St. Jude Medical Identity SR Model 5172, Identity DR Model… recallZ-0298-2007 · St Jude Medical CRMD · initiated Oct 6, 2006

Terminated All recalls posted in December 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from St Jude Medical CRMD or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

St. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.

FDA root cause category: Other

What the company is doing

The notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on St. Jude's website, www.sim.com Completion of this advisory is estimated to be three months after its initiation.

Recall details

Recall number
Z-0298-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 6, 2006
Posted by the FDA
Dec 19, 2006
Recalling firm
St Jude Medical CRMD
Quantity
199,065
Product code
DXY
FDA submissions
P880086
Distribution
Worldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN, SOUTH KOREA, LATVIA, LICHTENSTEIN, LUXEMBURG, MALAYSIA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PORTUGAL, SAUDI ARABIA, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, UNITED KINGDOM & VENEZUELA

Product description

St. Jude Medical Identity SR Model 5172, Identity DR Model 5370, & Identity XL DR Model 5376; Pulse Generators (pacemakers)

Code information (models, lots, serial numbers)

All Devices (all serial numbers for the devices referenced in the advisory.)

Other devices under the same FDA submission

38 devices cleared under P880086 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 36619).