Medical device recalls: September 2005
The FDA posted 140 medical device recall records in September 2005, from 39 recall events by 34 companies. 140 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Baxter Healthcare Corp. · 3 events, 5 records
- North Coast Medi-Tek Inc · 3 events, 3 records
- Boston Scientific Corporation · 2 events, 63 records
- Becton Dickinson Medical Systems · 1 event, 9 records
- Angiomed Gmbh · 1 event, 7 records
- Scanlan International Inc · 1 event, 7 records
- Linvatec Corp. · 1 event, 5 records
- Fischer Imaging Corporation · 1 event, 4 records
FDA root cause
- Other · 136
- Labeling design · 2
- Nonconforming Material/Component · 1
- Device Design · 1
Not yet classified (140)
The FDA has posted these records but has not published a class for them yet.
- Z-1596-05 · Lifescan Inc · posted Sep 29, 2005 · Terminated
LifeScan brand OneTouch Ultra Meter - Owner's Booklets are missing instructions on how to set the Units of Measurement (UOM)--mg/dL versus mmol/L. This error has the potential to lead to acute hyperglycemic complications if the user fails to realize that…
- Z-1459-05 · Boston Scientific Corporation · posted Sep 28, 2005 · Terminated
During surgery, several reports of the grafts fraying and tearing upon suturing by the surgeon.
- Z-1589-05 · Cryolife Inc · posted Sep 28, 2005 · Terminated
This recall was initiated as part of the firm's retrospection review of tissues associated with complaints of alleged recipient infection.
- Z-1590-05 · Cryolife Inc · posted Sep 28, 2005 · Terminated
This recall was initiated as part of the firm's retrospection review of tissues associated with complaints of alleged recipient infection.
- Z-1591-05 · Cryolife Inc · posted Sep 28, 2005 · Terminated
This recall was initiated as part of the firm's retrospection review of tissues associated with complaints of alleged recipient infection.
- Z-1592-05 · Baxter Healthcare Corp. · posted Sep 28, 2005 · Terminated
Potential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
- Z-0187-06 · Guidant Corporation · posted Sep 24, 2005 · Terminated
A failure mode occurring at a low rate, has been identified within the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include: Intermittent or permanent loss of pacing output without warning, intermittent or…
- Z-0188-06 · Guidant Corporation · posted Sep 24, 2005 · Terminated
A failure mode occurring at a low rate, has been identified within the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include: Intermittent or permanent loss of pacing output without warning, intermittent or…
- Z-0189-06 · Guidant Corporation · posted Sep 24, 2005 · Terminated
A failure mode occurring at a low rate, has been identified within the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include: Intermittent or permanent loss of pacing output without warning, intermittent or…
- Z-0190-06 · Guidant Corporation · posted Sep 24, 2005 · Terminated
A failure mode occurring at a low rate, has been identified within the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include: Intermittent or permanent loss of pacing output without warning, intermittent or…
- Z-1577-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1578-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1579-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1580-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1581-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1582-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1583-05 · Angiomed Gmbh · posted Sep 24, 2005 · Terminated
Biliary Stents may malfunction and cause failure to expand, fracture, incomplete deployment, inaccurate positioning, migration, or lumen perforation.
- Z-1584-05 · Cook Urological, Inc. · posted Sep 24, 2005 · Terminated
Mislabeled as to size.
- Z-1585-05 · Bard Peripheral Vascular Inc · posted Sep 24, 2005 · Terminated
Angiomed Material Review Board reviewing the manufacturing process observed an increase in the internal scrap rate (failure mode: broken stent struts) of nitinol stents and determined the root cause to be a manufacturing change from polymer to aluminum…
- Z-1586-05 · Guidant Endovascular Solutions, Inc. · posted Sep 24, 2005 · Terminated
This action is being taken as a preventative measure as a result of an increased rate of complaints associated with stents becoming loose or dislodged from the delivery catheter.
