Skip to content

Cyberonics VNS Therapy System, Model 250-''HAND HELD''… recallZ-0341-2007 · Cyberonics, Inc · initiated Oct 17, 2006

Terminated All recalls posted in January 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cyberonics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician

FDA root cause category: Other

What the company is doing

Firm sent safety alert to physicians on 11/14/06. Safety Alert informs physicians of the problem and provides instructions on how to avoid an event and correct an event should it occur.

Recall details

Recall number
Z-0341-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 17, 2006
Posted by the FDA
Jan 25, 2007
Recalling firm
Cyberonics, Inc
Quantity
3,538 Flashcards containing software.
Product code
LYJ
FDA submissions
P970003
Distribution
Worldwide Distribution ---- USA and countries of Argentina, Australia, Austria, Belgium, Canada, Cyprus, Czech Rep, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, S. Korea, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, and Venezuela.

Product description

Cyberonics VNS Therapy System, Model 250-''HAND HELD'' programming software v6.1, v6.1.7, v7.0 and v7.1.3; manufactured by Cyberonics, Inc., Houston, TX.

Code information (models, lots, serial numbers)

For Software v6.1: 10-005-9400 and 10-006-0900; For Software v6.1.7: 10-0007-2400 and 10-0007-2500; For Software v7.0: 10-0007-4100 and 10-0007-2500; For Software v7.1.3: 10-0007-8400 and 10-0007-8500;

Other devices under the same FDA submission

23 devices cleared under P970003 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 36672).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page