Skip to content

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part… recallZ-1105-2007 · Smith & Nephew, Inc Endoscopy Division · initiated Oct 12, 2004

Terminated All recalls posted in August 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew, Inc Endoscopy Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.

FDA root cause category: Other

What the company is doing

Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.

Recall details

Recall number
Z-1105-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 12, 2004
Posted by the FDA
Aug 2, 2007
Recalling firm
Smith & Nephew, Inc Endoscopy Division
Quantity
274 units.
Product code
DWQ
FDA submissions
K032387
Distribution
Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.

Product description

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

Code information (models, lots, serial numbers)

Pnuematic Footswitch part of TRIVEX System with the following serial numbers: 105, 106, 108, 109, 111 -123, 125 - 182, 184 - 188, 192 - 197, 199 - 203, 205 - 219, 222 - 249, 252 - 254, and 7210386.

Other devices under the same FDA submission

3 devices cleared under K032387 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page