Medical device recalls: October 2007
The FDA posted 77 medical device recall records in October 2007, from 41 recall events by 36 companies. 77 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Becton Dickinson & Company · 2 events, 5 records
- Boston Scientific Corporation · 2 events, 4 records
- Biomet, Inc. · 2 events, 3 records
- Medtronic Neuromodulation · 2 events, 2 records
- Pml Microbiologicals Inc. · 2 events, 2 records
- Smith & Nephew Inc · 1 event, 18 records
- Medtronic Inc. Cardiac Rhythm Managment · 1 event, 4 records
- Olympus America Inc. · 1 event, 4 records
FDA root cause
- Packaging process control · 19
- Process control · 14
- Device Design · 11
- Process design · 10
- Software design · 6
- Packaging change control · 2
- Package design/selection · 2
- Labeling mix-ups · 2
Not yet classified (77)
The FDA has posted these records but has not published a class for them yet.
- Z-0063-2008 · Medtronic, Inc. · posted Oct 31, 2007 · Terminated
Additional Warning in Instructions for Use for guide wire management. Complaints of wire lumen tearing during use or band marker becoming dislodged.
- Z-0065-2008 · Bio-Rad Laboratories · posted Oct 31, 2007 · Terminated
Incorrect Instruction: the package insert states to add 100L of conjugate; it should state to add 100uL.
- Z-0076-2008 · Biomet, Inc. · posted Oct 31, 2007 · Terminated
Incorrect Product -- Mislabeled as to size; a 18 mm x 65 mm tibula implant is enclosed in a package labeled as 12 mm x 80 mm.
- Z-0077-2008 · Biomet, Inc. · posted Oct 31, 2007 · Terminated
Incorrect Product -- Mislabeled as to size; a 18 mm x 65 mm tibula implant is enclosed in a package labeled as 12 mm x 80 mm.
- Z-0078-2008 · Gambro BCT, Inc. · posted Oct 31, 2007 · Terminated
Under specific conditions, apheresis machine could return up to 60mL of air to the patient.
- Z-0105-2008 · Olympus America Inc. · posted Oct 31, 2007 · Terminated
Slider knob does not move smoothly causing the user difficulty in advancing the coil sheath
- Z-0106-2008 · Olympus America Inc. · posted Oct 31, 2007 · Terminated
Slider knob does not move smoothly causing the user difficulty in advancing the coil sheath
- Z-0107-2008 · Olympus America Inc. · posted Oct 31, 2007 · Terminated
Slider knob does not move smoothly causing the user difficulty in advancing the coil sheath
- Z-0108-2008 · Olympus America Inc. · posted Oct 31, 2007 · Terminated
Slider knob does not move smoothly causing the user difficulty in advancing the coil sheath
- Z-0109-2008 · Boston Scientific Corporation · posted Oct 31, 2007 · Terminated
Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire.
- Z-0110-2008 · Philips Medical Systems (Cleveland) Inc · posted Oct 31, 2007 · Terminated
Two potential artifacts may appear in some images from head scans on the Brilliance 40/64 systems. These artifacts may lead to misdiagnosis.
- Z-0111-2008 · Philips Medical Systems (Cleveland) Inc · posted Oct 31, 2007 · Terminated
Two potential artifacts may appear in some images from head scans on the Brilliance 40/64 systems. These artifacts may lead to misdiagnosis.
- Z-0116-2008 · Boston Scientific Corporation · posted Oct 31, 2007 · Terminated
Device lacked manufacturing finishing process resulting in endoscope damage.
- Boston Scientific Radial Jaw Large Capacity Biopsy Forceps with Needle, Sterile, 240 mm. Box of 20 ↗Z-0117-2008 · Boston Scientific Corporation · posted Oct 31, 2007 · Terminated
Device lacked manufacturing finishing process resulting in endoscope damage.
- Boston Scientific Radial Jaw Large Capacity Biopsy Forceps with Needle, Sterile, 240 mm. Box of 40 ↗Z-0118-2008 · Boston Scientific Corporation · posted Oct 31, 2007 · Terminated
Device lacked manufacturing finishing process resulting in endoscope damage.
