Datex-Ohmeda S/5 Aespire Anesthesia System 7900 recallZ-0038-06 · Datex-Ohmeda Inc · initiated Sep 27, 2005

Terminated All recalls posted in October 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datex-Ohmeda Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.

FDA root cause category: Other

What the company is doing

Urgent Medical Device Correction letter dated 09/23/05 was sent to all customers. The letter describes the issues, affected device serial numbers and recommends that customers DO NOT USE the internal auxiliary oxygen flowmeter, if installed in their system, and to use alternate auxiliary oxygen sources, i.e. wall-mounted oxygen flow meter, in the interim. A GE service representative will contact the customers to schedule an upgrade, performed at no charge.

Recall details

Recall number
Z-0038-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 27, 2005
Posted by the FDA
Oct 12, 2005
Recalling firm
Datex-Ohmeda Inc
Quantity
16
Product code
BSZ
FDA submissions
K050626
Distribution
CA, CO, FL, GA, IL, IN, MI, MS, NM, OK, PA, PR, SD, UT & WA. OUS to: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Canada, Caribbean, Czech Republic, Chile, China, Colombia, Dominican Republic, Ecuador, Finland, France, Germany, Ghana, Great Britain, Hong Kong, India, Indonesia, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Oman, Palestine, Peru, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Arab Emerites, Venezuela & Vietnam.

Product description

Datex-Ohmeda S/5 Aespire Anesthesia System 7900

Code information (models, lots, serial numbers)

ANCJ00100 thru ANCJ00113, ANCJ00116, ANCJ00118

Other devices under the same FDA submission

1 device cleared under K050626 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.