Mobilett XP, mobile X-ray system, Model number 1818454 recallZ-0749-2007 · Siemens Medical Solutions USA, Inc · initiated Mar 21, 2007
Terminated All recalls posted in April 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mounting bolts for tank fork assembly and collimator flange may become loose.
FDA root cause category: Other
What the company is doing
Siemens Service Personnel are visiting each consignee, beginning March 21, 2007, to replace the required mounting bolts and tighten them according to required torque specifications.
Recall details
- Recall number
- Z-0749-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 21, 2007
- Posted by the FDA
- Apr 19, 2007
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 186 units
- Product code
- IZL
- FDA submissions
- K033238
- Distribution
- The products were shipped to medical facilities nationwide.
Product description
Mobilett XP, mobile X-ray system, Model number 1818454
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Serial numbers 1109, 1153, 1205, 2128, 1075, 1315, 1316, 1335, 1130, 1281, 1282, 1176, 1339, 1224, 1301, 2008, 1021, 1154, 1180, 1181, 1009, 1189, 1211, 1212, 2003, 2010, 1341, 1342, 2049, 2077, 2108, 1314, 1334, 1192, 1213, 1043, 1126, 1135, 2047, 2090, 2123, 1024, 1031, 1034, 1035, 1111, 2004, 1113, 1178, 1240, 1261, 2040, 1058, 1115, 1159, 1203, 1247, 1309, 1345, 2032, 1010, 1049, 1151, 1297, 1298, 2051, 1206, 1218, 1238, 1332, 1333, 2034, 2035, 2065, 1138, 1299, 1300, 2020, 2021, 2022, 2052, 2017, 2027, 1086, 1029, 1145, 1027, 1164, 1234, 2115, 1110, 1152, 1156, 1235, 1089, 1022, 1241, 1254, 1286, 1073, 1079, 1084, 1131, 1165, 1208, 2028, 2120, 2124, 2132, 1037, 1045, 1185, 1251, 1252, 1259, 1294, 1295, 2048, 1025, 1087, 1146, 1173, 1186, 1187, 1198, 1199, 1258, 1275, 1307, 2009, 1042, 1060, 1076, 1092, 1228, 1229, 1230, 1231, 1239, 1246, 1078, 2080, 2109, 2113, 1088, 1103, 1049, 1040, 1055, 1059, 1090, 1177, 1204, 1210, 1214, 1216, 1263, 1284, 1285, 1289, 1290, 1306, 1308, 2029, 2041, 2042, 2075, 2099, 2110, 1023, 1050, 1267, 1336, 1277, 1278, 1279, 1280, 2093, 1028, 1080, 1217, 1349, 1136, 1137, 1143, and 2100.
Other devices under the same FDA submission
2 devices cleared under K033238 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Mobilett XP 01818363Siemens Healthcare GmbH · System, x-ray, mobile5 reports
- Mobilett XP hybrid 01818454Siemens Healthcare GmbH · System, x-ray, mobile2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page