- Z-0102-2008 · Biomet, Inc. · posted Oct 27, 2007 · Terminated
Incorrect or lack of threading: The plug component was manufactured incorrectly and does not fit with the screw hole as intended.
- Z-0103-2008 · Boston Scientific · posted Oct 27, 2007 · Terminated
Missing holes: Catheters are missing the catheter's side holes.
- Z-0104-2008 · Boston Scientific · posted Oct 27, 2007 · Terminated
Missing holes: Catheters are missing the catheter's side holes.
- Z-0016-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0017-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0018-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0019-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0020-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0021-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0022-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0023-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0024-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0025-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0026-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0027-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0028-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0029-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0030-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0031-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0032-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0033-2008 · Smith & Nephew Inc · posted Oct 25, 2007 · Terminated
The carton label may not reflect the correct size of the acetabular cup.
- Z-0062-2008 · Philips Medical Systems · posted Oct 25, 2007 · Terminated
Inadequate shipping package causing damage to control boards resulting in unit failure. Failure to deliver therapy (defibrillation, pacing or cardioversion) may occur as a result of undetected damage to the PCB assembly.
- Z-0067-2008 · Medtronic Inc. Cardiac Rhythm Managment · posted Oct 25, 2007 · Terminated
Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company recommends against new implants of the…
- Z-0068-2008 · Medtronic Inc. Cardiac Rhythm Managment · posted Oct 25, 2007 · Terminated
Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company recommends against new implants of the…
- Z-0069-2008 · Medtronic Inc. Cardiac Rhythm Managment · posted Oct 25, 2007 · Terminated
Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company recommends against new implants of the…
- Z-0070-2008 · Medtronic Inc. Cardiac Rhythm Managment · posted Oct 25, 2007 · Terminated
Medtronic Cardiac Rhythm Disease Management has voluntarily suspended worldwide distribution of the Sprint Fidelis family of defibrillation leads because of the potential for lead fractures. In addition, the company recommends against new implants of the…
- Z-0072-2008 · Stryker Howmedica Osteonics Corp. · posted Oct 25, 2007 · Terminated
Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will…
- Z-0073-2008 · Stryker Howmedica Osteonics Corp. · posted Oct 25, 2007 · Terminated
Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will…
- Z-0101-2008 · Steris Corporation Hopkins Facility · posted Oct 24, 2007 · Terminated
Loose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the attached endoscope.
- Z-0058-2008 · Remington Medical Inc. · posted Oct 23, 2007 · Terminated
Sterility, inadequate -- Pathway may not be sterile as labeled.
- Z-0066-2008 · Becton Dickinson and Company · posted Oct 23, 2007 · Terminated
Microkeratome blade may stop oscillating and transverse the eye causing injury to the cornea
- Z-0042-2008 · Esa Biosciences, Inc. · posted Oct 18, 2007 · Terminated
Level 1 and Level 2 Control values were incorrectly assigned.
- Z-0043-2008 · Esa Biosciences, Inc. · posted Oct 18, 2007 · Terminated
Level 1 and Level 2 Control values were incorrectly assigned.
- Z-0044-2008 · Becton Dickinson & Company · posted Oct 18, 2007 · Terminated
Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.
- Z-0045-2008 · Becton Dickinson & Company · posted Oct 18, 2007 · Terminated
Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.
- Z-0046-2008 · Becton Dickinson & Company · posted Oct 18, 2007 · Terminated
Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.
- Z-0061-2008 · Zoll Lifecor Corporation · posted Oct 18, 2007 · Terminated
Possible malfunction of the battery pack. A significant electrostatic discharge (ESD) can cause the battery pack to malfunction. The malfunction causes the battery pack to discharge more rapidly and prevents the pack from recharging.
- Z-0002-2008 · Beckman Coulter Inc · posted Oct 16, 2007 · Terminated
Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.
- Z-0056-2008 · Mallinckrodt Inc · posted Oct 13, 2007 · Terminated
Synchronization problems: units that have both a main programmable console and an auxiliary programmable console, when simultaneously, injection parameters are programmed into the unit's main console and the unit's auxiliary consoles, mis-synchronization…
- Z-0040-2008 · Medtronic Neuromodulation · posted Oct 12, 2007 · Terminated
Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).
