Medtronic Intrathecal Catheter, Model 8731, (part of… recallZ-1414-06 · Medtronic Neurological · initiated Jul 21, 2006
Terminated All recalls posted in September 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neurological or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Tip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.
FDA root cause category: Other
What the company is doing
An Urgent Worldwide Medical Device Recall letter was sent to affected customers 07/20/06 for delivery on 07/21/06. The letter describes the issue, identifies the affected product and list action required by the customer. Affected devices are to be removed from active inventory, quarantined and returned to Medtronic for a warranty credit. A Reply card is asked to be returned. Medtronic Representatives will follow-up by July 28, 2006.
Recall details
- Recall number
- Z-1414-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 21, 2006
- Posted by the FDA
- Sep 9, 2006
- Recalling firm
- Medtronic Neurological
- Quantity
- Recalls Z-1414-1415-06-USA - 359 devices. Foreign - 575 devices
- Product code
- LKK
- FDA submissions
- P860004
- Distribution
- Worldwide Distribution-USA and Austria, Belgium, Canada, Cyprus, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, and the United Kingdom
Product description
Medtronic Intrathecal Catheter, Model 8731, (part of infusion systems using Medtronic implantable pumps) The catheter is designed for use in the intrathecal space. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.
Code information (models, lots, serial numbers)
Model 8731 Base Lot Numbers: B011311N, B011421N, B011422N, B011500N, B011721N, B011737N, B011738N, B011786N, N0012097, N0012140, N0012141, N0012521, N0012692, N0012814, N0012815, N0012816, N0013149, N0013150, N0013352, N0013353, N0013354, N0013355, N0013707, N0013710, N0013936, N0014007, N0014236, N0014237, N0014330, N0014331, N0014332, N0014722, N0015050. *Model 8731 Catheters and Model 8598 Distal Revision kits have a Base Lot number of eight (8) characters plus a two (2) digit suffix that represents serialization within these recalled lots. This list provides just the Base lot Number (the first eight characters of the product Lot No. found on the product labeling).
Other devices under the same FDA submission
51 devices cleared under P860004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
- SynchroMed® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths1,920 reports
- SYNCHROMED™ III 8667-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 death1,937 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths1,766 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35931).
- Z-1415-06 · TerminatedMedtronic Intrathecal Catheter Distal Revision Kit, Model 8598, (provides replacement parts for the distal section of the 8731 Intrathecal Catheter) . Medtronic, Inc., 710 Medtron…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page