The unimplanted HiRes90K Implantable Cochlear Stimulators… recallZ-0759-06 · Advanced Bionics Corporation · initiated Mar 8, 2006
Terminated All recalls posted in April 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Bionics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain HiRes 90K cochlear implants could fail as a result of elevated moisture levels. The unimplanted HiRes 90K implants listed are being removed from the market. Patients are to be followed for signs of device failure: intermittent function, complete loss of sound, sudden discomfort, pain, noise, or popping.
FDA root cause category: Other
What the company is doing
The recall was initiated on 03/08/06. In North America, the clinician notification letters were distributed by fax. For other regions, an Advanced Bionics representative and/or distributor are providing notification to the affected clinicians. As part of the clinician notification process, each center was provided with a list of their affected devices. Customers were given a list of HiRes 90K implants with feedthrus from Supplier B that were shipped to their clinic and are not yet registered with Advanced Bionics. Customers were asked to return the implants listed in the attachment to Advanced Bionics according to the instructions. Customers were told not to implant any of the devices on the list. Customers were told that implants manufactured wth feedthrus from Supplier A will be delivered quickly to minimize disruptions to their schedule. Customers will be contacted through Advanced Bionics Customer Service within 48 hours to arrange retrieval of implants manufactured with feedthrus from Supplier B and replacement with implants manufactured with feedthrus from Supplier A.
Recall details
- Recall number
- Z-0759-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 8, 2006
- Posted by the FDA
- Apr 11, 2006
- Recalling firm
- Advanced Bionics Corporation
- Quantity
- 613 units for both models (186 in the U.S.), (427 ROW)
- Product code
- MCM
- FDA submissions
- P960058
- Distribution
- Nationwide to: AR, AL, FL, TX, NY, NE, NC, CO, GA, OH, IL, PA, IN, MN, NC, CA, HI, MA, AZ, VA, OR, SD, MI, MO, IA, UT, TN. Worldwide to: Argentina, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Cuba, Denmark, Ecuador, Egypt, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherlands, Norway, Poland, Portugal, Puerto Rico, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Venezuela.
Product description
The unimplanted HiRes90K Implantable Cochlear Stimulators (ICS) manufactured with feedthrus from a specific supplier are being removed from the market. Model number CI-1400-02H.
Code information (models, lots, serial numbers)
Model CI-1400-02H. Serial numbers: 350550, 350974, 350386, 350613, 350654, 350687, 350714, 350754, 350859, 350964, 350998, 351006, 351031, 351035, 351063, 351129, 351132, 351139, 350568, 350121, 350486, 350514, 350581, 350655, 350896, 350955, 350958, 350967, 350452, 350816, 350949, 350959, 350963, 351021, 351140, 350489, 350492, 350505, 350555, 350789, 350910, 250914, 350957, 351036, 350565, 351020, 350251, 350283, 350482, 350509, 350519, 350524, 350525, 350527, 350531, 350532, 350583, 350676, 350689, 351046, 350535, 350220, 350224, 350239, 350246, 350250, 350254, 350256, 350260, 350294, 351069.
Other devices under the same FDA submission
25 devices cleared under P960058 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HiRes™ Ultra CI HiFocus™ MS Electrode CI-1600-04Advanced Bionics AG · Implant, Cochlear2,918 reports
- HiRes™ Ultra 3D CI HiFocus™ MS Electrode CI-1601-04Advanced Bionics AG · Implant, Cochlear1 death931 reports
- HiRes™ Ultra 3D CI HiFocus™ SlimJ Electrode CI-1601-05Advanced Bionics AG · Implant, Cochlear795 reports
- HiRes™ Ultra CI HiFocus™ SlimJ Electrode CI-1600-05Advanced Bionics AG · Implant, Cochlear335 reports
- HiRes 90K™ Advantage CI HiFocus™ MS Electrode CI-1500-04Advanced Bionics AG · Implant, Cochlear315 reports
- HiRes 90K™ Advantage CI HiFocus™ 1j Electrode CI-1500-01Advanced Bionics AG · Implant, Cochlear280 reports
- HiRes™ Ultra 3D CI HiFocus™ SlimJ Electrode CI-1601-05Advanced Bionics, LLC · Implant, cochlear1 death257 reports
- HiRes™ Ultra 3D CI HiFocus™ SlimJ Electrode CI-1601-05Advanced Bionics AG · Implant, Cochlear254 reports
- HiRes™ Ultra 3D CI HiFocus™ MS Electrode CI-1601-04Advanced Bionics, LLC · Implant, cochlear1 recall149 reports
- HiRes™ Ultra 3D CI HiFocus™ MS Electrode CI-1601-04Advanced Bionics AG · Implant, Cochlear195 reports
- HiRes™ Ultra CI HiFocus™ SlimJ Electrode CI-1600-05Advanced Bionics AG · Implant, Cochlear178 reports
- HiRes 90K™ Advantage CI HiFocus™ MS Electrode CI-1500-04Advanced Bionics AG · Implant, Cochlear1 recall127 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 34873).
- Z-0758-06 · TerminatedThe unimplanted HiRes90K Implantable Cochlear Stimulators (ICS) manufactured with feedthrus from a specific supplier are being removed from the market. Model number CI-1400-01
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page