Depuy P.F.C. SIGMA stabilized insert recallZ-1183-05 · Depuy Orthopaedics, Inc. · initiated May 26, 2005
Terminated All recalls posted in August 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Depuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
FDA root cause category: Other
Recall details
- Recall number
- Z-1183-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 26, 2005
- Posted by the FDA
- Aug 12, 2005
- Recalling firm
- Depuy Orthopaedics, Inc.
- Quantity
- 12
- Product code
- JWH
- FDA submissions
- K033272
- Distribution
- Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
Product description
Depuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 3, 17.5 mm; Ref 1581-23-017.
Code information (models, lots, serial numbers)
Lots Y45KN4 and Y5CVW4.
Other devices under the same FDA submission
65 devices cleared under K033272 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- P.F.C. SIGMA 1581-22-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER23 reports
- P.F.C. SIGMA 1581-23-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER21 reports
- P.F.C. SIGMA 1581-24-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER18 reports
- P.F.C. SIGMA 1581-14-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- P.F.C. SIGMA 1581-14-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- P.F.C. SIGMA 1581-13-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER15 reports
- P.F.C. SIGMA 1581-23-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER14 reports
- P.F.C. SIGMA 1581-12-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER13 reports
- P.F.C. SIGMA 1581-13-010Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER13 reports
- P.F.C. SIGMA 1581-24-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER13 reports
- P.F.C. SIGMA 1581-21-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER12 reports
- P.F.C. SIGMA 1581-22-008Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 32192).
- Z-1154-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 2, 8 mm; Ref 1581-11-008.
- Z-1155-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert , Sz 2.5, 8 mm; Ref 1581-12-008.
- Z-1156-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 2.5, 10 mm; Ref 1581-12-010.
- Z-1157-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 2.5, 17.5 mm; Ref 1581-12-017.
- Z-1158-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 8 mm; Ref 1581-13-008.
- Z-1159-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 10 mm; Ref 1581-13-010.
- Z-1160-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 15 mm; Ref 1581-13-015.
- Z-1161-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 17.5 mm; Ref 1581-13-017.
- Z-1162-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 20 mm; Ref 1581-13-020.
- Z-1163-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4, 8 mm; Ref 1581-14-008.
- Z-1164-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4,10 mm; Ref 1581-14-010.
- Z-1165-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4, 15 mm; Ref 1581-14-015.
- Z-1166-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4, 17.5 mm; Ref 1581-14-017.
- Z-1167-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 8 mm; Ref 1581-15-008.
- Z-1168-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 10 mm; Ref 1581-15-010.
- Z-1169-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 12.5 mm; Ref 1581-15-012.
- Z-1170-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 15 mm; Ref 1581-15-015.
- Z-1171-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 5, 17.5 mm; Ref 1581-15-017.
- Z-1172-05 · TerminatedDepuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 6, 12.5 mm; Ref 1581-16-012.
- Z-1173-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 1.5, 8 mm; Ref 1581-20-008.
- Z-1174-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 1.5, 15 mm; Ref 1581-20-015.
- Z-1175-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2, 15 mm; Ref 1581-21-015.
- Z-1176-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2, 20 mm; Ref 1581-21-020.
- Z-1177-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2.5, 8 mm; Ref 1581-22-008.
- Z-1178-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2.5, 10 mm; Ref 1581-22-010.
- Z-1179-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 2.5, 12.5 mm; Ref 1581-22-012.
- Z-1180-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 3, 10 mm; Ref 1581-23-010.
- Z-1181-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 3, 12.5 mm; Ref 1581-23-012.
- Z-1182-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 3, 15 mm; Ref 1581-23-015.
- Z-1184-05 · TerminatedDepuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 4, 8 mm; Ref 1581-24-008.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page