- Z-0055-2008 · Advanced Bionics Corp · posted Oct 12, 2007 · Terminated
Mislabeled: Product labeled with the incorrect shelf-life information.
- Z-0037-2008 · Medtronic Neuromodulation · posted Oct 11, 2007 · Terminated
A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .
- Z-0038-2008 · Radiometer America Inc · posted Oct 11, 2007 · Terminated
Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur…
- Z-0039-2008 · Diagnostica Stago, Inc. · posted Oct 11, 2007 · Terminated
Incorrect results. Internal investigation determined the presence of a defective measurement head which can impact the results of clotting tests by either yielding no result or rendering an error message.
- Z-0041-2008 · Naviscan PET Systems · posted Oct 11, 2007 · Terminated
Naviscan has identified a potential safety issue with all versions of PEM View Software commonly used for analyzing patient images acquired on the PEMFIex product line. 1. The Positron Emission Mammography Standardized Uptake Value (PEMSUV) and distance…
- Z-0009-2008 · Encore Medical, Lp · posted Oct 10, 2007 · Terminated
Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site.
- Z-0010-2008 · Encore Medical, Lp · posted Oct 10, 2007 · Terminated
Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site.
- Z-0011-2008 · Pml Microbiologicals Inc. · posted Oct 10, 2007 · Terminated
Marketed a regulated product without a 510(k).
- Z-0014-2008 · Pml Microbiologicals Inc. · posted Oct 10, 2007 · Terminated
Plates manufactured without vancomycin. Possibility for false resistant results.
- Z-0057-2008 · I-Flow Corporation · posted Oct 6, 2007 · Terminated
Misassembly: The ON-Q PainBuster PMB01 package may contain a Soaker catheter instead of a Non-Soaker catheter. As a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead…
- Z-0003-2008 · Meretek Diagnostics, Inc. · posted Oct 4, 2007 · Terminated
Diagnostic test kit for Helicobacter pylori was distributed for adult and pediatric use, but was not approved for pediatric use.
- Z-0006-2008 · Datex - Ohmeda, Inc · posted Oct 4, 2007 · Terminated
Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
- Z-0008-2008 · Lemaitre Vascular, Inc. · posted Oct 4, 2007 · Terminated
Sterility of device may be compromised due to breach in sterile barrier.
- Z-0015-2008 · Thermo Fisher Scientific · posted Oct 4, 2007 · Terminated
Method is susceptible to positive interference from endogenous ALT in patient samples.
- Z-0034-2008 · Louisville Apl Diagnostics Inc · posted Oct 4, 2007 · Terminated
Product mislabeled; outer container labeled with incorrect lot number.
- Z-0007-2008 · Tosoh Bioscience Inc · posted Oct 3, 2007 · Terminated
Firm discovered that a deterioration value of approximately 20% was found when the product of this lot was measured. Tosoh AIA confirmed that deterioration of the product occurred over time and resulted in a deviation from the assigned value.
- Z-0004-2008 · Abbott Spine · posted Oct 2, 2007 · Terminated
Tip of PathFinder surgical instrument broke during procedure.
- Z-0005-2008 · Abbott Spine · posted Oct 2, 2007 · Terminated
Tip of PathFinder surgical instrument broke during procedure.
- Z-0012-2008 · Becton Dickinson & Company · posted Oct 2, 2007 · Terminated
Complaints received of rust on the blades of BD Bard-Parker Surgical blades.
- Z-0013-2008 · Becton Dickinson & Company · posted Oct 2, 2007 · Terminated
Complaints received of rust on the blades of BD Bard-Parker Surgical blades.
- Z-0035-2008 · Heraeus Kulzer Inc., Dental Products Division · posted Oct 2, 2007 · Terminated
The material may solidify in the bottle, and thus be rendered unusable.
- Z-0036-2008 · Heraeus Kulzer Inc., Dental Products Division · posted Oct 2, 2007 · Terminated
The material may solidify in the bottle, and thus be rendered unusable.