Why it was recalled

Synchronization problems: units that have both a main programmable console and an auxiliary programmable console, when simultaneously, injection parameters are programmed into the unit's main console and the unit's auxiliary consoles, mis-synchronization could occur resulting in a variance between the intended injection parameters (those programmed) and the actual parameters of the injection deliv

FDA root cause category: Device Design

What the company is doing

On 9/20/2007, a notification was sent from Mallinckrodt to all direct accounts via certified mail. The notification details the potential for mis-synchronization of injection parameters if the main and auxiliary consoles are simultaneously programmed. The notification reminds end-users that simultaneous programming should not be performed and to ensure it is not performed, a notification label (provided in the notification) is to be affixed on both consoles of the injector system. Once the label is affixed to the system, end-users are instructed to complete the notification survey and return the survey to the Mallinckrodt. In addition, the notification includes an updated chapter insert (Chapter 11.1) for the injector system's product manual that emphasize the separation between programming of the main and auxiliary consoles. Questions regarding this matter may be directed to Mallinckrodt's Product Monitoring Department at 1-888-744-1414, menu option 1 and then menu option 2.

Recall details

Recall number
Z-0056-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 24, 2007
Posted by the FDA
Oct 13, 2007
Recalling firm
Mallinckrodt Inc
Quantity
176
Product code
DXT
FDA submissions
K963071
Distribution
Worldwide to include:U.S.A. states of - AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, & WI. and Foreign countries of Argentina, Belgium, Brazil, Canada, Germany, Japan, Singapore, & Zambia.

Product description

Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO.

Code information (models, lots, serial numbers)

Product Model Numbers : 902040; 904045; and 902045CI

Other devices under the same FDA submission

2 devices cleared under K963071 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.