ADVIA Centaur TnI-Ultra Assay, Troponin Immunoassay, 500… recallZ-0580-2008 · Siemens Medical Solutions Diagnostics · initiated Jun 7, 2007

Terminated All recalls posted in June 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect results: Falsely elevated Troponin values have been reported which are inconsistent with the patient's clinical picture and test negative by other Troponin assays.

FDA root cause category: Material/Component Contamination

What the company is doing

Two Support Bulletins (078N0361-01 & 078N0361-02) were sent by e-mail to all affected Siemens Diagnostics Branches to inform them of this issue. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. Two Customer Bulletins (078D0647-01 & 078D0651-01) were sent by e-mail to the Siemens Diagnostics Branches for communication with the affected customers. The first one was sent on 6/07/07 and the second one was sent on 7/05/07. These bulletins state that if a false positive is suspected, the sample should be diluted and retested.

Recall details

Recall number
Z-0580-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 7, 2007
Posted by the FDA
Jun 5, 2008
Recalling firm
Siemens Medical Solutions Diagnostics
Quantity
Domestic: 31,497 kits; Foreign: 27,321 kits
Product code
MMI
FDA submissions
K053020
Distribution
Worldwide Distribution - USA, Argentina, Austria, Australia, Brazil, Canada, Chile, Colombia, Egypt, France, Germany, Greece, Hong Kong, India, Israel, Italy, Malaysia, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela

Product description

ADVIA Centaur TnI-Ultra Assay, Troponin Immunoassay, 500 tests - Part/Catalog # 02790309, and 100 tests - Part/Catalog # 02789602. For in vitro diagnostic use, Siemens Medical Solutions Diagnostics, Tarrytown, NY 10591

Code information (models, lots, serial numbers)

All lots. 500 tests: Part/Catalog #02790309, and 100 tests: Part/Catalog # 02789602.

Other devices under the same FDA submission

4 devices cleared under K053020 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.