St. Jude Medical Identity SR Model 5172, Identity DR Model… recallZ-0298-2007 · St Jude Medical CRMD · initiated Oct 6, 2006
Terminated All recalls posted in December 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St Jude Medical CRMD or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
St. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.
FDA root cause category: Other
What the company is doing
The notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on St. Jude's website, www.sim.com Completion of this advisory is estimated to be three months after its initiation.
Recall details
- Recall number
- Z-0298-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 6, 2006
- Posted by the FDA
- Dec 19, 2006
- Recalling firm
- St Jude Medical CRMD
- Quantity
- 199,065
- Product code
- DXY
- FDA submissions
- P880086
- Distribution
- Worldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN, SOUTH KOREA, LATVIA, LICHTENSTEIN, LUXEMBURG, MALAYSIA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PORTUGAL, SAUDI ARABIA, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, UNITED KINGDOM & VENEZUELA
Product description
St. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)
Code information (models, lots, serial numbers)
All Devices (all serial numbers for the devices referenced in the advisory.)
Other devices under the same FDA submission
38 devices cleared under P880086 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Zephyr™ XL DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable24 deaths944 reports
- Assurity™ PM1240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall20 deaths306 reports
- Victory™ 5816St. Jude Medical, Inc. · implantable pacemaker Pulse-generator5 deaths347 reports
- Endurity™ PM2160St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall2 deaths237 reports
- Zephyr® DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable9 deaths219 reports
- Accent™ DRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable20 deaths214 reports
- Accent™ SR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable12 deaths164 reports
- Zephyr™ XL SRSt. Jude Medical, Inc. · Pulse generator, permanent, implantable2 deaths161 reports
- Assurity™ PM2260St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall109 reports
- Identity™ 5386St. Jude Medical, Inc. · implantable pacemaker Pulse-generator3 deaths154 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 36619).
- Z-0296-2007 · TerminatedSt. Jude Medical APS III Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3500/3510
- Z-0297-2007 · TerminatedSt. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page