Guidant VENTAK PRIZM VR (model 1860) and VENTAK PRIZM DR… recallZ-1297-06 · Boston Scientific CRM Corp · initiated Jun 24, 2006
Terminated All recalls posted in September 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Guidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion in the affected devices.
FDA root cause category: Other
What the company is doing
A Press Release was issued on 06/26/06. An Urgent Medical Device Safety Information & Corrective Action letter, dated 06/23/06 was delivered to physicians describing preliminary information before Guidant has completed its own investigation and finalized patient care recommendations. Non-implanted inventory is being retrieved from hospital shelves and Guidant Represenatatives. Physicians are to perform an in-clinic follow-up exam as soon as possible with patients implanted with affected devices.
Recall details
- Recall number
- Z-1297-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 24, 2006
- Posted by the FDA
- Sep 28, 2006
- Recalling firm
- Boston Scientific CRM Corp
- Quantity
- 175
- Product code
- KRG
- FDA submissions
- P960040
- Distribution
- Worldwide- USA and Antigua Barbuda, Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Micronesia, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Peru, Philippines, Polynesia, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela
Product description
Guidant VENTAK PRIZM VR (model 1860) and VENTAK PRIZM DR (model 1861) Automatic Implantable Cardioverter Defibrillator (AICD), Dual Chamber, Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Quantities Distributed by model are as follows: VENTAK PRIZM VR model 1860 = 67; DR model 1861 = 108.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
model 1860, serial numbers: 144889, 144891, 144893 thru 144903, 144926, 144927, 144930, 144932, 144942, 144946, 144947, 144950, 144955 thru 144957, 144959, 144965 thru 144979, 144981, 144983, 144985 thru 144989, 144991, 144994, 145000 thru 145005, 145009, 145012, 145013, 145015 thru 145028, 145030 thru 145044, 145050 thru 145053, 145055, 145056, 145058 thru 145061, 145065 thru 145069, 145071 thru 145078. model 1861, serial numbers: 266473, 266475, 266477, 266478, 266480 thru 266483, 266485, 266487, 266488, 266492 thru 266494, 266496 thru 266506, 266510, 266511, 266516, 266518 thru 266521, 266523, 266525 thru 266527, 266532, 266533, 266535, 266536, 266553 thru 266557, 266561, 266576 thru 266578, 266580, 266582 thru 266587, 266608, 266617, 266632, 266637, 266641, 266663, 266665, 266668, 266671, 266676, 266684, 266685, 266694 thru 266696, 266699, 266702, 266705 thru 266712, 266714 thru 266728, 266732 thru 266741, 266744 thru 266749, 266753, 266764, 266769 thru 266775, 266777, 266779, 266780, 266782, 266784, 266786 thru 266794, 266796 thru 266798, 266802, 266803, 266805, 266807, 266812, 266820, 266823, 266824, 266829, 266832, 266833, 266835 thru 266837, 266839, 266841 thru 266843, 266849 thru 266857, 266859 thru 266863.
Other devices under the same FDA submission
45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- DYNAGEN™ EL ICD DR D152Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)12 deaths1,208 reports
- INOGEN™ EL ICD DR D142Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)6 deaths1,017 reports
- RESONATE™ EL ICD DR D433Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths965 reports
- DYNAGEN™ EL ICD DR D153Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths886 reports
- MOMENTUM™ EL ICD DR D121Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 deaths850 reports
- DYNAGEN™ EL ICD VR D150Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths831 reports
- INOGEN™ EL ICD VR D140Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths823 reports
- ENERGEN™ ICD E140Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)3 deaths819 reports
- VIGILANT™ EL ICD VR D232Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths750 reports
- ENERGEN™ ICD E142Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 death705 reports
- INCEPTA™ ICD E160Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 deaths601 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35754).
- Z-1298-06 · TerminatedCONTAK RENEWAL TR (models H120, H125) and CONTAK RENEWAL TR2 (models H140, H145). CONTAK RENEWAL TR2 devices are not available in the US. Quantities Distributed by model are as …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page