- Z-1587-05 · ela Medical Llc · posted Sep 24, 2005 · Terminated
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
- Z-1597-05 · ela Medical Llc · posted Sep 24, 2005 · Terminated
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
- Z-1598-05 · ela Medical Llc · posted Sep 24, 2005 · Terminated
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
- Z-1543-05 · Baxter Healthcare Corp. · posted Sep 23, 2005 · Terminated
Swelling of the sealed lead-acid batteries in the infusion pump can cause internal pump damage, and excessive battery discharge can damage the batteries if the pump is left on battery power for an extended period of time after teh Battery Depleted alarm…
- Z-1544-05 · Baxter Healthcare Corp. · posted Sep 23, 2005 · Terminated
Swelling of the sealed lead-acid batteries in the infusion pump can cause internal pump damage, and excessive battery discharge can damage the batteries if the pump is left on battery power for an extended period of time after teh Battery Depleted alarm…
- Z-1547-05 · Linvatec Corp. · posted Sep 21, 2005 · Terminated
Linvatec has determined that there is a possibility that the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for…
- Z-1548-05 · Linvatec Corp. · posted Sep 21, 2005 · Terminated
Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised…
- Z-1549-05 · Linvatec Corp. · posted Sep 21, 2005 · Terminated
Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised…
- Z-1550-05 · Linvatec Corp. · posted Sep 21, 2005 · Terminated
Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised…
- Z-1551-05 · Linvatec Corp. · posted Sep 21, 2005 · Terminated
Linvatec has determined there is a possibility the pouch Tyvek header to foil longitudinal seal was not properly sealed by Bemis Flexible Packaging, in all pouches, prior to sterilization by Linvatec. Therefore there is a potential for compromised…
- Z-1552-05 · Baxter Healthcare Corp. · posted Sep 21, 2005 · Terminated
The mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.
- Z-1553-05 · Baxter Healthcare Corp. · posted Sep 21, 2005 · Terminated
The mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.
- Z-1554-05 · Stryker Howmedica Osteonics Corp. · posted Sep 21, 2005 · Terminated
The threaded part of the instrument may break when connected to the screw and potentially result in an adverse health consequence.
- Z-1555-05 · Hill-Rom Manufacturing, Inc. · posted Sep 21, 2005 · Terminated
Sudden unexpected lowering of the head of the bed may occur due to failure of the had of the bed actuator.
- Z-1556-05 · Cordis Corporation · posted Sep 21, 2005 · Terminated
During competitive product testing, wire movement difficulties were observed in Powerflex P3 units. Due to the possibility of user inconvenience, Cordis is recalling the six affected lots of Powerflex P3. There is no patient impact.
- Z-1557-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1558-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1559-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1560-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1561-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1562-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1563-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1564-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1565-05 · Becton Dickinson Medical Systems · posted Sep 21, 2005 · Terminated
Actuating tab may detach from the stopcock body, allowing fluid leak.
- Z-1566-05 · Encore Medical, Lp · posted Sep 21, 2005 · Terminated
Incorrect alignment markings or no markings at all on a reverse shoulder prosthesis unit.
- Z-1567-05 · Datascope Corp · posted Sep 21, 2005 · Terminated
The software version contains an anomaly which could freeze the unit when the right click button is used on the mouse.
- Z-1568-05 · C.R. Bard, Inc., Urological Division · posted Sep 21, 2005 · Terminated
An incorrect Y-connector was placed in the kit.
- Z-1571-05 · North Coast Medi-Tek Inc · posted Sep 21, 2005 · Terminated
Customer observed unsealed packages.
- Z-1572-05 · North Coast Medi-Tek Inc · posted Sep 21, 2005 · Terminated
Outside box labeling, containing 10 trays, incorrectly identified the product as ultrasound trays. Product is Breast Biopsy trays, Cat. 255826R2. Individual tray package labeling correctly identifies the product as Breast Biopsy tray.
- Z-1573-05 · Enpath Medical, Inc. · posted Sep 21, 2005 · Terminated
Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole.
- Z-1574-05 · Power Surgical Solutions · posted Sep 21, 2005 · Terminated
Potential compromise of sterile packaging for disposable dissecting tools.
- Z-1575-05 · Alcon Research, Ltd · posted Sep 21, 2005 · Terminated
Unknown residue noted on suface of flap irrigators incorporated into cusom paks.
- Z-1546-05 · CHdiagnostics · posted Sep 15, 2005 · Terminated
A suspect material used in the manufacturer of the product has degraded over time and consequently may cause falsely high readings when testing your blood glucose level.
- Z-1398-05 · Avid Medical Inc · posted Sep 14, 2005 · Terminated
Defect in packaging of medical devices may compromise sterility of these products used in medical and surgical procedures.
