Baxter ClearLink Intravenous (IV) Solution Sets and… recallZ-0845-05 · Baxter Healthcare Corp. · initiated Feb 28, 2005
Terminated All recalls posted in May 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.
FDA root cause category: Other
What the company is doing
Recall letters dated 2/28/05 were sent to the direct accounts via first class mail on the same date to the attention of the Director of Nursing. The accounts were informed of the increase in reports of ClearLink valve occlusion, requested to examine their inventory for the affected lots of product, and call Baxter Healthcare Center for Service at 1-888-229-001 to return the affected product for credit. The letters included a copy of the Technical Letter revised 5/4/04, outlining the correct preparation and use ClearLink valve product, including detecting valve re-knits.
Recall details
- Recall number
- Z-0845-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 28, 2005
- Posted by the FDA
- May 17, 2005
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 1,171,116 sets
- Product code
- FPA
- FDA submissions
- K003225, K791496, K961225, K962581, K984381
- Distribution
- Nationwide, including Puerto Rico and American Samoa, and internationally to Canada, Panama, Australia, New Zealand and Venezuela.
Product description
Baxter ClearLink Intravenous (IV) Solution Sets and Extension Sets that have a ClearLink luer activated valve for IV access, which allows the administration of medications/solutions with a needleless luer syringe; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Product code 2C6255, lot R04J12126; Product code 2C8428, lot R04J11318; Product code 2C8515, lot R04J12043; Product code 2C8519, lots R04J12076, R04J12159; Product code 2C8537, lots R04J11128, R04J11201, R04J12027, R04J12084, R04J12183, R04J13033, R04J13116, R04J13173, R04J14031, R04J14098, R04J14130, R04J14155, R04J15061; Product code 2C8546, lot R04J11177; Product code 2C8571, lot R04J14205; Product code 2C8606, lot R04J13025; Product code 2C8612, lot R04J12019; Product code 2C8632, lot R04J12258; Product code 2C8634, lot R04J13090; Product code 2C8819, lots UR241323, UR242560, UR242669; Product code 2C8864, lot UR240697; Product code 2C8865, lots UR241638, UR241786, UR244012; Product code A2C9801, lots UR244020, UR245647; Product code 2H8401, lot R04J12068; Product code 2H8519, lots R04J15053, R04J16010; Product code 2N8371, lot UR245027; Product code 2N8373, lot UR243501; Product code 2N8374, lots UR229138, UR229526, UR230193, UR230193X, UR230433, UR230458, UR240747, UR241059, UR241356, UR241505, UR242065, UR246033; Product code 2N8378, lots UR229344, UR229781, UR230060, UR241133, UR241661, UR242107, UR242131, UR242362, UR242644, UR244632; Product code 2N8399, lots UR230128, UR230573, UR231027, UR231704, UR239301, UR239889, UR240200, UR240945, UR241265, UR242081, UR242230, UR243576;
Other devices under the same FDA submission
25 devices cleared under K003225, K791496, K961225, K962581, K984381 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Baxter Healthcare Corporation · Set, administration, intravascular2 recalls797 reports
- CLEARLINK/CONTINU-FLO 2C8537Baxter International Inc. · Set, administration, intravascular2 recalls490 reports
- CLEARLINK/CONTINU-FLO 2C8519Baxter Healthcare Corporation · Set, administration, intravascular3 recalls1 death221 reports
- CLEARLINK/CONTINU-FLO 2R8537Baxter International Inc. · Set, administration, intravascular1 recall136 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT UC8519Baxter International Inc. · Set, administration, intravascular1 recall66 reports
- CLEARLINK/CONTINU-FLO 2H8519Baxter Healthcare Corporation · Set, administration, intravascular93 reports
- CLEARLINK/CONTINU-FLO 2R8519Baxter International Inc. · Set, administration, intravascular1 recall37 reports
- CLEARLINK/CONTINU-FLO 2R8546Baxter International Inc. · Set, administration, intravascular59 reports
- CLEARLINK/DUO-VENT 2C7462Baxter Healthcare Corporation · Set, administration, intravascular57 reports
- DUO-VENT 1C8507Baxter Healthcare Corporation · Set, administration, intravascular1 recall4 reports
- CLEARLINK 2C7461Baxter Healthcare Corporation · Set, administration, intravascular49 reports
- CLEARLINK/DUO-VENT 2H7462Baxter International Inc. · Set, administration, intravascular47 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page