serial No. 2004126, 10040116, 20020088, 20020125, 20030104, 20030105, 20030133, 20030141, 20030146, 20030194, 20030199, 20030254, 20030287, 20030312, 20030394, 20030401, 20030408, 20030428, 20030460, 20030512, 20030522, 20030535, 20030627, 20030665, 20030759, 20030791, 20030818, 20030960, 20031030, 20031034, 20031039, 20031049, 20031056, 20031105, 20031151, 20031164, 20031172, 20031271, 20031336, 20031414, 20031417, 20031418, 20031423, 20031444, 20031488, 20031550, 20031622, 20031632, 20031661, 20031669, 20031681, 20040007, 20040010, 20040014, 20040021, 20040104, 20040106, 20040115, 20040124, 20040127, 20040153, 20040162, 20040164, 20040166, 20040202, 20040203, 20040208, 20040227, 20040290, 20040377, 20040378, 20040379, 20040380, 20040381, 20040382, 20040383, 20040385, 20040386, 20040387, 20040388, 20040389, 20040391, 20040392, 20040393, 20040394, 20040395, 20040396, 20040397, 20040398, 20040399, 20040400, 20040401, 20040402, 20040406, 20040408, 20040409, 20040410, 20040412, 20040414, 20040415, 20040416, 20040419, 20040420, 20040422, 20040427, 20040429, 20040505, 20040715, 20040817, 20040996, 20041156, 20041157, 20041158, 20041159, 20041160, 20041161, 20041162, 20041163, 20041164, 20041165, 20041166, 20041167, 20041168, 20041169, 20041170, 20041171, 20041172, 20041173, 20041174, 20041175, 20041176, 20041177, 20041178, 20041179, 20041180, 20041181, 20041182, 20041183, 20041184, 20041185, 20041186, 20041187, 20041188, 20041189, 20041190, 20041191, 20041192, 20041193, 20041194, 20041195, 20041196, 20041197, 20041198, 20041199, 20041200, 20041201, 20041202, 20041203, 20041204, 20041205, 20041206, 20041207, 20041208, 20041209, 20041210, 20041211, 20041212, 20041213, 20041214, 20041215, 20041216, 20041217, 20041218, 20041219, 20041220, 20041221, 20041222, 20041223, 20041224, 20041225, 20041227, 20041228, 20041229, 20041230, 20041231, 20041232, 20041233, 20041234, 20041235, 20041236, 20041237, 20041238, 20041239, 20041240, 20041241, 20041242, 20041244, 20041245, 20041246, 20041247, 20041249, 20041250, 20041251, 20041253, 20041256, 20041257, 20041258, 20041259, 20041260, 20041261, 20041262, 20041263, 20041264, 20041265, 20041266, 20041267, 20041268, 20041269, 20041270, 20041271, 20041272, 20041273, 20041274, 20041275, 20041276, 20041277, 20041278, 20041279, 20041280, 20041281, 20041282, 20041283, 20041284, 20041285, 20041286, 20041287, 20041288, 20041289, 20041290, 20041291, 20041293, 20041294, 20041295, 20041296, 20041297, 20041298, 20041299, 20041300, 20041301, 20041302, 20041303, 20041304, 20041305, 20041306, 20041307, 20041308, 20041309, 20041310, 20041311, 20041312, 20041313, 20041314, 20041315, 20041316, 20041317, 20041318, 20041319, 20041320, 20041321, 20041322, 20041323, 20041324, 20041325, 20041326, 20041327, 20041364, 20041581, 29941255
Other devices under the same FDA submission
5 devices cleared under K011907 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Level 1 EQ-5000Icu Medical, Inc. · System, thermal regulating46 reports
- Level 1 SW5-HOSE-7Icu Medical, Inc. · System, thermal regulating7 reports
- Level 1 SW-2003Icu Medical, Inc. · System, thermal regulating1 report
- Level 1 EQ-5000-INT-115VIcu Medical, Inc. · System, thermal regulating1 report
- Level 1 EQ-5000HF-EN-115VIcu Medical, Inc. · System, thermal regulating1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page