Advisor Vital Signs Monitor (model 9200) with Capnography… recallZ-0833-05 · Smiths Medical PM, Inc. · initiated Mar 16, 2005
Terminated All recalls posted in May 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
FDA root cause category: Other
What the company is doing
On March 16, 2005, customers were emailed Safety Action Bulletin 05-003 informing them of the low CO2 (End Tidal and Inspired) measurements in percent (%) and urging them to use mmHg or kPa as the unit of measurement . In addition, a Technical Worksheet will be distributed to customers who request a software upgrade. The Technical Worksheet will instruct them how to perform the software upgrade which corrects the CO2 measurements (End Tidal and Inspired) in percent (%).
Recall details
- Recall number
- Z-0833-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 16, 2005
- Posted by the FDA
- May 12, 2005
- Recalling firm
- Smiths Medical PM, Inc.
- Quantity
- 15
- Product code
- DPS
- FDA submissions
- K010770, K011177, K030407, K982279
- Distribution
- CA and NY within US. Mexico, Argentina, Columbia, Chile, Brazil, Ecuador, Guatemala and Pakistan.
Product description
Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925754335.
Code information (models, lots, serial numbers)
serial no. 706331000, 706331001, 706331002, 706331003, 706331004, 706331005, 706331006, 706331007, 706331008, 706331009, 706331010, 706331011, 706331012, 706331013 and 706331014
Other devices under the same FDA submission
2 devices cleared under K010770, K011177, K030407, K982279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BCI WW1100Icu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 reports
- BCI 920654230AIcu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 31505).
- Z-0824-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 920654225.
- Z-0825-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 920654335.
- Z-0826-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925474325.
- Z-0827-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925554325.
- Z-0828-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925574325.
- Z-0829-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925654220.
- Z-0830-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925654325.
- Z-0831-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925674320.
- Z-0832-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925674335.
- Z-0834-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925774325.
- Z-0835-05 · TerminatedAdvisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 92E654325.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page