ADVIA 2400 Chemistry System, Automated Clinical Chemistry… recallZ-0353-2007 · Bayer Healthcare, LLC (Diagnostics Division) · initiated Sep 8, 2006
Terminated All recalls posted in January 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bayer Healthcare, LLC (Diagnostics Division) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the customer was approximately -25%. Investigation into this issue was initiated promptly following receipt of this complaint.
FDA root cause category: Software design
What the company is doing
On 9/08/06, a Support Bulletin was sent to all affected Bayer HealthCare LLC Branches in order to inform them of the issue and to provide them with instructions for the 'field correction'. On 9/08/06, a Customer Bulletin was sent to all affected Bayer HealthCare LLC Branches both in the US and worldwide for communication with the affected end user customers. The bulletin advises of the problem and the corrective action. A Completion Confirmation (Fax-back) Form was included to ensure that all customers worldwide have been notified and have complied with the mandatory action. The Customer Bulletin is to be sent to all affected end user customers. Bayer Technical Service Specialists will follow-up with all of the ADVIA Chemistry System customers to make sure they have understood the bulletin and to ask that they return the FAX-back Completion Confirmation Form to Bayer HealthCare. If Bayer HealthCare does not receive the form in a timely manner, specialists in the Bayer Technical Care Center will contact the customer through whatever means possible in an effort to obtain 100% effectiveness on this communication.
Recall details
- Recall number
- Z-0353-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 8, 2006
- Posted by the FDA
- Jan 17, 2007
- Recalling firm
- Bayer Healthcare, LLC (Diagnostics Division)
- Quantity
- 1184 instruments total, all products
- Product code
- LEG
- FDA submissions
- K990346
- Distribution
- Worldwide, including USA, Puerto Rico, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Korea, Malaysia, Mexico, The Netherlands, New Zealand, Norway, Poland, Portugal, Russia, Serbia, Singapore, Slovakia, South Africa, Spain, Saudi Arabia, Sweden, Switzerland, Syria, Taiwan, Turkey, and United Kingdom.
Product description
ADVIA 2400 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests. The following are the measured parameters that are within the scope of this ''field correction'': Affected assay: Valproic Acid. Affecting assays: Digoxin; Phenytoin; Gentamicin; Theophylline.
Code information (models, lots, serial numbers)
All ADVIA 2400 Chemistry Systems are involved as well as all existing assay lots for Valproic Acid (VPA), Phenytoin (PHNY), Gentamicin (GENT), Digoxin (DIG), and Theophylline (THEO). ADVIA 2400 Part # 073-A010, VPA Part # B01-4784-01, PHNY Part # B01-4775-01, DIG Part # B01-4769-01, GENT Part # B01-4787-01, THEO Part # B01-4772-01.
Other devices under the same FDA submission
2 devices cleared under K990346 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry XPT System 10723034Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall174 reports
- ADVIA® 1800 Analyzer 10315183Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE150 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 36713).
- Z-0351-2007 · TerminatedADVIA 1200 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests. The following are the measured parameters that are within the scope of th…
- Z-0352-2007 · TerminatedADVIA 1650 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests. The following are the measured parameters that are within the scope of …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page