Skip to content

Advisor Vital Signs Monitor (model 9200) with Capnography… recallZ-0830-05 · Smiths Medical PM, Inc. · initiated Mar 16, 2005

Terminated All recalls posted in May 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.

FDA root cause category: Other

What the company is doing

On March 16, 2005, customers were emailed Safety Action Bulletin 05-003 informing them of the low CO2 (End Tidal and Inspired) measurements in percent (%) and urging them to use mmHg or kPa as the unit of measurement . In addition, a Technical Worksheet will be distributed to customers who request a software upgrade. The Technical Worksheet will instruct them how to perform the software upgrade which corrects the CO2 measurements (End Tidal and Inspired) in percent (%).

Recall details

Recall number
Z-0830-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 16, 2005
Posted by the FDA
May 12, 2005
Recalling firm
Smiths Medical PM, Inc.
Quantity
6
Product code
DPS
FDA submissions
K010770, K011177, K030407, K982279
Distribution
CA and NY within US. Mexico, Argentina, Columbia, Chile, Brazil, Ecuador, Guatemala and Pakistan.

Product description

Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925654325.

Code information (models, lots, serial numbers)

serial no. 706330001, 706330002, 706330003, 706330004, 706330005 and 706431007.

Other devices under the same FDA submission

2 devices cleared under K010770, K011177, K030407, K982279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 31505).

Sources and updates