Product is an Orthopedic device contained in a Peel Foil… recallZ-1551-05 · Linvatec Corp. · initiated Feb 22, 2005

Terminated All recalls posted in September 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Linvatec Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised sterility of the devices.

FDA root cause category: Other

What the company is doing

Linvatec will issue the recall notifications to all Linvatec direct consignees and customers by mailings using UPS overnight delivery or equivalent method.

Recall details

Recall number
Z-1551-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 22, 2005
Posted by the FDA
Sep 21, 2005
Recalling firm
Linvatec Corp.
Quantity
3,273 Units
FDA submissions
K013131
Distribution
Product was shipped to consignees and customers (sales representatives, domestic customers, international customers, and international distributors and affiliates).

Product description

Product is an Orthopedic device contained in a Peel Foil Pouch. Product is labeled in part: ''REF***QTY 1 BioScrew XtraLok ***STERILE***Linvatec, LINVATEC CORPORATION LARGO, FL 33773 USA***''.

Code information (models, lots, serial numbers)

Lot #''s BBB32078, BBB34094, BBB34999, BBB35880, BBB38626, BBB48838, BBB49880, BBB59210, BBB53860, BBB55923, BBB59491, BBB65474, BBB66408, BBB68488, BBB31428, BBB34095, BBB35002, BBB36673, BBB48839, BBB51999, BBB55022, BBB59493, BBB66410, BBB35055, BBB50780, BBB61023, BBB34089, BBB35879, BBB37714, BBB48840, BBB52911, BBB55921, BBB65475, BBB30602, BBB31323, BBB49881, BBB52912, BBB67472, BBB52000.

Other devices under the same FDA submission

1 device cleared under K013131 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates