Baxter Interlink System Continu-Flo Solution Set with… recallZ-0653-06 · Baxter Healthcare Corp. · initiated Dec 15, 2005
Terminated All recalls posted in March 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
FDA root cause category: Other
What the company is doing
Baxter sent Urgent Product Recall letters dated 8/3/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the microbore female luers that may crack during use, resulting in the possible under delivery or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter extended the recall to include additional lots and product codes via letter dated 12/15/05. The letters referenced the 8/3/05 letter, including a copy of that letter, and listed the affected lots of sets. The instructions for return of the product remained the same as that described in the 8/3/05 letter.
Recall details
- Recall number
- Z-0653-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 15, 2005
- Posted by the FDA
- Mar 24, 2006
- Recalling firm
- Baxter Healthcare Corp.
- Quantity
- 1,920 sets
- Product code
- FMG
- FDA submissions
- K022895, K921899, K961225, K962581
- Distribution
- Nationwide, including Puerto Rico and American Samoa, and internationally to Canada, Australia, Japan, South Korea, Saudi Arabia, India, Mexico, Venezuela, Malaysia, East Timor and New Zealand.
Product description
Baxter Interlink System Continu-Flo Solution Set with Standard Bore 4-Way Stopcock Manifold Extension Set and Minivolume Extension Set, product code 3C0078; An Rx sterile, nonpyrogenic fluid pathway; solution set: 74'', 60 drops/mL, with injection site and male Luer Lock adapter, manifold extension set: 29'', 4 mL Vol, male luer adapter, and minivolume extension set: 7'', 0.4 mL Vol with retractable T-connector; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.
Code information (models, lots, serial numbers)
product code 3C0078, lotsGR229013, GR229799
Other devices under the same FDA submission
28 devices cleared under K022895, K921899, K961225, K962581 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Baxter Healthcare Corporation · Set, administration, intravascular2 recalls797 reports
- CLEARLINK/CONTINU-FLO 2C8537Baxter International Inc. · Set, administration, intravascular2 recalls490 reports
- CLEARLINK/CONTINU-FLO 2C8519Baxter Healthcare Corporation · Set, administration, intravascular3 recalls1 death221 reports
- CLEARLINK/CONTINU-FLO 2R8537Baxter International Inc. · Set, administration, intravascular1 recall136 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT UC8519Baxter International Inc. · Set, administration, intravascular1 recall66 reports
- CLEARLINK/CONTINU-FLO 2H8519Baxter Healthcare Corporation · Set, administration, intravascular93 reports
- CLEARLINK/CONTINU-FLO 2R8519Baxter International Inc. · Set, administration, intravascular1 recall37 reports
- CLEARLINK/CONTINU-FLO 2R8546Baxter International Inc. · Set, administration, intravascular59 reports
- CLEARLINK/DUO-VENT 2C7462Baxter Healthcare Corporation · Set, administration, intravascular57 reports
- DUO-VENT 1C8507Baxter Healthcare Corporation · Set, administration, intravascular1 recall4 reports
- CLEARLINK 2C7461Baxter Healthcare Corporation · Set, administration, intravascular49 reports
- CLEARLINK/DUO-VENT 2H7462Baxter International Inc. · Set, administration, intravascular47 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 34290).
- Z-0626-06 · TerminatedBaxter Non-DEHP Y-Type Catheter Extension Set, product code 2N1191; An Rx sterile, nonpyrogenic fluid pathway, 4.8', 0.6 mL Vol., with male Luer Lock adapter; Baxter Healthcare Cor…
- Z-0629-06 · TerminatedBaxter Interlink System Non-DEHP Standard Bore 3-Lead Catheter Extension Set, product code 2N3335; An Rx sterile, nonpyrogenic fluid pathway with 2 Interlink injection sites and a …
- Z-0630-06 · TerminatedBaxter Interlink System T-Connector Extension Set, product code 2N3343; An Rx sterile, nonpyrogenic fluid pathway, 7.0'', 1.0 mL Vol., with 1 Interlink T-connector injection sites…
- Z-0634-06 · TerminatedBaxter Interlink System Non-DEHP I.V. Catheter Extension Set, product code 2N3384; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.5 mL Vol., with injection site and male Luer …
- Z-0637-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8290; An Rx sterile, nonpyrogenic fluid pathway, 11.0'', 0.86 mL Vol., with 2 injection sites and male Luer Lock ad…
- Z-0638-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8378; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.6 mL Vol., with injection site and male Luer Lock adapter…
- Z-0640-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8412; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 1.0 mL Vol., with injection site, 0.22 Micron Downstream F…
- Z-0641-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8462; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 4.2 mL Vol., with 2 injection sites, 0.22 Micron Downstrea…
- Z-0643-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8590; An Rx sterile, nonpyrogenic fluid pathway, 22'', 0.9 mL Vol, with injection site and Stationary Swivel Lock T…
- Z-0644-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8643; An Rx sterile, nonpyrogenic fluid pathway, 7'', 0.3 mL Vol, with Male Luer Slip T-Connector; Baxter Healthcar…
- Z-0646-06 · TerminatedBaxter Interlink System Extension Set with Standard Bore 4-Way Stopcock Manifold Extension Set and Minivolume Extension Set, product code 3C0071; An Rx sterile, nonpyrogenic fluid …
- Z-0648-06 · TerminatedBaxter Interlink System Y-Type Minivolume Extension Set, product code 1C8645; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.8 mL Vol., with 2 injection sites and male Luer Lo…
- Z-0652-06 · TerminatedBaxter Interlink System Minivolume Extension Set, product code 1C8740; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 1.0 mL Vol., with injection site, 0.22 Micron Downstream F…
- Z-0654-06 · TerminatedBaxter Interlink System Buretrol Solution Set with Large Bore 3-Way Stopcock with Rotating Male Luer Lock Extension Set and Minivolume Extension Set, product code 3C0108; An Rx ste…
- Z-0656-06 · TerminatedBaxter Interlink System Buretrol Solution Set with Large Bore 4-Way Stopcock with Rotating Male Luer Lock Extension Set and Extension Set, product code 3C0172; An Rx sterile, nonp…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page