ADVIA 1200 Chemistry System, Automated Clinical Chemistry… recallZ-0351-2007 · Bayer Healthcare, LLC (Diagnostics Division) · initiated Sep 8, 2006

Terminated All recalls posted in January 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bayer Healthcare, LLC (Diagnostics Division) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the customer was approximately -25%.

FDA root cause category: Software design

What the company is doing

On 9/08/06, a Support Bulletin was sent to all affected Bayer HealthCare LLC Branches in order to inform them of the issue and to provide them with instructions for the 'field correction'. On 9/08/06, a Customer Bulletin was sent to all affected Bayer HealthCare LLC Branches both in the US and worldwide for communication with the affected end user customers. The bulletin advises of the problem and the corrective action. A Completion Confirmation (Fax-back) Form was included to ensure that all customers worldwide have been notified and have complied with the mandatory action. The Customer Bulletin is to be sent to all affected end user customers. Bayer Technical Service Specialists will follow-up with all of the ADVIA Chemistry System customers to make sure they have understood the bulletin and to ask that they return the FAX-back Completion Confirmation Form to Bayer HealthCare. If Bayer HealthCare does not receive the form in a timely manner, specialists in the Bayer Technical Care Center will contact the customer through whatever means possible in an effort to obtain 100% effectiveness on this communication.

Recall details

Recall number
Z-0351-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 8, 2006
Posted by the FDA
Jan 17, 2007
Recalling firm
Bayer Healthcare, LLC (Diagnostics Division)
Quantity
1184 instruments total, all products
Product code
LEG
FDA submissions
K990346
Distribution
Worldwide, including USA, Puerto Rico, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Korea, Malaysia, Mexico, The Netherlands, New Zealand, Norway, Poland, Portugal, Russia, Serbia, Singapore, Slovakia, South Africa, Spain, Saudi Arabia, Sweden, Switzerland, Syria, Taiwan, Turkey, and United Kingdom.

Product description

ADVIA 1200 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests. The following are the measured parameters that are within the scope of this ''field correction'': Affected assay: Valproic Acid. Affecting assays: Digoxin; Phenytoin; Gentamicin; Theophylline.

Code information (models, lots, serial numbers)

All ADVIA 1200 Chemistry Systems are involved as well as all existing assay lots for Valproic Acid (VPA), Phenytoin (PHNY), Gentamicin (GENT), Digoxin (DIG), and Theophylline (THEO). ADVIA 1200 Part # 094-A001, VPA Part # B01-4784-01, PHNY Part # B01-4775-01, DIG Part # B01-4769-01, GENT Part # B01-4787-01, THEO Part # B01-4772-01.

Other devices under the same FDA submission

2 devices cleared under K990346 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 36713).