Medical device recalls: June 2008
The FDA posted 51 medical device recall records in June 2008, from 22 recall events by 19 companies. 51 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Stryker Howmedica Osteonics Corp. · 2 events, 8 records
- AGFA Corp. · 2 events, 2 records
- GE Healthcare · 2 events, 2 records
- Chattanooga Corp Inc · 1 event, 15 records
- Bio-Rad Laboratories · 1 event, 4 records
- California Medical Laboratories Inc · 1 event, 3 records
- Cardinal Health 303 Inc DBA Alaris Products · 1 event, 3 records
- Medline Industries Inc · 1 event, 3 records
FDA root cause
- Device Design · 23
- Process design · 7
- Process control · 7
- Software design · 5
- Nonconforming Material/Component · 4
- Software change control · 1
- Process change control · 1
- Other · 1
Not yet classified (51)
The FDA has posted these records but has not published a class for them yet.
- Z-1167-2008 · ZOLL Medical Corporation, World Wide Headquarters · posted Jun 28, 2008 · Terminated
Incorrect Patient Records: Patient records that have been stored, then later transmitted or printed may not be the correct record. Records that are transmitted or printed immediately following acquisition are correct.
- Z-1606-2008 · Siemens Healthcare Diagnostics Inc. · posted Jun 28, 2008 · Terminated
False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine…
- Z-2196-2008 · Siemens Medical Solutions USA, Inc · posted Jun 28, 2008 · Terminated
The firm sent an Update Instruction, "TH003/08/S" for Motion Enable Switch/holder assembly. The switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury may result.
- Z-1241-2008 · AGFA Corp. · posted Jun 27, 2008 · Terminated
MIsidentification: The electrocardiogram (ECG) data of one patient is misidentified as the data of another patient.
- Z-1242-2008 · AGFA Corp. · posted Jun 27, 2008 · Terminated
Incorrect Results: Erroneous echocardiographic measurement values due to mathematical formula being mis-configured.
- Z-1188-2008 · Kentec Medical Inc · posted Jun 26, 2008 · Terminated
Incompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional diode and higher resistor in its existing configuration making them inoperable when used to interface with…
- Z-1236-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 26, 2008 · Terminated
Delamination: Loosening of the titanium plasma coating on the product was observed on twelve non-distributed units.
- Z-1160-2008 · Beckman Coulter Inc · posted Jun 25, 2008 · Terminated
False negative results when used to test serum samples (as compared to plasma samples).
- Z-1162-2008 · Medline Industries Inc · posted Jun 25, 2008 · Terminated
Separation from Bushing: The breathing bag may separate from the bushing during use.
- Z-1163-2008 · Medline Industries Inc · posted Jun 25, 2008 · Terminated
Separation from Bushing: The breathing bag may separate from the bushing during use.
- Z-1164-2008 · Medline Industries Inc · posted Jun 25, 2008 · Terminated
Separation from Bushing: The breathing bag may separate from the bushing during use.
- Z-1218-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1219-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1220-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1221-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1222-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1223-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1224-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Devices have the potential to electric shock and burn
- Z-1225-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1226-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1227-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1228-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1229-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1230-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1231-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1232-2008 · Chattanooga Corp Inc · posted Jun 25, 2008 · Terminated
Device malfunction: causing electric shock and burn.
- Z-1161-2008 · GE Healthcare · posted Jun 24, 2008 · Terminated
Failure to alarm: An active monitor could become disconnected from the unit without any notification or alarm.
- Z-1155-2008 · GE Healthcare Clinical Systems · posted Jun 12, 2008 · Terminated
Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause…
- Z-1168-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1169-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1170-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1171-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1172-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1173-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1174-2008 · Stryker Howmedica Osteonics Corp. · posted Jun 12, 2008 · Terminated
Foreign material: Some of the parts tested exceeded Stryker Orthopaedics internal acceptance criteria for manufacturing residuals.
- Z-1180-2008 · Intuitive Surgical, Inc. · posted Jun 12, 2008 · Terminated
Delay in responding: In certain circumstances, the device may not respond immediately to a user's command such as master clutch or camera control.
- Z-1156-2008 · Bio-Rad Laboratories · posted Jun 11, 2008 · Terminated
False negative results due to reagent packs exhibiting low signal.
- Z-1157-2008 · Bio-Rad Laboratories · posted Jun 11, 2008 · Terminated
False negative results due to reagent packs exhibiting low signal.
- Z-1158-2008 · Bio-Rad Laboratories · posted Jun 11, 2008 · Terminated
False negative results due to reagent packs exhibiting low signal.
- Z-1159-2008 · Bio-Rad Laboratories · posted Jun 11, 2008 · Terminated
False negative results due to reagent packs exhibiting low signal.
- Z-1175-2008 · California Medical Laboratories Inc · posted Jun 11, 2008 · Terminated
Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip of the product.
- Z-1176-2008 · California Medical Laboratories Inc · posted Jun 11, 2008 · Terminated
Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip of the product.
- Z-1177-2008 · California Medical Laboratories Inc · posted Jun 11, 2008 · Terminated
Fragment of material: The product may have a loose, stainless steel stringer flash located at the distal tip of the product.
- Z-1178-2008 · Cardinal Health · posted Jun 11, 2008 · Terminated
The product has the potential of being non-sterile.
- Z-0842-2008 · GE Healthcare · posted Jun 10, 2008 · Terminated
The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
- Z-1140-2008 · B & K Medical Systems, Inc. · posted Jun 10, 2008 · Terminated
Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization.
- Z-1710-2008 · Cardinal Health 303 Inc DBA Alaris Products · posted Jun 10, 2008 · Terminated
A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failures of the U19 Socketed Integrated Circuits (U19 IC) on the logic board of the Alaris Pump module…
- Z-1711-2008 · Cardinal Health 303 Inc DBA Alaris Products · posted Jun 10, 2008 · Terminated
A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System modules…
- Z-1712-2008 · Cardinal Health 303 Inc DBA Alaris Products · posted Jun 10, 2008 · Terminated
A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the Alaris Syringe…
- Z-0580-2008 · Siemens Medical Solutions Diagnostics · posted Jun 5, 2008 · Terminated
Incorrect results: Falsely elevated Troponin values have been reported which are inconsistent with the patient's clinical picture and test negative by other Troponin assays.
- Z-1139-2008 · Visx Incorporated, A Subsidiary Of Amo Inc · posted Jun 5, 2008 · Terminated
Firm received two complaints (including one injury complaint) associated with the movement of the VISX chair while in the patient loading ("fully swung out" unlocked) position under an IntraLase FS laser.