Skip to content

Mobilett XP 01818363Recalls & reported problems

Siemens Healthcare GmbH · Mobile Generator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04056869009087), received between Dec 22, 2020 and Jul 8, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K033238); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Crack1
  • Loose or Intermittent Connection1
  • Unintended Collision1
  • Unintended Electrical Shock1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Electric Shock1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2024Unintended Electrical Shock; Unintended Electrical Shock

    SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE MOBILETT XP SYSTEM. ON JUNE 26, 2024, SIEMENS RECEIVED INFORMATION FROM THE USER THAT THE OPERATORS WERE EXPERIENCING AN ELECTRIC SHOCK LIKE SENSATION WHEN TOUCHING THE SYSTEM HANDLE. NO FURTHER INFORMATION REGARDING THIS EVENT WAS AVAILABLE, AND IT WAS INITIALLY CLASSIFIED AS NOT REPORTABLE. IT WAS ASSUMED THAT THE SENSATION WAS CAUSED BY ELECTROSTATIC DISCHARGE. ON JULY 2, 2024, ADDITIONAL INFORMATION WAS RECEIVED THAT THE GROUND WIRE OF THE MAIN CABLE WAS BROKEN. IT WAS THEN DECIDED TO REPORT THE ISSUE, SINCE A HAZARDOUS SITUATION COULD NOT BE EXCLUDED IN CASE A SECOND FAULT CONDITION WAS PRESENT IN THE SYSTEM. IT WAS CONFIRMED THAT THERE ARE NO INJURIES RELATED TO THIS EVENT.

    Patient problems: Electric Shock; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19687992 ↗
  • MalfunctionSep 25, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    SIEMENS BECAME AWARE OF AN ISSUE WITH THE MOBILETT XP UNIT. PER CUSTOMER¿S REQUEST, A LOCAL SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE FACILITY TO CHECK LIGHT FIELD/RADIATION FIELD ACCURACY. DURING THE INSPECTION THE SERVICE ENGINEER NOTICED THAT THREE OF THE FOUR SCREWS OF THE SINGLE TANK FLANGE FIXATIONS WERE NOT TIGHTENED ANYMORE. THE COLLIMATOR WAS DE-INSTALLED, THE SCREWS WERE TIGHTENED, AND THE COLLIMATOR RE-INSTALLED AGAIN. WE HAVE NO INDICATIONS OF ADVERSE EFFECTS ON THE HEALTH STATUS OF PATIENTS, USERS OR THIRD PARTIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17807116 ↗
  • MalfunctionMar 23, 2022Mechanical Problem; Mechanical Problem

    THE SCREWS HOLDING THE COLLIMATOR COVER ON THE MOBILETT XP UNITS BECAME LOOSE OVER TIME. PERSONNEL AT THE USER'S SITE TOOK THE COVER APART AND REINSTALLED WITHOUT SIEMENS INVOLVEMENT. HOWEVER, THE COLLIMATOR REMAINED UNSTABLE/WOBBLY. THE COLLIMATOR WAS UNABLE TO ROTATE TO 180 DEGREES AS DESIGNED. THE INSPECTION PERFORMED BY SIEMENS SERVICE ENGINEER CONFIRMED THAT THE SCREWS SECURING THE COLLIMATOR TO THE X-RAY TUBE HAD LOOSENED. FOLLOWING CLEANING AND LUBRICATION OF THE ROTATING PARTS, THE COLLIMATOR COULD BE ROTATED NORMALLY AGAIN AND BECAME STEADY FOLLOWING THE REPAIRS. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13869973 ↗
  • MalfunctionAug 26, 2021Mechanical Problem; Mechanical Problem

    SIEMENS BECAME AWARE OF AN INCIDENT WITH THE MOBILETT XP UNIT. THE OPERATOR WAS MOVING THE UNIT TO THE PATIENT ROOM WHEN HE HEARD A NOISE NEAR THE X-RAY TUBE. WHEN THE OPERATOR TOUCHED THE X-RAY TUBE TO CHECK IT, THE COLLIMATOR FELL DOWN. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12370298 ↗
  • MalfunctionDec 22, 2020Crack; Unintended Collision

    SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WITH THE MOBILETT XP DEVICE. DUE TO A VIOLENT IMPACT, THE RIGHT SIDE OF THE SUPPORT ARM HAS SEVERE CRACKS, WHICH MAY CAUSE THE TUBE TO FALL. THE USER WAS ADVISED TO ISOLATE THE UNIT AND STOP ALL USAGE UNTIL REPAIRS ARE COMPLETED. THERE ARE NO INJURIES ATTRIBUTED TO THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11059498 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033238. One submission often covers many models, so check the product and lot details in each record.

  • Z-1985-2014 · May 9, 2014 · Terminated
    Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and pre
  • Z-0171-2011 · Oct 4, 2010 · Terminated
    Mobilett XP and Mobilett XP Hybrid, Mobile radiology device System, x-ray, mobile
  • Z-0718-2011 · Oct 4, 2010 · Terminated
    Mobilett XP Hybrid/Digital/CR Intended use: mobile x-ray system
  • Z-0749-2007 · Mar 21, 2007 · Terminated
    Mobilett XP, mobile X-ray system, Model number 1818454

Other System, x-ray, mobile devices

All System, x-ray, mobile devices →

Sources and updates