- Z-1539-05 · Abbott Laboratories MPG · posted Sep 14, 2005 · Terminated
The XSYSTEMS Multiconstituent Controls accessory insert (commodity number 34-3067/R6) does not contain the correct control ranges for the AxSYM Opiates assay.
- Z-1540-05 · Trek Diagnostic Systems · posted Sep 14, 2005 · Terminated
Invitro diagnostic test kit used for testing to confirm the presence of tuberculosis contains interfering bioburden which can cause false positive results.
- Z-1541-05 · Boston Scientific Corp · posted Sep 14, 2005 · Terminated
Firm has received complaints of extended ablation times, lack of roll-off and patient burns.
- Z-1542-05 · Boston Scientific Corp · posted Sep 14, 2005 · Terminated
Firm has received complaints of extended ablation times, lack of roll-off and patient burns.
- Z-1532-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1533-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1534-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1535-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1536-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1537-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1538-05 · Scanlan International Inc · posted Sep 13, 2005 · Terminated
Some scissors purchased before May 1, 2005 may not perform properly. They may have a crack in the blade that could break.
- Z-1531-05 · Linvatec Imaging · posted Sep 10, 2005 · Terminated
Through the complaint handling system, the company has determine there are performance issued involving the LS7500, 300 Watt Xenon Light Source. The performance issues result in a low light output that can create poor illumination of the surgical site.
- Z-1524-05 · Fischer Imaging Corporation · posted Sep 9, 2005 · Terminated
Belt driving the Image Intensifier may break, causing the Intensifier to slip down, possibly causing injury to patient.
- Z-1525-05 · Fischer Imaging Corporation · posted Sep 9, 2005 · Terminated
Belt driving the Image Intensifier may break, causing the Intensifier to slip down, possibly causing injury to patient.
- Z-1526-05 · Fischer Imaging Corporation · posted Sep 9, 2005 · Terminated
Belt driving the Image Intensifier may break, causing the Intensifier to slip down, possibly causing injury to patient.
- Z-1527-05 · Fischer Imaging Corporation · posted Sep 9, 2005 · Terminated
Belt driving the Image Intensifier may break, causing the Intensifier to slip down, possibly causing injury to patient.
- Z-1528-05 · Siemens Medical Solutions USA, Inc · posted Sep 9, 2005 · Terminated
there is a misalignment of the patient table that may occur between the patient table and the opening to the magnet bore
- Z-1529-05 · Siemens Medical Solutions USA, Inc · posted Sep 9, 2005 · Terminated
there is a misalignment of the patient table that may occur between the patient table and the opening to the magnet bore
- Z-1530-05 · Varian Medical Systems Inc · posted Sep 9, 2005 · Terminated
Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.
- Z-1523-05 · North Coast Medi-Tek Inc · posted Sep 8, 2005 · Terminated
Product contains the incorrect lot number.
- Z-0612-05 · Biomerieux, Inc. · posted Sep 7, 2005 · Terminated
A Bac T/ALERT 3D system failed to detect a positive sample in a pediatric BAC T/ALERT PF bottle.
- Z-0613-05 · Biomerieux, Inc. · posted Sep 7, 2005 · Terminated
A Bac T/ALERT 3D system failed to detect a positive sample in a pediatric BAC T/ALERT PF bottle.
- Z-0614-05 · Biomerieux, Inc. · posted Sep 7, 2005 · Terminated
A Bac T/ALERT 3D system failed to detect a positive sample in a pediatric BAC T/ALERT PF bottle.
- Z-1461-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1462-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1463-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1464-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1465-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1466-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1467-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1468-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1469-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1470-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1471-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1472-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1473-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1474-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1475-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1476-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1477-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1478-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1479-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1480-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1481-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1482-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1483-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1484-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1485-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1486-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1487-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1488-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1489-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1490-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1491-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1492-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1493-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1494-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1495-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1496-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1497-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1498-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1499-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1500-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1501-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1502-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1503-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1504-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1505-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1506-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1507-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1508-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1509-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1510-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1511-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1512-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1513-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1514-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1515-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1516-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1517-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1518-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1519-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1520-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1521-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1522-05 · Boston Scientific Corporation · posted Sep 2, 2005 · Terminated
The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.
- Z-1460-05 · Ortho-Clinical Diagnostics · posted Sep 1, 2005 · Terminated
Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.