Skip to content

HA Solid Back Acetabular Shell recallZ-0073-2008 · Stryker Howmedica Osteonics Corp. · initiated Aug 30, 2007

Terminated All recalls posted in October 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will decrease the gap between shell and liner on the opposing side, resulting in interference.

FDA root cause category: Process control

What the company is doing

Recall notification letters were sent out on August 30, 2007 with return receipt.

Recall details

Recall number
Z-0073-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 30, 2007
Posted by the FDA
Oct 25, 2007
Recalling firm
Stryker Howmedica Osteonics Corp.
Quantity
142 units
Product code
MEH
FDA submissions
K983382
Distribution
Product was distributed nationwide and one unit to Brazil.

Product description

HA Solid Back Acetabular Shell; Arc Deposited HA Coated Included Dome hole Plug; Hip prosthesis component. Sterile; made in USA; Howmedica Osteonics Corp.; 325 Corporate Drive; Mahwah, NJ 07430 A Subsidiary of Stryker Corp.

Code information (models, lots, serial numbers)

540-11-52E TRIDENT PSL HA SOLID BACK 52mm 8NKMJD 540-11-54F TRIDENT PSL HA SOLID BACK 54mm 3H7MJD 540-11-58G TRIDENT PSL HA SOLID BACK 58mm 78AMJD 540-11-58G TRIDENT PSL HA SOLID BACK 58mm 9T9MJD 540-11-66I TRIDENT PSL HA SOLID BACK 66mm 4THMJD 540-11-66I TRIDENT PSL HA SOLID BACK 66mm 4TJMJD 540-11-66I TRIDENT PSL HA SOLID BACK 66mm RLKMHD 540-11-72J TRIDENT PSL HA SOLID BACK 72mm 1WNMJD

Other devices under the same FDA submission

9 devices cleared under K983382 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 44780).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page