Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for… recallZ-1536-06 · Medtronic Neurological · initiated Aug 2, 2006
Terminated All recalls posted in September 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neurological or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Model 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled 3778 lot V009546 may contain a 3777 standard-spaced lead. The result of this error is an overall increase or decrease of 14 mm in the electrode coverage.
FDA root cause category: Other
What the company is doing
Medtronic representatives retrieved non-implanted devices. An Urgent Medical Device Recall letter was sent to physicians to describe the issue and patient management information.
Recall details
- Recall number
- Z-1536-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 2, 2006
- Posted by the FDA
- Sep 29, 2006
- Recalling firm
- Medtronic Neurological
- Quantity
- 22 total
- Product code
- LGW
- FDA submissions
- P840001
- Distribution
- Nationwide distribution --- including states of CA, GA, KY, LA, and TX.
Product description
Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis, MN 55432-5604.
Code information (models, lots, serial numbers)
Model 3777, Lot V009545
Other devices under the same FDA submission
116 devices cleared under P840001 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- RestoreSensor 37714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,116 reports
- PrimeAdvanced® SureScan® 97702Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death2,263 reports
- RestoreUltra® SureScan® 97712Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,887 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1,339 reports
- Vanta™ AdaptiveStim™ 977006Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief3 recalls1,151 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy1,125 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy635 reports
- RestoreAdvanced® SureScan® 97713Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF632 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief513 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page